# Mechanical mobility and the FRF family (ISO 7626-1)

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/vibration/structural/mechanical-mobility/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/vibration/structural/mechanical-mobility/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-07-11
Updated: 2026-11-16T21:36:14Z
Standards: ISO 7626-1:2011; ISO 7626-2:2015
Verified: 7 conformance checks (ISO 7626-1:2011; ISO 7626-2:2015) https://docs.jmrp.io/phonometry/reference/conformance/#room-building-acoustics
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: ISO 7626-1:2011 FRF family: receptance, mobility, accelerance and their reciprocals, conversion, the SDOF resonator and ISO 7626-2 acceptance.

---

Mechanical **mobility** is the complex ratio of a velocity response to the
force that produces it, $Y = v/F$. It is one member of a family of
motion-per-force **frequency-response functions** (FRFs): which one is used
depends only on whether the motion is a displacement, a velocity or an
acceleration, and each has a force-per-motion reciprocal. **ISO 7626-1:2011**
defines the whole family (Table 1, with the 3.1.2 mobility definition), and the
classic closed-form single-degree-of-freedom (SDOF) resonator serves as the
reference for those definitions. **ISO 7626-2:2015** adds the measurement side:
FRF estimation from measured signals and its acceptance criteria. This FRF
backbone underpins the structure-borne source and transmission standards:
ISO 9611, ISO 10846, EN 15657 and EN 12354-5.

![Receptance, mobility and accelerance magnitudes of the same single-degree-of-freedom resonator on a log-log frequency axis, each multiplied by a constant so that the three share one axis and peak at the same height, about twenty-five. All three peak at the same frequency and differ in slope on either side of it by one power of frequency: the receptance is flat below the resonance and falls steeply above it, the accelerance rises steeply below it and flattens above it, and the mobility sits between them with a plus one and minus one slope.](https://docs.jmrp.io/phonometry/media/mechanical_mobility.svg)

*The whole Table 1 family in one picture. The three curves peak at the same
frequency because they describe the same resonator; what distinguishes them is
the **slope** on either side of it, which differs by one power of $\omega$ from
one row of the table to the next. That is why receptance emphasises the
low-frequency, stiffness-controlled end, accelerance the high-frequency,
mass-controlled end, and mobility sits between them, and it is why mobility is
the currency of power flow (section 1). Each curve is scaled by the constant
printed in its legend entry ($k$, $k/\omega_0$ and $k/\omega_0^2$) so the
three peak at the same height and their shapes are comparable; the absolute
levels differ by orders of magnitude.*

**Show the code for this figure**

```python
import matplotlib.pyplot as plt
import numpy as np
from phonometry import vibration

m, k, c = 2.0, 8000.0, 5.0
f0 = vibration.resonance_frequency(m, k)
f = np.logspace(np.log10(f0 / 20.0), np.log10(f0 * 20.0), 600)
w0 = 2.0 * np.pi * f0
h = vibration.sdof_receptance(f, m, k, c)
y = vibration.convert_frf(h, "receptance", "mobility", frequency=f)
a = vibration.convert_frf(h, "receptance", "accelerance", frequency=f)

# Scaled by a constant each, not normalised by their own maxima: the point
# is that the three differ in slope, and dividing by the peak hides it.
for label, frf in (("receptance |H| x k", np.abs(h) * k),
                   ("mobility |Y| x k/w0", np.abs(y) * k / w0),
                   ("accelerance |A| x k/w0^2", np.abs(a) * k / w0**2)):
    plt.loglog(f, frf, label=label)
plt.axvline(f0, ls="--", color="0.6")
plt.xlabel("Frequency [Hz]"); plt.ylabel("Scaled FRF magnitude")
plt.legend(); plt.show()
```

## 1. The frequency-response-function family (Table 1)

For a harmonic motion $x\,\mathrm{e}^{j\omega t}$ the velocity is $j\omega x$ and the
acceleration $-\omega^2 x$, so all three motion-per-force FRFs follow from
the receptance $H$ by a power of $j\omega$, and each has a force-per-motion
reciprocal:

| Motion | FRF (motion / force) | Unit | Reciprocal (force / motion) | Unit |
|---|---|---|---|---|
| displacement | receptance $H = x/F$ | m/N | dynamic stiffness $1/H$ | N/m |
| velocity | mobility $Y = j\omega H$ | m/(N·s) | impedance $1/Y$ | N·s/m |
| acceleration | accelerance $A = -\omega^2 H$ | 1/kg | apparent mass $1/A$ | kg |

The figure at the top of this page is that table drawn: the same resonator seen
through all three rows, differing by one power of $\omega$ each.

