# Microphone Calibration by Reciprocity (IEC 61094-2/-3)

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/signals/metrology/reciprocity-calibration/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/signals/metrology/reciprocity-calibration/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-10-06
Updated: 2026-11-16T21:36:14Z
Standards: IEC 61094-2:2009; IEC 61094-3:2016; ISO 9613-1:1993
Verified: 34 conformance checks (IEC 61094-2:2009; IEC 61094-3:2016; ISO 9613-1:1993) https://docs.jmrp.io/phonometry/reference/conformance/
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Calibration by reciprocity (IEC 61094-2/-3): laboratory standard microphones without a reference, three pair measurements, coupler or free field.

---

A laboratory standard microphone is calibrated without a reference
microphone. Two microphones are coupled acoustically, one drives as a sound
source and the other receives, and the electrical transfer impedance
$Z_{\mathrm{e},12} = U_2/i_1$ that is measured is the product of their two
sensitivities times the acoustic transfer impedance of whatever couples them,
which is computed:

$$
M_1 M_2 = \frac{Z_{\mathrm{e},12}}{Z_{\mathrm{a},12}}
$$

In a **coupler** (IEC 61094-2) $Z_{\mathrm{a},12}$ is the acoustic impedance
of a small enclosed volume of gas; in a **free field** (IEC 61094-3) it is
that of a spherical wave between the two acoustic centres. Three microphones
measured in pairs give three products, and each sensitivity follows from them
alone. The three modules of `metrology` that do this are
`reciprocity_coupler`, `reciprocity_free_field` and `reciprocity_calibration`,
which holds the result and the budget both share.

The readings on this page are synthetic: electrical transfer impedances built
from three sensitivities the page knows, so every step can be checked against
them.

![Two panels against frequency. Left, the pressure sensitivity level of three LS1P microphones calibrated by reciprocity in a plane-wave coupler, 25 Hz to 10 kHz: flat near -26.1, -26.5 and -25.8 dB re 1 V/Pa up to 1 kHz, rising by about 1 dB to a peak near 5 kHz and falling to about -29 dB at 10 kHz; the expanded uncertainty, 0.015 dB to 0.064 dB, is a band hardly wider than the curves, and the legend states it. Right, the change of the acoustic transfer impedance of the pair 1-2 that each correction causes: heat conduction from -0.32 dB at 25 Hz to 0 dB near 8 kHz, the two capillary tubes from -0.42 dB at 25 Hz up to +0.14 dB near 150 Hz and back to 0, and both together](https://docs.jmrp.io/phonometry/media/reciprocity_pressure.svg)

*Three LS1P microphones calibrated in a plane-wave coupler (left) and the
corrections to the transfer impedance of one pair (right), from the results'
own `.plot()`.*

## 1. How the measurement goes

Both parts use three microphones, two of which must be reciprocal, or two
microphones and an auxiliary sound source, one of which must be reciprocal
(5.1.1). Each pair is measured in turn: one microphone is driven by a current,
measured as the voltage across a calibrated impedance in series with it, and
the other's open-circuit voltage is read by the insert voltage technique
(5.3). The quotient is the electrical transfer impedance. In a coupler the two
face each other across a cavity of known shape; in a free field they face
each other on their principal axes in an anechoic room.

![The readings of a reciprocity calibration in two columns. Left, a pressure field in a coupler (IEC 61094-2, 5.3 to 5.7): two laboratory standard microphones face each other across a small cavity, the transmitter (1) driven by a current i_1 measured across a series impedance and the receiver (2) giving its voltage U_2 by the insert voltage technique, the pairs 1-2, 2-3 and 3-1 in turn; Formula (2) gives M_1 M_2 = Z_e,12 / Z_a,12 with Z_e,12 = U_2 / i_1 measured and Z_a,12 computed by Formula (3) or (4) and Annexes A and B, and Formula (7) gives M_1 = (P_12 P_31 / P_23) to the power one half, P_ij being the product M_i M_j of each pair. Right, the same pair face to face in a free field (IEC 61094-3, 5.5 to 5.7), with the distance d_m12 between the diaphragms and the shorter distance d_12 between the two acoustic centres, which lie in front of the diaphragms; Formula (7) gives M_1 M_2 = -j (2 d_12 / rho f)(U_2 / i_1) exp(j k d_12) exp(alpha d_m12), the phase and spreading taken over d_12 and the attenuation over d_m12, and Formula (8) gives M_1 from the three products as in the coupler. The footer says the free-field distance is above ten diameters (7.3) and the supports twenty diameters long (6.4)](https://docs.jmrp.io/phonometry/media/diagram_reciprocity_setup.svg)

