Errata in IEC 62585

These are the 3 entries of the errata registry on IEC 62585:2012, in the order the registry lists them. The registry’s introduction says how each claim is checked against the printed page and what the status of an entry means.

IEC 62585:2012, Table I.2 (a coverage factor its own degrees of freedom do not give)

Section titled “IEC 62585:2012, Table I.2 (a coverage factor its own degrees of freedom do not give)”
  • Location: Annex I, Table I.2 “Uncertainty example for a frequency of 1 kHz” (printed folio 38), the last row, “Expanded uncertainty of ”.
  • The print: “(normal) k = 2,11” beside “Effective degree of freedom = 29,98”, with the expanded uncertainty printed “0,12” and a subscript guard digit “(4)”, that is 0,124 dB, over a combined standard uncertainty of “0,059 0” dB.
  • The problem: clause 5 asks for the coverage factor that gives a level of confidence of 95 %, and Annex I computes the effective degrees of freedom by Welch-Satterthwaite “thereby enabling the coverage factor k to be selected to provide a level of confidence of 95 %”. The table’s own components give dB and , both as printed, and the Student factor for 95 % at 29.98 degrees of freedom is 2.04, not 2.11; 2.11 is the factor for about 17 degrees of freedom. The expanded uncertainty that follows is dB, printed with its guard digit 0.12(1); the printed 0.12(4) is dB. To the two decimals the correction is quoted to (component a14), both read 0.12 dB.
  • Evidence: the budget recomputed from the fifteen printed values and divisors, and the Student quantile evaluated at the printed degrees of freedom. Verified on PDF page 40 (printed p. 38) of BS EN 62585:2012, the English text of EN 62585:2012, which is IEC 62585:2012 unchanged.
  • Library behaviour: metrology.correction_uncertainty_budget takes the coverage factor from the effective degrees of freedom, 2.042 here, and the conformance rows on Table I.2 pin that factor and the expanded uncertainty it gives, 0.121 dB, with the printed ones named as the erratum (tests/metrology/test_free_field_corrections.py).
  • Status: unreported.

IEC 62585:2012, Table H.1 (the exponent of index 31)

Section titled “IEC 62585:2012, Table H.1 (the exponent of index 31)”
  • Location: Annex H, Table H.1 “Exact frequencies for one-twelfth-octave steps over one decade” (printed folio 35), the row of index 31.
  • The print: “Exact ” , beside “Exact calculated” 5,956 621 kHz.
  • The problem: every other row prints the exponent , which is Formula (H.1), with , and the calculated value beside it is . would be 2.441 kHz, outside the decade the table covers. The exponent reads .
  • Evidence: Formula (H.1) evaluated for the 41 indices; all 41 calculated values reproduce to the seven significant digits printed. Verified on PDF page 37 (printed p. 35) of BS EN 62585:2012, the English text of EN 62585:2012, which is IEC 62585:2012 unchanged.
  • Library behaviour: metrology.exact_frequencies evaluates Formula (H.1) and never reads the exponent column, so no change was required; the conformance row on Table H.1 checks the 41 calculated values.
  • Status: unreported.

IEC 62585:2012, Formulas (E.4) to (E.6) (the two readings in the coupler exchanged)

Section titled “IEC 62585:2012, Formulas (E.4) to (E.6) (the two readings in the coupler exchanged)”
  • Location: Annex E, Formulas (E.4), (E.5) and (E.6) (printed folio 26), against Figure E.1 and its Formulas (E.3A) and (E.3B) (printed folio 25), the list of symbols under (E.6), and descriptors a3 and a4 of Table I.1 (printed folio 37).
  • The print: Figure E.1: ” (E.3A)” and ” (E.3B)”, with drawn at the reference microphone and at the sound level meter; the symbol list: ” is the sound pressure level at reference microphone in the comparison coupler”, ” is the sound pressure level at sound level meter”; Table I.1: ” Level measurement – reference microphone in comparison coupler”, ” Level measurement – sound level meter in comparison coupler”. Then ” (E.6)”.
  • The problem: the correction is the meter’s free-field response less its response in the coupler, , as (E.5) states, and with the definitions of Figure E.1 that is . Formulas (E.4) to (E.6) are (D.5) to (D.7) of the calibrator with and renamed, so they take as the meter and as the reference, and at the meter. Read with the figure’s own labels, (E.6) comes out away from the correction. Those are the deviations of each channel’s indication from the level in the coupler, and the method “requires neither absolute measurements nor an absolutely calibrated sound level meter” (E.1, printed folio 25) while the reference channel reads “the level of the output voltage from the microphone” (E.2 step 2, printed folio 26), so the error carries the difference of the two channels’ absolute sensitivities and is there at every frequency, tens of decibels when one channel reads in decibels re 1 V and the other in sound pressure level. Only when both channels read sound pressure level does it reduce to twice the difference of the two pressure responses. Either the figure, the symbol list and Table I.1 exchange the two labels, or the formulas do; the page cannot hold both.
  • Evidence: (E.1) to (E.3B) substituted into (E.4), which does not reduce to with the figure’s labels and does with the formulas’. Verified on PDF pages 27 and 28 (printed pp. 25 and 26) and PDF page 39 (printed p. 37) of BS EN 62585:2012, the English text of EN 62585:2012, which is IEC 62585:2012 unchanged.
  • Library behaviour: metrology.comparison_coupler_correction names its inputs by what each reading is of, slm_coupler_level and reference_coupler_level, so neither labelling reaches it; the conformance row on Annex E builds the readings with Figure E.1’s labels and the tests pin the size of the discrepancy (tests/metrology/test_free_field_corrections.py).
  • Status: unreported.
Edit page

Published , updated