`convert_frf` moves between any two of the six FRFs, pivoting through the
receptance. A **driving-point** FRF has the response and force at the same point
($i = j$); a **transfer** FRF has them at different points. Note that the
force-per-motion kinds are element-wise reciprocals: the *free* quantities of
ISO 7626-1, 3.1.4; the *blocked* matrix quantities of Table 1 do not invert
element-wise for multi-coordinate systems (Table 1 also names $F/a$ the
"effective mass", the quantity called apparent mass here).

```python
from phonometry import vibration

# A mobility of 2e-3 m/(N.s) at 80 Hz, expressed as the other FRFs:
Y = 2e-3
print(round(abs(vibration.convert_frf(Y, "mobility", "impedance", frequency=80.0)), 1))     # 500.0  N.s/m
print(f"{abs(vibration.convert_frf(Y, 'mobility', 'accelerance', frequency=80.0)):.3f}")    # 1.005  1/kg
```

The choice between the three motion FRFs is one of convenience, not physics:
they carry the same information and `convert_frf` moves between them exactly.
Accelerance is what an accelerometer-based measurement delivers directly;
mobility is the natural currency of the structure-borne power standards
(power is force times velocity, so
$P = \tfrac{1}{2}\operatorname{Re}\{Y\}\,|F|^2$ at a contact); the
reciprocals appear whenever a source is described by what it imposes rather
than by how it responds. A driving-point mobility is read as three regimes
(a stiffness line, a peak and a mass line), which section 4 shows on the plot.

### Why two mobilities decide the power

That $P = \tfrac{1}{2}\operatorname{Re}\{Y\}|F|^2$ is only half the story,
because the force at a contact is not given: it is what the two structures
agree on. Couple a source of free velocity $v_\mathrm{f}$ and mobility
$Y_\text{source}$ to a receiver of mobility $Y_\text{receiver}$ and the contact
force is

$$
F = \frac{v_\mathrm{f}}{Y_\text{source} + Y_\text{receiver}},
$$

so the power that crosses depends on **both** mobilities through their sum. Two
limits fall out, and every downstream standard is built on one of them.

- When $|Y_\text{source}| \gg |Y_\text{receiver}|$ the source behaves as a
  **velocity source**: it delivers its free velocity whatever the receiver
  does, and the receiver's mobility sets the force.
- When $|Y_\text{source}| \ll |Y_\text{receiver}|$ it behaves as a **force
  source**, delivering its **blocked force**. That is the idealisation
  ISO 10846 and EN 15657 are written around, and it is why
  [transfer stiffness](https://docs.jmrp.io/phonometry/vibration/structural/transfer-stiffness/)
  can characterise an isolator by a blocking force at all.
- Maximum power transfer sits between them, at matched mobilities. Which is
  the useful way to read a resilient mount: it works by making the source side
  *more* mobile than the receiver, not by absorbing anything.

Everything the rest of this page measures exists to put a number on one of
those two mobilities.

## 2. The SDOF reference resonator (closed form)

The canonical closed-form reference, expressed in the Table 1 / 3.1.2 FRF
taxonomy, is a mass $m$, viscous damping $c$ and stiffness $k$, whose
receptance is

$$
H(\omega) = \frac{1}{k - \omega^2 m + j\,\omega c}, \qquad
\omega_0 = \sqrt{k/m}.
$$

At the resonance $\omega_0$ the driving-point mobility is **purely real** and
equal to $1/c$ (the mobility peak measures the damping) while the static
receptance ($\omega \to 0$) is the compliance $1/k$. "Purely real" is a
statement about phase, and it generalises: the phase of a *driving-point*
mobility runs from $+90°$ on the stiffness line through $0°$ at each resonance
to $-90°$ on the mass line and never leaves that band, whatever the structure
(ISO 7626-2, A.4). A measured driving-point FRF whose phase leaves $\pm 90°$ is
not a driving-point FRF: the accelerometer and the force transducer are not at
the same point, or the attachment is compliant. A transfer FRF, by contrast,
may sit in any quadrant. The same clause adds a second free check: a
driving-point FRF shows an antiresonance between every pair of resonances, and
a missing one indicates an offset between the two transducers.

```python
import numpy as np
from phonometry import vibration

m, k, c = 2.0, 8000.0, 5.0
f0 = vibration.resonance_frequency(m, k)             # 10.07 Hz

y0 = complex(vibration.sdof_mobility(f0, m, k, c))
print(round(y0.real, 4), round(y0.imag, 6))   # 0.2 0.0   -> |Y(f0)| = 1/c
print(round(complex(vibration.sdof_receptance(1e-6, m, k, c)).real, 7))  # 0.000125 = 1/k
```

## 3. Measured FRFs and their acceptance criteria (ISO 7626-2)

In the usual ISO 7626-2 arrangement the structure hangs on a compliant
suspension (the standard admits freely suspended or grounded structures; clause
5.2 asks a grounded support to be representative of the intended application),
an exciter drives one point through an **impedance head** (a transducer stack
measuring force and acceleration at the same point, which is what gives the
attached-exciter setup its driving-point FRF), and accelerometers pick up
the response
elsewhere for the transfer FRFs. ISO 7626-5 covers the alternative of impact
excitation with an exciter that is not attached to the structure, in
practice usually an instrumented hammer: it trades the attached exciter's
controlled spectrum for speed, with an excitation spectrum set by the
impactor mass and tip stiffness, and it has [a page of its
own](https://docs.jmrp.io/phonometry/vibration/structural/impact-mobility/).