*A pair of the triad in a coupler and in a free field: what is measured, what
is computed and what each part's formulas make of them.*

| Requirement | Value | Clause |
| :--- | :--- | :--- |
| Reference conditions | 23.0 °C, 101.325 kPa, 50 % relative humidity | 61094-2 and 61094-3, 4 |
| Polarizing voltage | 200.0 V recommended, and the one used reported | 61094-2, 6.2; 61094-3, 6.2 |
| Plane-wave coupler | Distance between the diaphragms 0.5 to 0.75 times the diameter, up to about 10 kHz (LS1P) and 20 kHz (LS2P) | 61094-2, C.2 |
| Large-volume coupler | Length to diameter about 0.3, up to about 2.5 kHz (LS1P) and 5 kHz (LS2P) with a wave-motion correction | 61094-2, C.3 |
| Free-field distance | Greater than ten nominal diameters of the microphone | 61094-3, 7.3 |
| Free-field mounting | On a cylinder of the microphone's diameter at least twenty diameters long, tapered to its support | 61094-3, 6.4 |
| Reported uncertainty | Expanded, with $k = 2$ (61094-2) or for a 95 % coverage probability (61094-3), at each frequency | 61094-2, 7.5; 61094-3, 7.8 |

## 2. The microphones and the coupler

Each microphone enters the acoustic transfer impedance through its acoustic
impedance and its front cavity. Annex E writes the impedance as a lumped
compliance, mass and resistance, or equivalently as the equivalent volume at
low frequencies, the resonance frequency and the loss factor; E.4 states the
representation good to about 1.3 times the resonance frequency. The front
cavity's measured volume, depth and diameter come from E.2 and E.3. The
coupler is either a plane-wave coupler of the diameter of the front cavities
or a large-volume coupler, with the nominal dimensions of Tables C.1 and C.2
in `metrology.IEC61094_2_TABLE_C1` and `metrology.IEC61094_2_TABLE_C2`.
Figure C.2 draws each microphone of a large-volume coupler with its face a
distance F back from the cavity, through a bore of the front-cavity diameter
B: `LargeVolumeCoupler` takes that length as `port_length_m`, and the two
bores of the LS1P coupler add 2.4 % to its volume, a tenth of a decibel on
each sensitivity.

```python
import numpy as np
from phonometry import fluids, metrology

mics = tuple(
    metrology.ReciprocityMicrophone(
        equivalent_volume_m3=v,
        resonance_frequency_hz=f0,
        loss_factor=d,
        front_cavity_volume_m3=0.534e-6,   # measured, E.3
        front_cavity_depth_m=1.95e-3,      # D of Table C.1
        front_cavity_diameter_m=18.6e-3,   # B of Table C.1
    )
    for v, f0, d in ((148e-9, 8200.0, 1.05), (141e-9, 8350.0, 1.00), (152e-9, 8050.0, 1.10))
)
coupler = metrology.PlaneWaveCoupler(
    length_m=7.5e-3,
    diameter_m=18.6e-3,
    capillary=metrology.CapillaryTube(length_m=0.05, radius_m=0.2e-3, n_tubes=2),
)
conditions = {"temperature_c": 22.4, "static_pressure_pa": 100200.0, "relative_humidity_percent": 46.0}
f = metrology.exact_frequencies(20, 10000, fraction=3)
check = metrology.check_coupler(f, coupler, mics, **conditions)
print(check.passes, round(check.length_to_diameter_ratio, 3), check.ratio_recommended)   # True 0.613 True

large = metrology.LargeVolumeCoupler(length_m=12.55e-3, diameter_m=42.88e-3, port_length_m=0.80e-3)   # E, C, F of Table C.2
print(round(large.cavity_volume_m3 * 1e9), round(large.closed_volume_m3(mics[:2]) * 1e9))   # 18124 19626 (mm³)
```

`check_coupler` holds the coupler to what its formulas need: Formulas (A.3)
and (A.4) "are valid for the frequency range given by $\omega\rho a^2 >
100\eta$", 3 Hz and 12 Hz for the LS1P and LS2aP couplers of Table C.1; the
approximation (A.2) of a large-volume coupler to $0.125 < R < 8$ and
$X > 5$; and the conditions to the domain Annex F states for its air. Two
things are reported without entering the verdict: the ratio C.2 recommends,
0.5 to 0.75 between the distance of the diaphragms and the diameter, and a
frequency below 20 Hz with the approximation, where A.2 asks for the full
solution "or the corresponding uncertainty component shall be increased
accordingly" (`full_solution_advised`).

