# Errata in ANSI S1.4

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/reference/errata/ansi-s1-4/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/reference/errata/ansi-s1-4/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-10-28
Updated: 2026-11-16T21:36:14Z
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Errata in ANSI S1.4: 1 defect in the printed text, with the evidence and what phonometry does about it.

---

This is the entry of the [errata registry](https://docs.jmrp.io/phonometry/reference/errata/) on ANSI S1.4-1983. The registry's introduction says how each claim is checked against the printed page and what the status of an entry means.

## ANSI S1.4-1983, Table V, 20 Hz type 2 cell (a plus sign that lost its bar)

- **Location:** clause 5.2, Table V "Tolerance limits on relative response
  levels for sound at random incidence measured on an instrument's calibration
  range", 20 Hz row, type 2 column (printed p. 6).
- **The print:** the cell reads "**+ 3**", with no second term. Its column
  neighbours at 10, 12.5 and 16 Hz read "+ 5, − ∞", and the type 0 and type 1
  cells of its own row read "± 2" and "± 2.5".
- **The problem:** the table has one notation for an upper-only limit, a pair
  "+ n, − ∞", and it is used three rows above this cell in the same column.
  This cell uses neither that notation nor the "± n" of its row, so it is
  either a limit written in a form the table uses nowhere else or a "±" whose
  bar failed to print. IEC 651:1979 Table V, of which this table is the US
  counterpart and with which the type 2 column agrees at all thirty-three
  other rows, prints "**±3**" at exactly this cell. The intended reading is
  ±3 dB.
- **Evidence:** the cell and its column neighbours, read on PDF page 16
  (printed p. 6) of ANSI S1.4-1983, against the same cell on PDF page 10
  (printed p. 8, marked "[IEC page 19]") of BS 5969:1981, the identical
  British adoption of IEC 651:1979.
- **Library behaviour:** `_ANSI_S14_TABLE5_12` in
  [`weighting_compliance.py`](https://github.com/jmrplens/phonometry/blob/main/src/phonometry/filters/weighting_compliance.py)
  and its `reference_data` twin carry −3 dB as the 20 Hz type 2 lower limit,
  the stricter of the two readings, with the note beside them.
  `test_b_masks_match_reference_data` pins the two transcriptions to each
  other. No shipped verdict moves: the realized B weighting sits 0.05 dB below
  nominal at 20 Hz and clears either reading.
- **Status:** unreported.