![ISO 7626 mobility measurement: a free-free beam on soft suspension driven by an exciter through a drive rod and an impedance head at the driving point, an accelerometer at a transfer point, and an impact hammer as the alternative excitation. Below, a three-panel inset shows the accelerometer routed through the drive rod marked invalid, the force transducer at the structure marked valid, and the force transducer at the exciter end marked with caution, with the ten-times mobility criteria for the exciter attachment and the suspension stated underneath.](https://docs.jmrp.io/phonometry/media/diagram_mobility_rig.svg)

### What the suspension has to do (clause 5.3)

"Soft enough" is not the criterion; ISO 7626-2 clause 5.3 gives three numbers,
and the first of them is a *mobility* criterion, which is checkable with the
functions this page has already introduced.

- **The driving-point mobility of the suspension** at each attachment point
  should be at least **ten times** that of the structure at the same point,
  over the whole frequency range of interest. This is the requirement that
  matters, because it is what keeps the suspension out of the answer.
- As a **minimum** requirement, every rigid-body resonance of the suspended
  structure shall be **below half the lowest frequency of interest**: the
  lowest frequency of *interest*, not the first elastic mode.
- The masses of suspension components near the structure (hooks, turnbuckles)
  shall be **less than one tenth** of the free effective mass of the structure
  at each frequency of interest.

And four practical rules that go with them: attach near nodal points, found by
preliminary testing, so the suspension interacts as little as possible; run the
cables normal to the excitation direction where practical, and expect transverse
string vibrations to affect the data even then; keep the suspension resonances
well away from the structure's modal frequencies, because shock cords and foam
pads carry mass but little damping; and watch the damping the suspension itself
adds, which the clause raises as a CAUTION. The suspension used must be
described in the test report.

The 10× rule is worth executing rather than quoting. For the resonator of
section 2 the worst case is its own peak, where the structure is at its most
mobile:

```python
# `vibration` is imported by the figure block above; the blocks of a page
# run in reading order.
f0 = vibration.resonance_frequency(2.0, 8000.0)                   # 10.07 Hz
y_structure = abs(complex(vibration.sdof_mobility(f0, 2.0, 8000.0, 5.0)))
print(round(y_structure, 3))                                      # 0.2 = 1/c

# The suspension seen at the same point, disconnected from the structure:
# a shock cord of 31.6 N/m, whose own mobility is omega / k.
y_suspension = abs(complex(
    vibration.convert_frf(31.6, "dynamic_stiffness", "mobility", frequency=f0)))
print(round(y_suspension / y_structure, 1))       # 10.0 -> clause 5.3 met
```

That cord puts the rigid-body bounce of the 2 kg structure at 0.63 Hz, so the
minimum requirement is met with room to spare for any lowest frequency of
interest above 1.3 Hz. The mobility rule is the binding one, and it is the same
criterion, in the same units, as the exciter-attachment rule below: **everything
you attach to the structure must be at least ten times more mobile than the
structure itself**.

### Attaching the exciter (clause 6.4.4)

ISO 7626-2 calls exciter attachment "often the most difficult problem
encountered when using fixed vibration exciters to measure the mobility of
lightweight structures", and gives it a criterion and a design.

- **The restraint criterion.** With the exciter and its hardware *disconnected*
  from the structure, the lateral and rotational driving-point mobilities of
  the attachment shall be at least ten times larger, at all frequencies of
  interest, than the corresponding elements of the structure's driving-point
  mobility matrix. An exciter that fails it clamps the structure laterally and
  rotationally, and the low-order modes are the first casualty.
- **The drive rod.** High stiffness along the excitation axis and enough
  flexibility in every other direction: slender short rods are usual, but thick
  rods with thin flexible sections near each end can do better. Align the
  exciter and the rod with the force-transducer axis. Two things to watch:
  drive-rod bending modes inside the frequency range can interfere with the
  measurement, and bending of the exciter's own moving system can inject
  moments the force transducer never sees.
- **Where the two transducers go**, which is where validity is decided and
  which the inset of the diagram above draws. With a flexible drive rod the
  accelerometer shall be attached **directly to the structure**, never through
  the rod, whose axial compliance makes the response measurement invalid. The
  force transducer shall measure the force transmitted **from the rod into the
  structure**; putting it at the exciter end is admitted "only with extreme
  caution", and then the rod compliance must be checked per ISO 7626-1 and the
  rod mass compensated by the mass-cancellation procedure below.
- **Area-reducing cones** may be needed to approximate a point force, and can
  themselves introduce a spurious moment if used carelessly.

### What the transducers do to the answer (clauses 6.4.2, 6.4.3 and 7.3)

Every transducer you attach adds mass, and the spurious force needed to
accelerate that mass is measured as though the structure had produced it. The
symptom is the damaging part: uncompensated inertia loading **shifts the
frequencies of the response peaks**, and a resonance measured in the wrong place
is inherited by everything downstream: the modal fit, the damping from the
half-power bandwidth, the $|Y| = 1/c$ read of section 2. Rotational inertia does
the same to moments (6.4.3), and impedance heads are the worst offenders because
their moment of inertia about the mounting point can be large.

The standard's sequence is: choose the smallest mass and lowest inertia
consistent with sensitivity; then, if that is not enough, compensate.