## 3. The acoustic transfer impedance and its corrections

In the plane-wave coupler, Formula (4) treats the whole system as a
transmission line of length $l_0$ between the diaphragms, the coupler plus
the two front-cavity depths, terminated at each end by the microphone's
acoustic impedance in parallel with the excess volume of its front cavity
(7.3.3.1):

$$
\frac{1}{Z'_{\mathrm{a},12}} = \frac{1}{Z_{\mathrm{a},0}}\left[
\left(\frac{Z_{\mathrm{a},0}}{Z_{\mathrm{a},1}} + \frac{Z_{\mathrm{a},0}}{Z_{\mathrm{a},2}}\right)\cosh\gamma l_0
+ \left(1 + \frac{Z_{\mathrm{a},0}^2}{Z_{\mathrm{a},1} Z_{\mathrm{a},2}}\right)\sinh\gamma l_0\right]
\tag{4}
$$

The heat conduction and the viscous losses at the cylindrical wall enter
through the propagation coefficient and wave impedance of Formulas (A.3) and
(A.4), and the heat conduction at the two diaphragms through the admittance of
Formula (A.5). In a large-volume coupler the gas is a pure compliance,
Formula (3), and the heat conduction makes its volume look larger by the
complex factor of Formula (A.1),

$$
\Delta_\mathrm{H} = \frac{\kappa}{1 + (\kappa - 1)\,E_V},
\qquad
X = \frac{f\,l^2}{\kappa\,\alpha_t}
$$

with $E_V$ Gerber's temperature transfer function of the cylinder and $l$ its
volume to surface ratio. The capillary tubes that equalise the static
pressure shunt the cavity, and Formula (6) divides the transfer impedance by
$\Delta_\mathrm{C} = 1 + n Z''_{\mathrm{a},12}/Z_{\mathrm{a,C}}$, with the
input impedance of an open tube of Formulas (B.1) to (B.3).

```python
pairs = ((0, 1), (1, 2), (2, 0))
za = [metrology.coupler_transfer_impedance(f, coupler, mics[i], mics[j], **conditions) for i, j in pairs]
print(za[0].heat_conduction_correction[[1, 16, 25]].round(3))   # [-0.287 -0.051  0.004]
print(za[0].capillary_correction_db[[1, 16, 25]].round(3))         # [-0.213  0.011 -0.   ]
```

At 31.6 Hz the heat conduction lowers the transfer impedance by 0.29 dB and
the two tubes by another 0.21 dB; at 1 kHz both are a few hundredths. The two
annexes print tables meant for testing exactly this arithmetic.
`temperature_transfer_function` sums the full solution of the annex's
reference [A.1], which A.2 requires below 20 Hz unless the uncertainty
component is increased instead, and which gives every entry of
Table A.1 to the 0.000 01 the annex states for them; Formula (A.2) is the
approximation, accurate to 0.01 % in modulus for $X > 5$:

```python
print(np.round(metrology.temperature_transfer_function(1.0, 100.0), 5))                      # (0.97179+0.02758j)
print(np.round(metrology.temperature_transfer_function(1.0, 100.0, method="approximation"), 5))  # (0.97179+0.02758j)
air = fluids.air(temperature_c=23.0, static_pressure_pa=101325.0, relative_humidity_percent=50.0)
tube = metrology.capillary_tube_impedance([20.0, 1000.0], length_m=0.05, radius_m=1 / 6000, gas=air)
print((tube.impedance_pa_s_m3 / 1e9).round(3))   # [3.015+0.097j 8.331+4.288j]
```

Table B.1 prints 3.015 and 8.331 GPa·s/m³ for this tube, Table B.2 0.097 and
4.287. The radius the table heads "0,1667 mm" is 1/6 mm: with 0.1667 mm the
values come out 0.08 % low, beyond the last printed figure (an erratum). Even
with 1/6 mm the air of Annex F does not give every entry to its last figure:
the 186 complex entries come back within 0.002 GPa·s/m³, the residue reaching
0.0011 in the real part and 0.0018 in the imaginary near the resonances of the
tubes, and 0.0005 and 0.0007 from 20 Hz to 250 Hz, as the 4.288 here shows.

