# Build a sound level meter

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/signals/sound-level-meter/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/signals/sound-level-meter/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-07-21
Updated: 2026-11-08T02:07:59Z
Standards: IEC 61672-1:2013; IEC 61260-1:2014; IEC 60942:2017
Verified: 87 conformance checks (IEC 61672-1:2013; IEC 61260-1:2014; IEC 60942:2017) https://docs.jmrp.io/phonometry/reference/conformance/
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Sound level meter built from the API: IEC 60942 calibration, IEC 61672-1 weightings, Leq, SEL, percentiles, IEC 61260-1 bands and class checks.

---

A sound level meter is not one algorithm but a short pipeline of them, and
IEC 61672-1 specifies every stage. phonometry implements each stage as an
independent, composable function; this page assembles them, in order, into a
working meter. Every snippet runs as written (the signals are synthesized so
the page is self-contained), and each stage links to the deep guide that
explains it fully.

![The sound level meter pipeline of this page: the calibrator tone feeding sensitivity() to obtain the factor in pascals per digital unit, the measurement recording joining it as the calibrated pressure, and three readout branches from there, weighting_filter with time_weighting for the displayed LAF(t) and the percentile levels, laeq, sel and lc_peak for the integrated levels, and octave_filter for the one-third-octave spectrum, all closed by verify_weighting_class and verify_filter_class against the IEC 61672-1 Table 3 and IEC 61260-1 Table 1 acceptance limits](https://docs.jmrp.io/phonometry/media/diagram_slm_pipeline.svg)

This is the same chain IEC 61672-1 draws for the physical instrument: the
class 1 calibrator anchors the microphone to 94 dB at 1 kHz, and every stage
that follows is one function of this page.

![IEC 61672-1 sound level meter chain: a class 1 sound calibrator at 94 dB and 1 kHz coupled onto the measurement microphone with its windscreen, followed by the four instrument stages of microphone plus preamplifier, A, C or Z frequency weighting, squaring with the exponential F or S time weighting of 125 ms or 1 s, and the display of LAF(t) in dB re 20 µPa](https://docs.jmrp.io/phonometry/media/diagram_slm_chain.svg)

The snippets on this page build on each other: run them top to bottom in one
session (or paste the whole page into a script).

## 1. The scenario

A meter needs two recordings from the *same* input chain: the calibrator tone
that anchors the digital numbers to pascals, and the measurement itself. Here
both are synthesized so you can run the page anywhere; in a real measurement
they come from your microphone.

```python
import numpy as np
from phonometry import filters, metrology, signals

fs = 48000

# Calibrator tone: 94 dB SPL = 1 Pa RMS at 1 kHz (IEC 60942).
#   Synthesized here; in the field, record a few seconds of your calibrator.
calibrator = np.sqrt(2) * np.sin(2 * np.pi * 1000 * np.arange(3 * fs) / fs)

# "Street" measurement: 10 s of pink background noise plus a 1 s horn-like
#   1 kHz event, so the statistical levels have something to separate.
recording = signals.noise_signal(fs, duration_s=10.0, color="pink", rms=0.02, seed=7)
recording[4 * fs : 5 * fs] += 0.2 * np.sqrt(2) * np.sin(
    2 * np.pi * 1000 * np.arange(fs) / fs
)
```

## 1b. Recording the two files

A sound level meter built in software inherits every defect of the file it is
handed, and three recorder settings decide whether the numbers below mean
anything. **Set the input gain by hand and lock it.** Automatic gain control,
limiting and noise suppression all move the sensitivity the next step measures
once and step 4 then applies to every sample, so any of them silently voids the
calibration. **Record uncompressed PCM, 24 bit, at 48 kHz or more.** A lossy
codec redistributes the high-frequency energy the one-third-octave stage of
step 5 reports, and a 44.1 kHz file has no band left above 22 kHz in which to
support the class 1 A-weighting of step 3 near 16 kHz. **Choose the gain from
the loudest event the measurement must be able to prove**, leaving roughly
12 dB of headroom above its peak: the $L_\mathrm{Cpeak}$ of step 4 reads the true
inter-sample peak, and one clipped run invalidates it and the $L_\mathrm{Aeq}$ with
it.