- **When to act.** Electronic compensation shall be considered when the
  magnitude of the structure's driving-point mobility exceeds
  $0.01/(f\,m_\mathrm{t})$ at all frequencies of interest, with $m_\mathrm{t}$ the total
  effective mass of the attachment hardware plus the effective end mass of the
  force transducer or impedance head (and of a separate driving-point
  accelerometer, if one is fitted).
- **What it is.** "Mass cancellation": multiply the driving-point acceleration
  by $m_\mathrm{t}$ and subtract the product from the force signal, in analogue or
  digitally. A *transfer*-mobility measurement therefore needs a separate
  accelerometer at the driving point to supply that signal.
- **What it cannot do.** It compensates translational inertia at the driving
  point, in the excitation direction, only: never rotational inertia. The
  standard **strongly recommends** reconsidering the transducer selection and
  redesigning the attachment hardware ahead of using it, and confines it to the
  range where the ratio of attachment-and-transducer effective mass to the
  structure's free effective mass at the driving point is between **0.05 and
  0.5**.
- Clause 7.2 adds the rule that makes any of this repeatable: transducers
  mounted on studs are tightened to the torque the manufacturer recommends, and
  attachment compliance is checked per ISO 7626-1.

For the resonator of section 2 the trigger is easy to place. Its driving-point
mobility peaks at $1/c = 0.2$ m/(N·s) at 10.07 Hz; with a 20 g impedance head
and attachment hardware ($m_\mathrm{t} = 0.02$ kg) the trigger $0.01/(f\,m_\mathrm{t})$ is
0.050 m/(N·s) there, so at its peak the structure is four times above it. Take
a 2 g head instead and the trigger rises to 0.50 m/(N·s), above the peak. The
trigger falls as the added mass rises, which is the whole point of it: the same
structure needs compensation with a heavy transducer and not with a light one.

Processing measured random-excitation records per ISO 7626-2, 8.1.3 (the H1
estimator $\hat{H} = G(\text{response}, \text{force})/G(\text{force}, \text{force})$)
and the ordinary coherence $\gamma^2 = |G_{xy}|^2/(G_{xx}\,G_{yy})$ used for
its data-quality checks are the library's
existing spectral estimators [`transfer_function` and
`coherence`](https://docs.jmrp.io/phonometry/devices/electroacoustics/electroacoustics/) (H1 is their default).

### Reading the coherence, and the one bias H1 has (Annex A.1)

Annex A.1 is a diagnosis table, and it is the most immediately useful page of
the standard. A coherence substantially below 1 says something specific
depending on *where* it drops.

- **Notches at the resonances and antiresonances** mean inadequate frequency
  resolution (which also biases the computed FRF), inadequate time-domain
  weighting, a nonlinear structure or a saturating amplifier, more than one
  input force, or digitising and electronic noise from an inadequate exciter
  force.
- **Notches at the resonances specifically** can also mean the exciter's force
  drops there, which is inherent when a shaker drives a lightly damped
  structure. That leads to the bias worth knowing before recommending H1: if
  the force signal falls below the noise floor, the H1 estimate (the
  cross-spectrum over the *force* auto-spectrum) reads **too low**, exactly at
  the resonance peak the reader is being taught to measure the damping from.
  H2 (response auto-spectrum in the denominator) has the opposite bias, being
  the one to reach for when the noise is on the force; Hv sits between them.
  `transfer_function` provides H1 by default.
- **A broad droop across the range** is poor signal-to-noise, usually
  inadequate dynamic range: shape the excitation spectrum, or use synchronous
  time-domain averaging.
- **Low coherence anywhere** can simply be nonlinearity.

Two notes catch beginners. A coherence computed from a *single* record is
always a fictitious 1.0, so it means nothing until several averages are in. And
a high coherence does not prove the data valid: it can equally indicate
cross-talk between the force and response channels.

### The acceptance criteria

On top of the estimators, two ISO 7626-2 acceptance criteria are provided:

* **Operational rigid-mass calibration (7.5.2).** The measured FRF of a freely
  suspended rigid block of known mass must agree within ±5 % with
  $|A| = 1/m$ (accelerance) or $|Y| = 1/(2\pi f m)$ (mobility).
* **Random error (Annex A + 8.1.3).** Enough spectra must be averaged that the
  normalized random error $\varepsilon = \sqrt{(1-\gamma^2)/(2n\gamma^2)}$ at
  each resonance of a
  driving-point mobility is below 5 %.

```python
import numpy as np
from phonometry import vibration

# A 10 kg calibration block: |A| must be 1/m = 0.100 1/kg at every frequency.
f = np.array([20.0, 100.0, 500.0])
res = vibration.verify_rigid_mass_calibration([0.100, 0.102, 0.097], frequencies=f, mass=10.0)
print(res.passes, res.within_tolerance.tolist())   # True [True, True, True]

# The Annex A example: coherence 0.8 needs about 75 averages for < 5 %.
print(round(float(vibration.random_error_percent(0.8, 75)), 2))   # 4.08  %
```

The random-error criterion is a decision taken *before* a measurement, not a
check after one, and how many averages it costs depends steeply on the
coherence you expect:

![Normalized random error against the number of averaged spectra on logarithmic axes, one straight falling line per coherence value from 0.5 to 0.95. A horizontal dashed line marks the five per cent criterion, and the number of averages each coherence needs to reach it is marked where its curve crosses: eleven averages at a coherence of 0.95, twenty-two at 0.9, fifty at 0.8, eighty-six at 0.7 and two hundred at 0.5. A star marks the standard's own worked example, a coherence of 0.8 with seventy-five averages giving 4.08 per cent.](https://docs.jmrp.io/phonometry/media/mobility_random_error.svg)

*What the criterion costs, as a function of how good the measurement already
is. Reaching 5 % takes 11 averages at $\gamma^2 = 0.95$ and 200 at
$\gamma^2 = 0.5$: a factor of twenty for a factor of two in coherence. That is
the argument for fixing the measurement, with the Annex A.1 diagnosis above,
rather than averaging through it.*

**Show the code for this figure**

```python
import matplotlib.pyplot as plt
import numpy as np

averages = np.unique(np.round(np.logspace(np.log10(2.0), np.log10(1000.0), 260)))
for coherence in (0.5, 0.7, 0.8, 0.9, 0.95):
    error = [float(vibration.random_error_percent(coherence, int(n)))
             for n in averages]
    plt.loglog(averages, error, label=f"coherence {coherence}")
    # The n at which this coherence reaches the 5 % criterion.
    print(coherence, round((1.0 - coherence) / (2.0 * coherence * 0.05**2)))
plt.axhline(5.0, ls="--", color="0.4")
plt.xlabel("Number of averaged spectra"); plt.ylabel("Random error [%]")
plt.legend(); plt.show()
```

### Three more tests the standard asks for, and this page does not run

Clause 10 is explicit that the Annex A.1 averaging procedure applies to random
excitation, and that "additional tests, applicable to **all** types of
excitation, are described in A.2 to A.4". Those three are not implemented here,
but they are cheap, and a measurement that has passed only the two criteria
above is not yet validated.

- **A.2, linearity.** Bolted joints, support clearances and similar features
  make practical structures nonlinear. Measure the FRF, then repeat it with the
  excitation amplitude significantly increased and significantly decreased; if
  the results disagree, the reason needs further analysis. The clause asks for
  this in the course of *each* test series, and it is the check that catches
  what a coherence notch only hints at.
- **A.3, reciprocity.** For a linear elastic structure $Y_{ij}(f) = Y_{ji}(f)$:
  swap the exciter and the response accelerometer and the transfer mobility must
  come back the same. It is a strong end-to-end test of the whole chain,
  including the attachment restraints above, and it costs one extra
  measurement. (Some elements, hydrodynamic bearings among them, are genuinely
  non-reciprocal; between such an element and the drive point the test does not
  apply.)
- **A.4, driving point against transfer.** The phase and antiresonance checks
  of section 2: a driving-point FRF stays inside $\pm 90°$ and shows an
  antiresonance between every pair of resonances.

The distinction is worth keeping: A.1 is a **precision** criterion, answering
how many averages a given random error costs, while A.2 to A.4 are **validity**
criteria that no amount of averaging will fix.

The calibration check returns a `RigidMassCalibrationResult` carrying the
per-frequency deviation and pass flags, and a `.plot()`: the measured FRF
magnitude against the rigid-mass line with its ±5 % tolerance band (upper
panel) and the relative deviation against the same band (lower panel, where a
few-percent tolerance is actually readable). A calibration that drifts out of
the band towards a few kHz points at a transducer or attachment-compliance
error, exactly what the check is meant to catch:

```python
import numpy as np
from phonometry import vibration

m = 10.0                                                  # calibration block mass
f = np.logspace(np.log10(20.0), np.log10(5000.0), 400)
drift = 0.05 * (f / 2500.0) ** 2                          # high-frequency drift
measured = (1.0 / m) * (1.0 + 0.015 * np.sin(2 * np.pi * np.log10(f)) + drift)
res = vibration.verify_rigid_mass_calibration(measured, frequencies=f, mass=m)
print(res.passes)                                         # False (drift exceeds 5 %)
res.plot()
```

![Rigid-mass calibration check of a 10 kg block: the measured accelerance magnitude follows the flat rigid-mass line inside the plus-or-minus five percent tolerance band across most of the range, then drifts above the band towards a few kilohertz where the out-of-tolerance points are marked, and the lower panel shows the same deviation in percent crossing the plus five percent limit](https://docs.jmrp.io/phonometry/media/rigid_mass_calibration.svg)

