This is the entry of the errata registry on IEC 61672-1:2013. The registry’s introduction says how each claim is checked against the printed page and what the status of an entry means.
IEC 61672-1:2013, Table 3 (the class 1 lower limit at 6 300 Hz printed with a decimal point)
Section titled “IEC 61672-1:2013, Table 3 (the class 1 lower limit at 6 300 Hz printed with a decimal point)”- Location: Table 3, “Frequency weightings and acceptance limits”, row 6 300 Hz, column “Performance class 1”, printed p. 22.
- The print: “+1,5; -2.0”.
- The problem: every other number of the table writes its decimals with a comma, as the IEC does: the class 2 cell of the same row reads “±4,5” and the class 1 cell of the next row “+1,5; -2,5”. The lower limit at 6 300 Hz is the one cell written with a decimal point. The value is not in doubt, -2.0 dB, between the -1.5 dB above it and the -2.5 dB below it.
- Evidence: Verified on PDF page 24 (printed p. 22) of BS EN 61672-1:2013, the UK implementation of IEC 61672-1:2013 (second edition); the cell is legible and reads “-2.0” beside “+1,5;”.
- Library behaviour: none needed. The library holds the number,
-2.0 dB, in
filters.weighting_class_limits, whichverify_sound_level_meter_periodicreads for the frequency weightings. - Status: unreported.
Created and maintained by José Manuel Requena PlensORCIDAbout the maintainerRelease notesChangelogAll projects