Two more decisions belong to the session rather than to the recorder. Record
the calibrator tone through that same locked chain immediately before *and*
immediately after the measurement, so the pair brackets the drift that step 7
checks; a factor measured on another day, or after a gain control moved, is not
a calibration of this file. Outdoors, keep the windscreen fitted and point the
microphone in the reference direction its free-field calibration assumes.

A field sheet worth copying:

- Gain set by hand and written down; AGC, limiter and noise suppression off.
- Uncompressed PCM, 24 bit, $f_\mathrm{s} \ge 48$ kHz, one channel per microphone.
- Peak headroom of about 12 dB above the loudest expected event.
- Calibrator tone before and after, same gain, same chain, same session.
- Windscreen fitted, microphone on a tripod, reference direction at the source.
- Sample rate, gain setting, microphone and calibrator identifiers logged.

## 2. Calibrate: give the samples physical meaning

Digital samples are dimensionless; the **sensitivity factor** converts them
to pascals. `sensitivity()` computes it from the calibrator recording and, when
you pass `fs` as below and leave `validate` at its default `True`, also
validates the recording's short-term stability the way IEC 60942 qualifies the
calibrator itself, so a badly coupled microphone is caught here instead of
corrupting every level downstream. Drop either one and you get the factor on
its own, with no check.

```python
cal = metrology.sensitivity(calibrator, target_spl=94.0, fs=fs)
# cal is in Pa per digital unit; every level function accepts it as
# calibration_factor. For this synthetic tone it is ~1.0.
```

Deep guide: [Calibration and dBFS](https://docs.jmrp.io/phonometry/signals/metrology/calibration/), which
covers the stability check this call runs on the calibrator recording, the
integer-format rule, and the digital dBFS mode used when no physical reference
exists. With no calibrator at all the factor can be computed from the chain's
rated sensitivity rather than measured: a microphone rated $M$ mV/Pa through a
preamplifier of linear gain $G$ into a converter whose full scale is $V_\mathrm{FS}$
volts gives $S = V_\mathrm{FS} / (M G / 1000)$ pascals per digital unit; but a rated
factor carries the datasheet's tolerance rather than the ±0.4 dB of a class 1
calibrator, so it cannot be reported as a calibrated measurement.

## 3. Weight: frequency and time (IEC 61672-1)

The meter never shows raw pressure. The signal first passes the **A
frequency weighting** (the ear-response curve of IEC 61672-1), is squared,
and is then smoothed by the **Fast exponential detector** (time constant
125 ms). The result is the moving level a meter's display follows,
$L_\mathrm{AF}(t)$:

```python
pressure = cal * recording                                # digital units -> Pa
weighted = filters.weighting_filter(pressure, fs, weighting="A")
envelope = filters.time_weighting(weighted, fs, mode="fast")  # mean-square Pa^2
laf_t = 10 * np.log10(np.maximum(envelope, 1e-12) / (2e-5) ** 2)
# laf_t peaks near 80 dB during the event and settles near 55 dB between.
```

**Video:** [The Fast, Slow and Impulse detector ballistics](https://docs.jmrp.io/phonometry/media/anim_time_weighting.webm): A tone burst drives the RC exponential detector: the capacitor charges and drains while the Fast, Slow and Impulse meter needles follow their own ballistics, and the three level traces build up below to show the fast attack, the slow average and the asymmetric impulse hold.

The needle in the clip is that `time_weighting` call: a first-order low-pass
charging and draining on the squared signal. The three needles differ in one
number, the time constant, which is why Fast catches an event that Slow
smooths away. You rarely write this chain yourself: every level function of
the next step applies the frequency weighting internally, and the percentile
levels rebuild this Fast envelope for you. The energy metrics ($L_\mathrm{eq}$, SEL)
integrate the squared weighted signal directly, with no ballistics at all,
which is why they show no needle movement to follow. The chain is shown here
because it *is* the meter's display.

Deep guides: [Frequency Weighting (A, C, Z)](https://docs.jmrp.io/phonometry/signals/levels/weighting/)
and [Time Weighting](https://docs.jmrp.io/phonometry/signals/levels/time-weighting/), which takes
this same clip apart against the IEC 61672-1 tone-burst table.