**Show the code for this figure**

```python
import matplotlib.pyplot as plt
import numpy as np
from phonometry import vibration

m = 10.0
f = np.logspace(np.log10(20.0), np.log10(5000.0), 400)
drift = 0.05 * (f / 2500.0) ** 2
measured = (1.0 / m) * (1.0 + 0.015 * np.sin(2 * np.pi * np.log10(f)) + drift)
res = vibration.verify_rigid_mass_calibration(measured, frequencies=f, mass=m)
bad = ~res.within_tolerance

fig, (top, bot) = plt.subplots(2, 1, sharex=True, figsize=(10, 7),
                               gridspec_kw={"height_ratios": [1.5, 1.0]})
top.fill_between(f, res.expected * 0.95, res.expected * 1.05, color="C1",
                 alpha=0.15, label="±5 % tolerance band")
top.semilogx(f, res.expected, "--", color="C1", label="expected |A| = 1/m")
top.semilogx(f, res.measured, color="C0", label="within tolerance")
top.semilogx(f[bad], res.measured[bad], "o", color="C1", label="out of tolerance")
top.set_ylabel("Accelerance |A| [1/kg]"); top.legend()

bot.axhspan(-5.0, 5.0, color="C1", alpha=0.15)
bot.semilogx(f, 100.0 * res.relative_deviation, color="C0")
bot.semilogx(f[bad], 100.0 * res.relative_deviation[bad], "o", color="C1")
bot.set_xlabel("Frequency [Hz]"); bot.set_ylabel("Deviation [%]")
plt.show()
```

## 4. The `MobilityResult` bundle

`sdof_mobility_result` bundles the FRF over frequency into a `MobilityResult`,
which exposes `.magnitude`, `.phase_rad`, `.to(target)` (any Table-1 kind) and a
`.plot()` of $|Y(f)|$ with the resonance marked:

```python
import numpy as np
from phonometry import vibration

f = np.logspace(np.log10(0.5), np.log10(200.0), 400)
res = vibration.sdof_mobility_result(f, mass=2.0, stiffness=8000.0, damping=5.0)
z = res.to("impedance")                        # impedance = 1/Y per frequency
print(res.frequencies[int(np.argmax(res.magnitude))].round(1))   # ~10.1 Hz

res.plot()   # |Y(f)| with the resonance marked (needs matplotlib)
```

![Driving-point mobility magnitude of a single-degree-of-freedom resonator on log-log axes, climbing along the stiffness line below resonance, falling along the mass line above it, and peaking at one over the damping coefficient at the resonance](https://docs.jmrp.io/phonometry/media/mobility_result_lines.svg)

*Reading a driving-point mobility is a structural diagnosis. Below the
resonance the magnitude climbs along the **stiffness line** $\omega/k$, above
it it falls along the **mass line** $1/(\omega m)$, and the height of the peak
between them is $1/c$: a direct read of the damping, for this isolated
resonator; on a real structure with overlapping modes the damping is estimated
by modal fitting or from the half-power bandwidth instead. The lower panel is
the same statement in phase: $+90°$ on the stiffness line, $-90°$ on the mass
line, and exactly $0°$ at the resonance, which is what "purely real" means and
why the peak equals $1/c$.*

**Show the code for this figure**

```python
import matplotlib.pyplot as plt
import numpy as np
from phonometry import vibration

m, k, c = 2.0, 8000.0, 5.0
f = np.logspace(np.log10(0.5), np.log10(200.0), 400)
res = vibration.sdof_mobility_result(f, mass=m, stiffness=k, damping=c)

# One line — |Y(f)| with the resonance marked:
res.plot()
plt.show()

# By hand, adding the stiffness and mass asymptotes the prose describes:
w = 2.0 * np.pi * f
fig, ax = plt.subplots()
ax.loglog(f, res.magnitude, label="driving-point |Y(f)|")
ax.loglog(f, w / k, ":", label="stiffness line ω/k")
ax.loglog(f, 1.0 / (w * m), ":", label="mass line 1/(ωm)")
ax.axhline(1.0 / c, ls="--", color="0.6", label="peak |Y| = 1/c")
ax.set_xlabel("Frequency [Hz]")
ax.set_ylabel("Mobility |Y| [m/(N·s)]")
ax.set_title("Reading a driving-point mobility (ISO 7626-1)")
ax.legend()
plt.show()
```

## 5. Reference mobilities when nothing has been measured

The SDOF is a reference for the *definitions*, not a model of anything a
building acoustician owns. For a large plate or a long beam, driven away from
its edges, the reflected waves that make a real structure resonate have decayed
before they return, and the driving-point mobility takes the closed form of the
equivalent **infinite** structure, which is exactly the stand-in EN 12354-5
wants for a receiver mobility when no measurement exists. Three characters,
from Cremer, Heckl & Petersson Table 5.1:

- the **infinite plate** is a pure real resistance, $Y = 1/(8\sqrt{B'm''})$,
  independent of frequency: it absorbs at every frequency alike and has no peak
  to read a damping from;
- the **infinite beam** is $Y = (1-\mathrm{j})/(4m'c_\mathrm{B})$, so $-45°$ of phase at
  every frequency and a magnitude falling as $\omega^{-1/2}$, because the
  bending wave speed $c_\mathrm{B}$ itself grows as $\sqrt{\omega}$;
- a **rod in longitudinal motion** is $Y = 1/(\rho c_\mathrm{L} S)$, real and flat
  again: the mechanical analogue of a matched transmission line.

```python
import numpy as np

f = np.array([50.0, 500.0])

# A 140 mm concrete slab: E = 30 GPa, nu = 0.2, rho = 2400 kg/m3.
b_plate = vibration.plate_bending_stiffness(3.0e10, 0.14, 0.2)
plate = vibration.infinite_plate_point_mobility(f, b_plate, 336.0)
print(f"{float(plate.magnitude[0]):.3e}", round(float(np.degrees(plate.phase_rad[0]))))
#  2.551e-06 0     -> real, and the same at 500 Hz

# A 100 x 200 mm steel beam, driven at a point away from its ends.
b_beam = 2.1e11 * 0.1 * 0.2**3 / 12.0
beam = vibration.infinite_beam_point_mobility(f, b_beam, 7800.0 * 0.02)
print(f"{float(beam.magnitude[0]):.3e}", round(float(np.degrees(beam.phase_rad[0]))))
#  7.388e-06 -45
print(np.round(vibration.beam_bending_wave_speed(f, b_beam, 7800.0 * 0.02), 1))
#  [306.8 970.1]  -> c_B grows as sqrt(f), so |Y| falls as 1/sqrt(f)
```