![Two panels against frequency. Left, the heat-conduction correction 20 lg\|Delta_H\| of the LS1P large-volume coupler of Table C.2 closed by two microphones, with the two bores of length F, R = 0.293 and l = 4.005 mm: 0.61 dB at 2.5 Hz falling smoothly to 0.01 dB at 10 kHz. Right, the real and imaginary parts of the input impedance of an open capillary tube 50 mm long and 1/6 mm in radius at the reference conditions, in GPa s per cubic metre, 20 Hz to 20 kHz: the real part near 3 up to 300 Hz, then six resonance peaks, 12.5 at 1.33 kHz, 8.3 at 4.5 kHz and 6.9, 6.3, 5.9 and 5.7 near 7.7, 11, 14.4 and 17.7 kHz, with troughs between rising from 2.6 to 4.2; the imaginary part rising from 0.1 at 20 Hz to 4.3 near 1 kHz, swinging to -5.6 at 1.7 kHz and oscillating about zero above with a shrinking swing](https://docs.jmrp.io/phonometry/media/reciprocity_coupler_physics.svg)

*The heat conduction in the large-volume coupler of Table C.2 (left) and the
impedance of a capillary tube of Tables B.1 and B.2 at the reference
conditions the tables are printed for (right).*

Table C.3 prints the wave-motion corrections of the air-filled large-volume
coupler for LS1P microphones, to be added to the sensitivity level when they
are not determined for the individual coupler; in hydrogen C.3 scales the
frequency by the ratio of the speeds of sound:

```python
print(metrology.large_volume_wave_motion_correction([1000.0, 2000.0]).correction)   # [-0.002 -0.06 ]
```

## 4. The pressure sensitivities

Formula (2) gives the product of each pair, and Formula (7) the three
sensitivities from the three products:

$$
|M_{\mathrm{p},1}| = \left\{\left|\frac{Z_{\mathrm{e},12}Z_{\mathrm{e},31}}{Z_{\mathrm{e},23}}\right|
\left|\frac{Z''_{\mathrm{a},23}}{Z''_{\mathrm{a},12}Z''_{\mathrm{a},31}}\right|
\left|\frac{\Delta_\mathrm{C,12}\Delta_\mathrm{C,31}}{\Delta_\mathrm{C,23}}\right|\right\}^{1/2}
\tag{7}
$$

```python
true = [m0 * m.complex_equivalent_volume_m3(f) / m.equivalent_volume_m3 for m0, m in zip((0.0496, 0.0473, 0.0512), mics)]
ze = [true[i] * true[j] * z.transfer_impedance_pa_s_m3 for (i, j), z in zip(pairs, za)]   # Formula (2)

c = metrology.pressure_reciprocity(ze, za, frequencies=f)
print(c.sensitivity_level[:, 16].round(2))    # [-26.03 -26.44 -25.76]
print(c.sensitivity_mv_per_pa[:, 16].round(2))   # [49.93 47.64 51.51]
print(np.abs(c.sensitivity_v_per_pa - true).max() < 1e-15)   # True
```

The phase comes from the same products: the complex sensitivity is solved, not
only its modulus. A reciprocity measurement fixes products, and a product does
not change when every sensitivity changes sign together, so the sign of the
square root is a convention. The library takes the branch that is continuous
in frequency with a non-negative real part at the lowest frequency, and
`phase_deg` refers the phase to 0 to 360° as 5.7.1 of both parts asks. With two
microphones and an auxiliary source, the ratio of their sensitivities measured
against the source replaces the third pair, Formula (8):

```python
pair = metrology.pressure_reciprocity_pair(ze[0], za[0], true[0] / true[1], frequencies=f)
print(pair.sensitivity_level[:, 16].round(2))   # [-26.03 -26.44]
```

## 5. A free-field calibration

In a free field the transmitter is, seen from far enough, a simple source at
its acoustic centre. The product of Formula (7) of IEC 61094-3 carries the
distance $d_{12}$ between the acoustic centres, the density of the air, the
phase of the spherical wave and its attenuation over the distance
$d_{\mathrm{m}12}$ the wave travels between the diaphragms:

$$
M_{\mathrm{f},1} M_{\mathrm{f},2} = -\mathrm{j}\,\frac{2 d_{12}}{\rho f}\,
\frac{U_2}{i_1}\,\mathrm{e}^{\mathrm{j}k d_{12}}\,\mathrm{e}^{\alpha d_{\mathrm{m}12}}
\tag{7}
$$