## 4. Integrate: the numbers a meter reports

One pass over the calibrated recording yields the standard readouts: the
energy-equivalent **$L_\mathrm{Aeq}$**, the **percentile levels** that describe how
the level fluctuated ($L_{90}$ is the background, $L_{10}$ the events), the
**sound exposure level** that normalizes the event to one second, and the
C-weighted **peak** for impulsive content.

```python
la_eq = signals.laeq(recording, fs, calibration_factor=cal)     # ~70.1 dB
ln = signals.ln_levels(
    recording, fs, n=(10, 50, 90), weighting="A", calibration_factor=cal
)                                                # L10 ~78.0, L50 ~55.1, L90 ~54.9
lae = signals.sel(recording, fs, weighting="A", calibration_factor=cal)  # ~80.1
lc_pk = signals.lc_peak(recording, fs, calibration_factor=cal)           # ~84.4

print(f"LAeq {la_eq:.1f} dB | L10 {ln[10]:.1f} | L90 {ln[90]:.1f} "
      f"| LAE {lae:.1f} | LCpeak {lc_pk:.1f}")
```

Note the arithmetic the numbers encode. The event second measures 80.0 dB on its
own and the nine background seconds 55.0 dB, so in energy the event contributes
$1 \times 10^{8.00}$ and the background $9 \times 10^{5.50} = 2.9 \times 10^{6}$:
a ratio of about 35:1, or 15.5 dB. The event's 25 dB head start beats the
9.5 dB ($10\log_{10}9$) the background wins back on duration, which is why the
10 s $L_\mathrm{Aeq}$ of 70.1 dB sits only 0.1 dB above the 70.0 dB the event alone
would give spread over the same 10 s: the background is almost invisible in the
energy average. $L_{\mathrm{A}E}$ is that $L_\mathrm{Aeq}$ plus $10\log_{10}(10\ \mathrm{s}/1\
\mathrm{s}) = 10$ dB, the whole energy compressed into one second. The
percentiles read the other way round: the event fills exactly 10 % of the
record, so it lifts $L_{10}$ to 78.0 dB and leaves $L_{90}$ at 54.9 dB, the
background level.

![Fast A-weighted level history of the ten-second recording of this page, rising from about 55 dB of pink background to 80 dB during the one-second event at four seconds, with horizontal lines at LAeq 70.1 dB, L10 78.0 dB, L50 55.1 dB and L90 54.9 dB, the event second shaded, and an annotation reading LAE 80.1 dB, the whole event energy compressed into one second](https://docs.jmrp.io/phonometry/media/slm_level_track.svg)

*The readouts of this step drawn on the recording that produced them: the
event second lifts $L_\mathrm{Aeq}$ and $L_{10}$, and leaves $L_{50}$ and $L_{90}$ on
the background.*

**Show the code for this figure**

```python
import matplotlib.pyplot as plt

t = np.arange(recording.size) / fs
fig, ax = plt.subplots(figsize=(9, 4))
ax.axvspan(4.0, 5.0, color="C1", alpha=0.15)
ax.plot(t, laf_t, linewidth=0.8, label=r"$L_\mathrm{AF}(t)$")
for value, name, style in [(la_eq, "LAeq", "--"), (ln[10], "L10", ":"),
                           (ln[50], "L50", "-."), (ln[90], "L90", (0, (5, 1)))]:
    ax.axhline(value, linestyle=style, linewidth=1.0,
               label=f"{name} = {value:.1f} dB")
ax.annotate(f"LAE = {lae:.1f} dB: the whole event energy in 1 s",
            xy=(4.5, lae), xytext=(5.6, lae + 2),
            arrowprops={"arrowstyle": "->"})
ax.set(xlabel="Time [s]", ylabel="Level [dB re 20 µPa]", xlim=(0, 10))
ax.legend(loc="lower right", ncols=2)
plt.show()
```

Deep guide: [Integrated and Statistical Levels](https://docs.jmrp.io/phonometry/signals/levels/levels/),
which adds the maximum time-weighted level $L_\mathrm{AFmax}$ that this same `laf_t`
track yields, noise dose and octave spectrograms; the $L_\mathrm{den}$ and rating
levels continue in
[Environmental Levels](https://docs.jmrp.io/phonometry/environment/assessment/environmental-levels/).

