Errata in ANSI S1.4
This is the entry of the errata registry on ANSI S1.4-1983. The registry’s introduction says how each claim is checked against the printed page and what the status of an entry means.
ANSI S1.4-1983, Table V, 20 Hz type 2 cell (a plus sign that lost its bar)
Section titled “ANSI S1.4-1983, Table V, 20 Hz type 2 cell (a plus sign that lost its bar)”- Location: clause 5.2, Table V “Tolerance limits on relative response levels for sound at random incidence measured on an instrument’s calibration range”, 20 Hz row, type 2 column (printed p. 6).
- The print: the cell reads ”+ 3”, with no second term. Its column neighbours at 10, 12.5 and 16 Hz read ”+ 5, − ∞”, and the type 0 and type 1 cells of its own row read ”± 2” and ”± 2.5”.
- The problem: the table has one notation for an upper-only limit, a pair ”+ n, − ∞”, and it is used three rows above this cell in the same column. This cell uses neither that notation nor the ”± n” of its row, so it is either a limit written in a form the table uses nowhere else or a ”±” whose bar failed to print. IEC 651:1979 Table V, of which this table is the US counterpart and with which the type 2 column agrees at all thirty-three other rows, prints “±3” at exactly this cell. The intended reading is ±3 dB.
- Evidence: the cell and its column neighbours, read on PDF page 16 (printed p. 6) of ANSI S1.4-1983, against the same cell on PDF page 10 (printed p. 8, marked “[IEC page 19]”) of BS 5969:1981, the identical British adoption of IEC 651:1979.
- Library behaviour:
_ANSI_S14_TABLE5_12inweighting_compliance.pyand itsreference_datatwin carry −3 dB as the 20 Hz type 2 lower limit, the stricter of the two readings, with the note beside them.test_b_masks_match_reference_datapins the two transcriptions to each other. No shipped verdict moves: the realized B weighting sits 0.05 dB below nominal at 20 Hz and clears either reading. - Status: unreported.
Created and maintained by José Manuel Requena PlensORCIDAbout the maintainerRelease notesChangelogAll projects