Both constructors return a `MobilityResult`, so they plot, convert with `.to()`
and report exactly like the SDOF above. Two more cover the degrees of freedom
the pair does not: `infinite_beam_moment_mobility` is the rotational mobility a
junction line or a bracket bolted at two points needs, in rad/(N·m·s), for when
the excitation is a moment rather than a force; and `longitudinal_rod_mobility`
(with its `longitudinal_rod_impedance` reciprocal) covers a strut or a pipe run
loaded along its axis.

![Two panels sharing a logarithmic frequency axis. The upper panel is the driving-point mobility magnitude of four structures: the single-degree-of-freedom resonator with its sharp peak drawn dashed, a horizontal line for the infinite concrete plate, a straight falling line of slope minus one half for the infinite steel beam, and a second horizontal line for the steel strut in longitudinal motion. The lower panel is the phase of each: the resonator sweeps from plus ninety degrees through zero at its resonance to minus ninety, while the plate and the rod stay at zero and the beam stays at minus forty-five degrees at every frequency.](https://docs.jmrp.io/phonometry/media/infinite_mobilities.svg)

*The contrast the whole section is for. An infinite structure never reflects
anything back, so it has no resonance and no peak: the plate and the rod are
real and flat, the beam falls as $f^{-1/2}$ at a constant $-45°$. Reading a
mobility plot as "stiffness line, peak, mass line" is a statement about finite
structures; a large plate is the case where none of the three appears.*

```python
# One line each, on shared axes with the SDOF of section 4.
plate.plot()
beam.plot()
```

Where do these numbers go? Into
[panel sound insulation](https://docs.jmrp.io/phonometry/buildings/design/panel-sound-insulation/),
which uses the same functions to price a panel's radiation, and into the
EN 12354-5 receiver mobility of
[installed structure-borne sound](https://docs.jmrp.io/phonometry/buildings/design/installed-structure-borne/).

### Test-report fiche

`MobilityResult.report(path)` renders a one-page mechanical-mobility measurement
report (ISO 7626-1:2011 FRF definitions, measurement per ISO 7626-2:2015).
Mobility is a continuous frequency-response function, not an octave-band
quantity, so the sheet presents it honestly as the $|Y(f)|$ magnitude spectrum
plus a compact table of characteristic points (the FRF type, driving-point or
transfer, the frequency range, the peak frequency, the peak mobility magnitude
and the phase there), and a boxed peak mobility $|Y|$ at the frequency it occurs
at (for a driving-point FRF a resonance, where $|Y| = 1/c$ measures the
damping). It is a characterisation, so there is no pass/fail verdict;
`language="es"` renders the Spanish fiche. The fiche always embeds the $|Y(f)|$
spectrum, so it needs both the report and plot extras
(`pip install "phonometry[report,plot]"`).

```python
from phonometry import ReportMetadata, vibration

res = vibration.sdof_mobility_result(f, mass=2.0, stiffness=8000.0, damping=5.0)
res.report(
    "mobility.pdf",
    metadata=ReportMetadata(
        specimen="Machine support bracket (driving point)",
        measurement_standard="ISO 7626-2",
    ),
)   # one-page fiche (needs phonometry[report,plot])
```

**Report:** [ISO 7626 mechanical-mobility example report (PDF)](https://docs.jmrp.io/phonometry/media/iso7626_mobility_example.pdf): One-page mechanical-mobility fiche: a metadata header, a table of the FRF characteristic points (the FRF type, the frequency range, the peak frequency, the peak mobility magnitude and the phase there) beside the mobility magnitude spectrum, and the boxed peak mobility.

Mechanical-mobility fiche (MobilityResult.report): the FRF characteristic points and the mobility magnitude spectrum.

## What this guide covers

- **Covered.** ISO 7626-1:2011's FRF family (Table 1): receptance, mobility and accelerance,
  with their force-per-motion reciprocals, moved between through `convert_frf`.
  Also covered are the driving-point/transfer distinction and the closed-form
  SDOF resonator used as their reference (`sdof_receptance`, `sdof_mobility`,
  `sdof_accelerance`, `resonance_frequency`, `sdof_mobility_result`), and the
  infinite-structure point mobilities that stand in for a real plate, beam or
  strut (`infinite_plate_point_mobility`, `infinite_beam_point_mobility`,
  `infinite_beam_moment_mobility`, `longitudinal_rod_mobility` and their
  impedance reciprocals, with `plate_bending_stiffness` and
  `beam_bending_wave_speed`). On the measurement side, ISO 7626-2:2015's H1
  processing of random excitation is covered too (`transfer_function`,
  `coherence`, shared with the electroacoustics guide), along with two
  acceptance criteria: the 7.5.2 rigid-mass operational calibration
  (`verify_rigid_mass_calibration`) and the Annex A random-error criterion
  (`random_error_percent`).