## 5. Band-filter: the spectrum view (IEC 61260-1)

A class 1 meter with a filter set reports band levels. `octave_filter`
decomposes the calibrated signal into fractional-octave bands whose design is
anchored to the IEC 61260-1 band edges; `nominal=True` labels them with the
preferred frequencies you would read on an instrument.

```python
filtered = filters.octave_filter(
    recording, fs, fraction=3, nominal=True,
    calibration=filters.LevelCalibration(factor=cal),
)
spl, bands = filtered.levels, filtered.frequencies
# 33 one-third-octave band levels in dB SPL, labeled '12.5' ... '20k'.
# The '1k' band holds the event: ~70 dB, while its neighbors stay ~25 dB below.
print(dict(zip(bands, np.round(spl, 1))))
```

These are **unweighted (Z) band levels**: `octave_filter` ran on `recording`,
not on the `weighted` signal of step 3, so their energy sum reproduces the
unweighted $L_\mathrm{eq}$ and not the $L_\mathrm{Aeq}$ of step 4. That sum is the sanity
check worth running once on every measurement:

```python
print(f"sum of the Z bands  {10 * np.log10(np.sum(10 ** (spl / 10))):.1f} dB")
print(f"Leq(Z)              {signals.leq(recording, calibration_factor=cal):.1f} dB")
# sum of the Z bands  70.3 dB
# Leq(Z)              70.4 dB
```

The tenth or two of deficit is the energy outside the 12.5 Hz–20 kHz analysis
range plus the band-edge shapes. A discrepancy larger than about half a decibel
points at a calibration or a sample-rate mistake rather than at the filters.

To obtain an *A-weighted* spectrum, band-filter the already weighted signal of
step 3 instead of the raw recording:

```python
spl_a = filters.octave_filter(weighted, fs, fraction=3, nominal=True).levels
print(f"sum of the A bands  {10 * np.log10(np.sum(10 ** (spl_a / 10))):.1f} dB")
# sum of the A bands  70.1 dB -> the LAeq of step 4, to a hundredth of a decibel
```

That identity closes the loop between the two halves of the meter: the
weighting stage and the filter stage measure the same energy.

![Thirty-three one-third-octave band levels of the recording of this page as a step plot from 12.5 Hz to 20 kHz, unweighted bands flat near 43 dB with the 1 kHz band highlighted at 70 dB, drawn beside the A-weighted spectrum of the same recording which falls away below 500 Hz and above 4 kHz](https://docs.jmrp.io/phonometry/media/slm_third_octave.svg)

*The same recording read in bands, unweighted and A-weighted. Pink noise gives
nearly constant fractional-octave band levels, so the flat floor is the
background and the single bar 27 dB above it is the event; A-weighting tilts
the floor without touching the 1 kHz band, where the A curve is 0 dB.*

**Show the code for this figure**

```python
centres = filters.octave_filter(recording, fs, fraction=3).frequencies  # exact centres

fig, ax = plt.subplots(figsize=(9, 4))
ax.step(centres, spl, where="mid", label="Z (unweighted)")
ax.step(centres, spl_a, where="mid", linestyle="--", label="A-weighted")
ax.axvspan(891.3, 1122.5, color="C1", alpha=0.15)   # the 1 kHz band
ax.set(xscale="log", xlabel="Band centre frequency [Hz]",
       ylabel="Band level [dB re 20 µPa]")
ax.legend(loc="lower center")
plt.show()
```

Deep guides: [Filter Banks](https://docs.jmrp.io/phonometry/signals/filters/filter-banks/) for the filter
architectures and zero-phase mode,
[Block Processing](https://docs.jmrp.io/phonometry/signals/filters/block-processing/) for streaming, and
[Multichannel and Performance](https://docs.jmrp.io/phonometry/signals/filters/multichannel/) for arrays.