The attenuation is computed by the five steps of Annex B, ISO 9613-1 adjusted
to the quantities of IEC 61094-2, and the wave number takes the speed of
sound of IEC 61094-2 Annex F with the dispersion of its NOTE to F.3, which
Annex B.1 points to.

```python
ff = metrology.exact_frequencies(1000, 40000, fraction=3)
ffc = {"temperature_c": 23.1, "static_pressure_pa": 101100.0, "relative_humidity_percent": 52.0}
att = metrology.reciprocity_air_attenuation(ff, **ffc)
print(att.attenuation_db_per_m[[0, 10, 13, 15]].round(4))   # [0.0053 0.1348 0.4713 0.9877]
```

Three LS2aP microphones 0.25 m apart, with acoustic centres a few millimetres
in front of their diaphragms that recede towards the diaphragm at high
frequency (Annex A shows such curves; these are illustrative):

```python
ratio = ff / 23000.0
true_ff = [m0 / (1.0 - ratio**2 + 1.3j * ratio) for m0 in (0.0128, 0.0124, 0.0131)]
centre = 4.5e-3 / (1.0 + (ff / 30000.0) ** 2)
ze_ff = [
    metrology.free_field_transfer_impedance(
        true_ff[i], true_ff[j], frequencies=ff, diaphragm_distance_m=0.25, acoustic_centers_m=(centre, centre), **ffc
    )
    for i, j in pairs
]   # Formula (D.1)
cf = metrology.free_field_reciprocity(
    ze_ff, frequencies=ff, diaphragm_distances_m=(0.25, 0.25, 0.25), acoustic_centers_m=(centre, centre, centre), **ffc
)
print(cf.sensitivity_level[:, 10].round(2))   # [-37.76 -38.03 -37.55]
print(np.abs(cf.sensitivity_v_per_pa - true_ff).max() < 1e-15)   # True
```

Formula (8) as printed omits the factor $-\mathrm{j}$ that it inherits from
the three products of Formula (7), and Formula (9), which the corrigendum
makes the complex sensitivity, omits it too. The modulus is unaffected; the
phase of the printed formulas is 45° off at every frequency, and Formula (D.1)
keeps the factor. The library follows Formula (7) (an erratum).

![Two panels against frequency. Left, the free-field sensitivity level of three LS2aP microphones calibrated by reciprocity, 1 kHz to 40 kHz: flat near -37.8, -38.1 and -37.6 dB re 1 V/Pa up to 10 kHz, falling to about -39 dB at 20 kHz and -47.5 dB at 40 kHz, with the expanded uncertainty of 0.049 dB to 0.080 dB stated in the legend and drawn as a thin band. Right, the attenuation of sound in air by Annex B at 23.1 degrees, 101.1 kPa and 52 percent on logarithmic axes: the total rising from 0.005 dB/m at 1 kHz to 1.4 dB/m at 50 kHz, dominated above 2 kHz by the oxygen relaxation term, with the nitrogen term flat near 0.0045 dB/m and the classical term rising as the square of frequency](https://docs.jmrp.io/phonometry/media/reciprocity_free_field.svg)

*Three LS2aP microphones calibrated in a free field (left) and the
attenuation of Annex B (right).*

**The acoustic centre** (6.5). Plotting the inverse of the sound pressure a
microphone radiates, corrected for the attenuation, against the distance from
a reference point on it gives a straight line in the far field, and where the
line crosses the axis is the acoustic centre:

```python
r = np.array([0.15, 0.20, 0.25, 0.30, 0.40, 0.50])
alpha = metrology.reciprocity_air_attenuation([10000.0], **ffc).attenuation_np_per_m[0]
p = 0.85 * np.exp(-alpha * r) / (r - 2.4e-3)     # your readings at one frequency
print(round(1000 * metrology.acoustic_center(r, p, attenuation_np_per_m=alpha).position_m, 2))   # 2.4
```

![The inverse of the measured sound pressure corrected for the attenuation, in 1/Pa, against the distance from the reference point, 0 m to 0.5 m: six readings from 0.15 m to 0.50 m, rising from 0.17 to 0.59 1/Pa, lie on a least-squares straight line that crosses zero at 2.4 mm, marked as the acoustic centre](https://docs.jmrp.io/phonometry/media/reciprocity_acoustic_centre.svg)