- **Not covered.** Impact-hammer excitation (ISO 7626-5) is not processed on this page but on
  [its own](https://docs.jmrp.io/phonometry/vibration/structural/impact-mobility/). The *blocked*
  matrix quantities of Table 1, needed for multi-coordinate systems, are not
  built.
  `convert_frf` returns only the element-wise *free* reciprocals of ISO 7626-1,
  3.1.4, correct for driving-point or single-path use but not for a full FRF
  matrix. Of the acceptance procedure, only the two criteria above are computed:
  the clause 5.3 suspension rules, the clause 6.4.4 exciter-attachment
  criterion, the clause 7.3 mass-cancellation trigger and the A.2 to A.4
  linearity, reciprocity and driving-point tests are described here and left to
  the operator, in the same way the linearity test of ISO 10846 (clause 7.7 of
  Parts 2, 4 and 5, clause 7.6 of Part 3) is described but not computed on the [transfer-stiffness
  page](https://docs.jmrp.io/phonometry/vibration/structural/transfer-stiffness/).

## See also

- [Mobility by impact excitation (ISO 7626-5)](https://docs.jmrp.io/phonometry/vibration/structural/impact-mobility/):
  the same FRFs measured with an impact hammer, with the windows, the average
  over impacts and the Annex A correction that measurement needs.
- [Transfer stiffness of resilient elements (ISO 10846)](https://docs.jmrp.io/phonometry/vibration/structural/transfer-stiffness/):
  the blocked-force limit of section 1, applied to characterise an isolator.
- [Structure-borne sound power of equipment (EN 15657)](https://docs.jmrp.io/phonometry/buildings/design/structure-borne-power/):
  where a source's mobility and free velocity become an installed power.
- [Installed structure-borne sound (EN 12354-5)](https://docs.jmrp.io/phonometry/buildings/design/installed-structure-borne/):
  the prediction that consumes a receiver mobility, measured or taken from
  section 5.
- [Bending-wave transmission at plate junctions](https://docs.jmrp.io/phonometry/vibration/structural/junction-transmission/):
  what the plate mobilities of section 5 become at a junction between two of
  them.
- [Frequency response and coherence](https://docs.jmrp.io/phonometry/devices/electroacoustics/electroacoustics/):
  the `transfer_function` and `coherence` estimators section 3 runs on.
- API reference: [`vibration.structural.mechanical_mobility`](https://docs.jmrp.io/phonometry/reference/api/vibration/mechanical-mobility/).
- Theory: [Point mobilities and radiation efficiency](https://docs.jmrp.io/phonometry/reference/theory/vibration/#point-mobilities-and-radiation-efficiency-cremer-5-hopkins-29): the infinite-structure point mobilities and the radiation efficiency, and why they are averages a finite structure oscillates about.

## References

- **Cremer, L., Heckl, M., & Petersson, B. A. T. (2005).** *Structure-borne sound: Structural vibrations and sound radiation at audio frequencies* (3rd ed.). Springer. https://doi.org/10.1007/b137728
  The standard monograph on structural vibration: point and transfer mobilities of beams and plates, and the power flow P = ½·Re{Y}·\|F\|² that makes mobility the working quantity of this page. ISBN 978-3-540-22696-3.
- **International Organization for Standardization. (2011).** [*Mechanical vibration and shock — Experimental determination of mechanical mobility — Part 1: Basic terms and definitions, and transducer specifications*](https://www.iso.org/standard/50426.html) (ISO 7626-1:2011).
  The FRF family and its reciprocals (Table 1, the 3.1.2 mobility and 3.1.4 free-quantity definitions), the free/blocked distinctions and the driving-point / transfer distinction implemented here. Conformance is anchored on the closed-form SDOF identities consistent with those definitions: the driving-point mobility peak \|Y(ω0)\| = 1/c, the static receptance H(0) = 1/k and the exact Table-1 reciprocity impedance·mobility = 1.
- **International Organization for Standardization. (2015).** [*Mechanical vibration and shock — Experimental determination of mechanical mobility — Part 2: Measurements using single-point translation excitation with an attached vibration exciter*](https://www.iso.org/standard/62483.html) (ISO 7626-2:2015).
  The measurement side: the 8.1.3 H1 processing of random excitation, the 7.5.2 rigid-mass operational calibration (±5 %) and the Annex A random-error criterion (< 5 % at resonances). Conformance is anchored on the rigid-mass calibration values (\|A\| = 0.100 1/kg for 10 kg; \|Y\| = 1.59155e-4 m/(N·s) at 100 Hz) and the Annex A example (γ² = 0.8, n = 75 → ε = 4.08 %).