## 6. Verify: is this meter class 1?

A real instrument is only a "class 1 sound level meter" after its weightings
and filters pass the acceptance limits of the standards. The library ships
the same verifiers it applies to itself in CI: `verify_weighting_class`
sweeps a `WeightingFilter` against the IEC 61672-1 Table 3 limits, and
`verify_filter_class` sweeps an `OctaveFilterBank` against the IEC 61260-1
Table 1 limits.

```python
wf = filters.WeightingFilter(fs, weighting="A")
print(filters.verify_weighting_class(wf).overall_class)   # 1

bank = filters.OctaveFilterBank(fs, fraction=3)
print(filters.verify_filter_class(bank).overall_class)    # 1
```

The verdicts also come per band, so you can see exactly where a design would
leave its class corridor. Deep guides:
[Frequency Weighting](https://docs.jmrp.io/phonometry/signals/levels/weighting/) (section on class
verification) and
[Filter Class Verification](https://docs.jmrp.io/phonometry/signals/filters/filter-compliance/) (the
Table 1 mask, class 0 and the compliance fiche).

## 7. Is this measurement valid?

Step 6 verified the *design* of the weighting and filter stages. It says
nothing about the file. A hardware meter refuses to show a level it cannot
stand behind: IEC 61672-1 clause 5.11 makes an overload indicator mandatory
(5.11.5 latches it whenever time-averaged or exposure levels are being
measured), and clause 5.12 adds an under-range indication.
A meter assembled from functions has neither, so three checks on the recording
replace them, and each of the three yields a number that belongs in the report.

```python
# (a) Overload: any run of two or more samples pinned at full scale.
at_full_scale = np.flatnonzero(np.abs(recording) >= 1.0 - 1e-6)
runs = (np.split(at_full_scale, np.flatnonzero(np.diff(at_full_scale) > 1) + 1)
        if at_full_scale.size else [])
clipped_runs = sum(1 for run in runs if run.size >= 2)

# (b) Under-range: the same chain capturing its own noise (dummy capsule).
self_noise = signals.noise_signal(fs, duration_s=3.0, color="pink", rms=2e-4, seed=11)
margin = la_eq - signals.laeq(self_noise, fs, calibration_factor=cal)

# (c) Drift: the calibrator tone recorded again after the measurement.
calibrator_post = 0.985 * calibrator          # 1.5 % lower in this session
cal_post = metrology.sensitivity(calibrator_post, target_spl=94.0, fs=fs)
drift = abs(20 * np.log10(cal_post / cal))

print(f"clipped runs {clipped_runs} | S/N margin {margin:.1f} dB "
      f"| drift {drift:.2f} dB")
# clipped runs 0 | S/N margin 55.3 dB | drift 0.13 dB
```

Refuse the record if any clipped run exists, require at least 10 dB of margin
over the chain's self-noise (below that, $L_{90}$ is reading the preamplifier
rather than the background), and hold the pre/post sensitivity difference to
the 0.5 dB gate of the
[calibration guide](https://docs.jmrp.io/phonometry/signals/metrology/calibration/), which also
explains why that same difference is the drift term of the uncertainty budget.

Learn the fingerprint of the failure the broadband level hides. Clipping barely
moves $L_\mathrm{Aeq}$ (hard-clipping a 1 kHz tone at 80 % of its peak costs 0.9 dB)
while injecting an odd-harmonic ladder that lifts the 3.15 kHz one-third-octave
band by 37 dB, the 5 kHz band by 37 dB and the 8 kHz band by 25 dB. A
one-third-octave spectrum carrying regularly spaced peaks above anything the
source can plausibly radiate is clipping until proven otherwise. The opposite
failure looks nothing like it: a signal far below the converter's range gives a
spectrum that barely changes when the source is switched off, which is what
check (b) catches.