*The acoustic centre at 10 kHz: the line through the inverse readings
crosses zero 2.4 mm in front of the reference point.*

**The arrangement.** `check_free_field_arrangement` holds the distance to the
ten nominal diameters 7.3 recommends, the supporting cylinder to the twenty
6.4 recommends and the conditions to the domain in which B.2 states its ±10 %
accuracy: its verdict is whether the arrangement is the recommended one. It
also reports whether the distances are within the 150 mm to 500 mm for which
Annex A gives acoustic centres:

```python
arrangement = metrology.check_free_field_arrangement(
    ff, diaphragm_distances_m=(0.25, 0.25, 0.25), microphone_diameter_m=13.2e-3, support_length_m=0.3, **ffc
)
print(arrangement.passes, arrangement.annex_a_range)   # True True
```

## 6. The uncertainty budgets

Both parts list the components of a calibration in their Table 1
(`IEC61094_2_TABLE_1`, `IEC61094_3_TABLE_1`) and ask for each as a standard
uncertainty as a function of frequency. The components of the acoustic
transfer impedance are found, as both describe, "by repeating a calculation
while the various components are changed one at a time by their associated
uncertainty": `coupler_parameter_uncertainty` and
`free_field_parameter_uncertainty` do that, each microphone's parameter moved
in turn and the three changes combined into the one row of the table. The
standard uncertainties they move travel together in a
`CouplerInputUncertainties` or a `FreeFieldInputUncertainties`, one keyword
per input quantity, and one left at zero adds no component.

```python
pu = metrology.coupler_parameter_uncertainty(
    f, coupler, mics,
    metrology.CouplerInputUncertainties(
        u_coupler_length_m=3e-6, u_coupler_diameter_m=3e-6, u_capillary_radius_m=2e-6,
        u_static_pressure_pa=10.0, u_temperature_k=0.05, u_relative_humidity_percent=2.0,
        u_front_cavity_depth_m=3e-6, u_front_cavity_volume_m3=1e-9, u_equivalent_volume_m3=1e-9,
        u_resonance_frequency_hz=50.0, u_loss_factor=0.03,
    ),
    **conditions,
)
print(pu.component_uncertainties_db["capillary_tube_dimensions"][[1, 16]].round(4))   # [0.0086 0.0002]

x = f / 10000
b = metrology.reciprocity_uncertainty_budget(
    {
        **pu.component_uncertainties_db,
        "series_impedance": 0.002, "voltage_ratio": 0.003, "cross_talk": 0.001 + 0.004 * x**2,
        "polarizing_voltage": 0.0025, "radial_wave_motion": 0.03 * x**2,
        "heat_conduction_theory": 0.002, "repeatability": 0.004, "rounding": 0.0006,
    },
    frequencies=f,
)
print(b.expanded_uncertainty_db[[1, 16, 25]].round(3))   # [0.023 0.015 0.041]

pf = metrology.free_field_parameter_uncertainty(
    ff,
    metrology.FreeFieldInputUncertainties(
        u_distance_m=0.1e-3, u_acoustic_center_m=0.5e-3, u_static_pressure_pa=20.0, u_temperature_k=0.1,
        u_relative_humidity_percent=2.0, u_air_attenuation_ratio=0.1 / np.sqrt(3.0),
    ),
    diaphragm_distances_m=(0.25, 0.25, 0.25), acoustic_centers_m=(centre, centre, centre), **ffc,
)
xf = ff / 40000
bf = metrology.reciprocity_uncertainty_budget(
    {
        **pf.component_uncertainties_db,
        "series_impedance": 0.002, "voltage_ratio": 0.003, "cross_talk": 0.002 + 0.01 * xf,
        "noise": 0.005, "reflections": 0.01 + 0.02 * xf, "polarizing_voltage": 0.0025, "repeatability": 0.01,
    },
    frequencies=ff,
    field="free",
)
print(bf.expanded_uncertainty_db[[0, 10, 15]].round(3))   # [0.049 0.054 0.072]
```

The expanded uncertainty is $k = 2$ times the root-sum-square, which IEC
61094-2 7.5 states; IEC 61094-3 7.8 asks for the factor of a 95 % coverage
probability, which `coverage_factor` sets when a budget is dominated by a
component with few degrees of freedom. The free-field budget moves each
acoustic centre by 0.5 mm in turn, and the attenuation by the ±10 % B.2
states, taken as a rectangular distribution. The components on this page are
illustrative.
Passed as `expanded_uncertainty_db`, the budget draws the band of the
calibration's plot, and its legend states how wide the band is.

