Errata in ISO 8253
These are the 4 entries of the errata registry on ISO 8253-1:2010 and ISO 8253-2:2009, in the order the registry lists them. The registry’s introduction says how each claim is checked against the printed page and what the status of an entry means.
ISO 8253-1:2010, 11.1 (the filter standard cited as IEC 61620)
Section titled “ISO 8253-1:2010, 11.1 (the filter standard cited as IEC 61620)”- Location: 11.1, the paragraph on how the ambient noise is measured, printed p. 16.
- The print: “The measurements shall meet the requirements for class 1 sound level meters of IEC 61672-1 and IEC 61620 and have a noise floor at least 6 dB below the sound pressure level being measured.”
- The problem: the one-third-octave measurement 11.1 asks for needs the filter standard, IEC 61260, “Octave-band and fractional-octave-band filters”, which Clause 2 lists among the normative references and 12.4 names for the “one-third-octave-band filter set”. Clause 2 lists no IEC 61620 and no other clause cites one: the digits are transposed.
- Evidence: the citation read against the list it should come from. Verified on PDF page 24 (printed p. 16), PDF page 9 (printed p. 1) and PDF page 30 (printed p. 22) of BS EN ISO 8253-1:2010, whose endorsement notice adopts ISO 8253-1:2010 without modification.
- Library behaviour: none needed.
check_audiometric_ambient_noisetakes the one-third-octave levels as measured, and its noise-floor flag is the 6 dB margin of the same sentence. - Status: unreported (cross-reference defect, no numerical consequence).
ISO 8253-1:2010, Table 3, footnote a (the earphone names marked with the diffuse-field footnote)
Section titled “ISO 8253-1:2010, Table 3, footnote a (the earphone names marked with the diffuse-field footnote)”- Location: Table 3, footnote a, printed p. 18.
- The print: “The values given are based on measurements using pure tones in a free sound field and using Telephonics TDH39 with MX 41/AR cushions and Beyer DT48 earphones”, with the mark of footnote d after each of the three product names. Footnote d reads “Data are valid for an artificial diffuse field according to ISO 4869-1”; footnote e reads “This is a product available commercially. This information is given for the convenience of users of this International Standard and does not constitute an endorsement by ISO of this product.”
- The problem: the three product names carry the marker of footnote d, which describes the ER-3A and HDA 200 data and says the opposite of the sentence it is attached to: a diffuse field where footnote a says a free field with pure tones. The note the product names need is footnote e, the disclaimer the table’s own column headings attach to the other two products: the ER-3A heading carries the marks d, e and f, the HDA 200 heading d, e and g.
- Evidence: the footnote read against the two footnotes it cites and the column headings. Verified on PDF page 26 (printed p. 18) of BS EN ISO 8253-1:2010.
- Library behaviour: none needed. The attenuation values of
EARPHONE_ATTENUATIONare the table’s cells and do not depend on the footnote marks. - Status: unreported (editorial, no numerical consequence).
ISO 8253-1:2010, A.3.6 (a misspelt word in the masking-noise uncertainty)
Section titled “ISO 8253-1:2010, A.3.6 (a misspelt word in the masking-noise uncertainty)”- Location: A.3.6, “Masking noise, ”, printed p. 25.
- The print: “However, a standard uncertainty of 2 dB may provisionably be attributed to with a normal probability distribution if masking noise is applied.”
- The problem: “provisionably” for “provisionally”. The word is the one that tells the reader the 2 dB is a stopgap until better figures exist, so the misspelling sits on the qualifier of the value.
- Evidence: the sentence as printed. Verified on PDF page 33 (printed p. 25) of BS EN ISO 8253-1:2010, whose endorsement notice adopts ISO 8253-1:2010 without modification.
- Library behaviour: none needed.
audiometric_uncertaintytakes the 2 dB of A.3.6 whenmasked=True; no number depends on the word. - Status: unreported (editorial, no numerical consequence).
ISO 8253-2:2009, Annex B (Table B.1 announced from 200 Hz, printed from 125 Hz)
Section titled “ISO 8253-2:2009, Annex B (Table B.1 announced from 200 Hz, printed from 125 Hz)”- Location: Annex B (informative), the paragraph above Table B.1 and the table, printed p. 15.
- The print: “Table B.1 gives figures for the increased sound pressure levels at test frequencies from 200 Hz to 12 500 Hz at angles of incidence 45° and 90°”. The table’s first rows are 125 Hz (0,5 dB and 1 dB) and 160 Hz (1 dB and 1,5 dB).
- The problem: the table runs from 125 Hz, two rows below the range the text announces for it. A reader who trusts the sentence discards two printed rows; one who trusts the table has a text that says they are not there. The 125 Hz row is the one sound field audiometry needs, since its primary range starts at 125 Hz (Clause 1).
- Evidence: the sentence read against the rows it introduces. Verified on PDF page 21 (printed p. 15) of ISO 8253-2:2009.
- Library behaviour:
INCIDENCE_CORRECTIONScarries every printed row, 125 Hz to 12 500 Hz, andincidence_correctionreads them. The conformance check “ISO 8253-2:2009 Table B.1” counts all 48 cells, the four below 200 Hz among them. - Status: unreported.