## What this guide covers

- **Covered.** This page composes stages already implemented elsewhere into the pipeline IEC
  61672-1:2013 describes: the A frequency weighting and Fast exponential time
  weighting, $L_\mathrm{Aeq}$, percentile levels, sound exposure level and the
  C-weighted peak, the IEC 61260-1:2014 octave-band filters of `octave_filter`,
  and the Table 3 (weighting) and Table 1 (filter) class acceptance limits
  checked by `verify_weighting_class` and `verify_filter_class`. Each stage's
  own guide states its coverage in detail. The calibrator the pipeline starts
  from has a verdict of its own: `metrology.verify_sound_calibrator` grades what
  a laboratory measured on it against every IEC 60942:2017 requirement that has
  an acceptance limit, by the conformance rule of IEC TC 29. So has a working
  meter's periodic test: `metrology.verify_sound_level_meter_periodic` grades
  what a laboratory measured in the tests of IEC 61672-3:2013, clause by
  clause ([Periodic tests of a sound level
  meter](https://docs.jmrp.io/phonometry/signals/metrology/sound-level-meter-periodic-tests/)).

- **Not covered.** `verify_weighting_class` and `verify_filter_class` check the
  frequency-response design of the digital filters against the standards'
  tables; they do not perform the IEC 61672-2:2013 pattern-evaluation tests a
  physical instrument needs for type approval, such as self-generated noise,
  linearity range, overload indication or directional response, nor the IEC
  61672-3:2013 periodic tests a working instrument receives, whose results a
  laboratory returns and `verify_sound_level_meter_periodic` grades. A class verdict
  from this page describes the algorithm, not a built device, and the three
  checks of step 7 screen the *recording*, not the instrument. The same
  holds for the calibrator: nothing here performs an IEC 60942:2017 test on
  one. `verify_sound_calibrator` grades the numbers a laboratory measured, and
  `sensitivity()` screens the short-term stability of the take against Table 2;
  the [calibration guide](https://docs.jmrp.io/phonometry/signals/metrology/calibration/#verifying-the-calibrator-itself-iec-609422017)
  says exactly what each checks and what it does not.

## See also

The meter built here is the trunk; the rest of the core grows from it.

- [Measurement uncertainty (GUM and Monte Carlo)](https://docs.jmrp.io/phonometry/signals/metrology/gum-uncertainty/):
  attach an uncertainty to the $L_\mathrm{Aeq}$ you just computed, calibration term
  included.
- [Calibrated spectral analysis](https://docs.jmrp.io/phonometry/signals/spectra/spectral-analysis/): when
  bands are too coarse, the Welch PSD with confidence intervals.
- [Correlation, time delay and envelope](https://docs.jmrp.io/phonometry/signals/spectra/correlation-delay/):
  two microphones instead of one, and the delay between them.
- [Block Processing](https://docs.jmrp.io/phonometry/signals/filters/block-processing/): turn this page's
  offline meter into a streaming one with carried filter state.
- API reference: [`metrology.calibration`](https://docs.jmrp.io/phonometry/reference/api/metrology/calibration/),
  [`filters.weighting`](https://docs.jmrp.io/phonometry/reference/api/filters/weighting/),
  [`signals.levels`](https://docs.jmrp.io/phonometry/reference/api/signals/levels/),
  [`phonometry`](https://docs.jmrp.io/phonometry/reference/api/filters/phonometry/) and
  [`filters.compliance`](https://docs.jmrp.io/phonometry/reference/api/filters/compliance/).

## References

- **International Electrotechnical Commission. (2013).** [*Electroacoustics — Sound level meters — Part 1: Specifications*](https://webstore.iec.ch/en/publication/5708) (IEC 61672-1:2013).
  The blueprint of the instrument assembled on this page: the A frequency weighting and the Fast exponential time weighting of the level chain, the C-weighted peak and the sound exposure level, and the Table 3 class acceptance limits checked by verify_weighting_class.
- **International Electrotechnical Commission. (2014).** [*Electroacoustics — Octave-band and fractional-octave-band filters — Part 1: Specifications*](https://webstore.iec.ch/en/publication/5063) (IEC 61260-1:2014).
  The fractional-octave-band filters behind the spectrum stage, and the Table 1 class acceptance limits checked by verify_filter_class.
- **International Electrotechnical Commission. (2017).** [*Electroacoustics — Sound calibrators*](https://webstore.iec.ch/en/publication/30045) (IEC 60942:2017).
  The acoustic calibrator assumed by the sensitivity stage: the 94 dB principal level and the short-term stability check applied to the reference recording.