![Two panels of standard uncertainty in dB against frequency, each with a legend of its components below it. Left, the budget of microphone 1 of the pressure triad, 25 Hz to 10 kHz: the expanded uncertainty with k = 2 from 0.035 dB at 25 Hz, where the capillary tubes dominate, down to 0.015 dB to 0.017 dB from 50 Hz to 2 kHz and up to 0.064 dB at 10 kHz, where the radial wave motion dominates. Right, the budget of the free-field triad, 1 kHz to 40 kHz: the expanded uncertainty rising from 0.049 dB to 0.080 dB, the acoustic centres near 0.018 dB and the reflections from 0.010 to 0.030 dB the largest components](https://docs.jmrp.io/phonometry/media/reciprocity_budgets.svg)

*The budgets of microphone 1 of each triad.*

**Show the code for these figures**

```python
import matplotlib.pyplot as plt

# The two calibrations again, with their budgets as the band.
c = metrology.pressure_reciprocity(ze, za, frequencies=f, expanded_uncertainty_db=b.expanded_uncertainty_db)
cf = metrology.free_field_reciprocity(
    ze_ff, frequencies=ff, diaphragm_distances_m=(0.25, 0.25, 0.25), acoustic_centers_m=(centre, centre, centre),
    expanded_uncertainty_db=bf.expanded_uncertainty_db, **ffc,
)

fig, (ax_cal, ax_corr) = plt.subplots(1, 2, figsize=(13.5, 5.6))
c.plot(ax_cal)
za[0].plot(ax_corr)
fig.tight_layout()

fh = metrology.exact_frequencies(1, 10000, fraction=3)
heat = metrology.heat_conduction_correction(
    fh[fh >= 2.0],
    volume_m3=large.closed_volume_m3(mics[:2]),
    surface_area_m2=large.closed_surface_m2(mics[:2]),
    length_to_diameter_ratio=large.length_m / large.diameter_m,
    gas=fluids.air(**conditions),
)
curve = metrology.capillary_tube_impedance(np.geomspace(20, 20000, 1500), length_m=0.05, radius_m=1 / 6000, gas=air)
fig, (ax_heat, ax_tube) = plt.subplots(1, 2, figsize=(13.5, 5.6))
heat.plot(ax_heat)
curve.plot(ax_tube)
fig.tight_layout()

fig, (ax_cal, ax_air) = plt.subplots(1, 2, figsize=(13.5, 5.6))
cf.plot(ax_cal)
metrology.reciprocity_air_attenuation(metrology.exact_frequencies(1000, 50000, fraction=3), **ffc).plot(ax_air)
fig.tight_layout()

fig, (ax_p, ax_f) = plt.subplots(1, 2, figsize=(14.5, 8.2))
b.plot(ax_p)
bf.plot(ax_f)
fig.tight_layout()

fig, ax = plt.subplots(figsize=(7.5, 5.2))
metrology.acoustic_center(r, p, attenuation_np_per_m=alpha).plot(ax)
fig.tight_layout()
plt.show()
```

## What this guide covers

- **Covered.** **IEC 61094-2:2009:** the products of Formula (2) and the sensitivities of
  Formulas (7) and (8), complex; the acoustic transfer impedance of a
  large-volume coupler by Formula (3) and of a plane-wave coupler by Formula
  (4), with the excess volume of each front cavity (7.3.3.1), the heat
  conduction of Annex A (Formula (A.1) with Gerber's full solution of [A.1] or
  the approximation (A.2), Formulas (A.3) to (A.5)), and the capillary tubes of
  Formula (6) and Annex B; the lumped microphone impedance and Formula (E.1) of
  Annex E; Tables C.1, C.2 and C.3 as published data, the wave-motion
  correction of Table C.3 with the frequency scaling of C.3 for hydrogen; the
  check of what the formulas require (A.2, A.3, Annex F); Table 1 and the
  one-at-a-time recalculation of 7.5. Table A.1 is reproduced to its stated
  0.000 01; Tables B.1 and B.2 to within 0.002 GPa·s/m³, more than their
  printed half-unit near the resonances of the tubes (section 3).
  **IEC 61094-3:2016 with COR1:2016:** the products of Formula (7), the
  sensitivities of Formulas (8) and (9) with the factor of (7), Formula (D.1);
  the attenuation of Annex B and the speed of sound with the dispersion of IEC
  61094-2 F.3; the acoustic centre of 6.5; the arrangement of 6.4 and 7.3;
  Table 1 and the recalculation of 7.8. Table B.1 is reproduced by ISO 9613-1,
  as its text says it was computed.

- **Covered, by a reading of the text.** **The branch of the square root.** IEC 61094-2 asks for the phase "by a
  similar procedure from the phase angle of each term" and IEC 61094-3 derives
  it from Formula (8), both referred to 0 to 2π; the library takes the branch
  continuous in frequency with a non-negative real part at the lowest
  frequency, and fixes the other two microphones by the products. **Dimension F
  of Table C.2** is the length of the bore that Figure C.2 draws, of the
  front-cavity diameter B, between each microphone face and the cavity; its
  volume and its wall join those of the coupler. **The surface of a
  large-volume coupler** closed by microphones is that of the closed cylinder
  plus the wall of each bore, of each front cavity and of its thread, the
  diaphragm taking the place of the port it closes (7.3.2.2). **The ratio of
  C.2** is read on the distance between the diaphragms, which 5.4 and C.2 call
  the length of the coupler.

- **Not covered.** The measurements themselves: the insert voltage technique, cross-talk and
  the ground-shield configuration (5.3, 6.3, 7.2). The radial wave-motion
  corrections of plane-wave couplers, which C.2 leaves to the literature, and
  the environmental corrections of Annex D (IEC 61094-2) and Annex C (IEC
  61094-3), which are the first-order correction of the comparison guide. The
  determination of the front cavity and equivalent volumes of E.3 and the data
  fitting of E.4. The free-field to pressure relation of Formula (4) of IEC
  61094-3, which needs the radiation impedance and the scattering factor, and
  the time-selective processing of Annex D, which follows IEC 61094-8 Annex B.
  The acoustic centres of Annex A, printed only as a figure.

## See also

- [Microphone calibration by comparison](https://docs.jmrp.io/phonometry/signals/metrology/comparison-calibration/):
  IEC 61094-5 and IEC 61094-8, which carry a reciprocity-calibrated
  reference's sensitivity over to a working standard microphone, with the
  environmental correction of Annex D.
- [Humid air](https://docs.jmrp.io/phonometry/fluids/humid-air/): the IEC 61094-2 Annex F air the
  couplers and the free field are computed with.
- [Outdoor propagation](https://docs.jmrp.io/phonometry/environment/propagation/outdoor-propagation/):
  ISO 9613-1, which Table B.1 of IEC 61094-3 is computed by.
- [Measurement uncertainty](https://docs.jmrp.io/phonometry/signals/metrology/gum-uncertainty/):
  the GUM the budgets are combined by.
- API reference: [`metrology.reciprocity_calibration`](https://docs.jmrp.io/phonometry/reference/api/metrology/reciprocity-calibration/),
  [`metrology.reciprocity_coupler`](https://docs.jmrp.io/phonometry/reference/api/metrology/reciprocity-coupler/),
  [`metrology.reciprocity_free_field`](https://docs.jmrp.io/phonometry/reference/api/metrology/reciprocity-free-field/).

## References

- **International Electrotechnical Commission. (2009).** [*Electroacoustics — Measurement microphones — Part 2: Primary method for pressure calibration of laboratory standard microphones by the reciprocity technique*](https://webstore.iec.ch/en/publication/4486) (IEC 61094-2:2009).
  The pressure calibration: Formulas (2) to (8), the heat conduction of Annex A with Gerber's full solution and the approximation (A.2), the capillary tubes of Annex B, the couplers and wave-motion corrections of Annex C, the microphone parameters of Annex E, Annex F for the air and the uncertainty components of Table 1. Read from BS EN 61094-2:2009, the English text of EN 61094-2:2009, which is IEC 61094-2:2009 unchanged.
- **International Electrotechnical Commission. (2016).** *Electroacoustics – Measurement microphones – Part 3: Primary method for free-field calibration of laboratory standard microphones by the reciprocity technique* (IEC 61094-3:2016).
  The free-field calibration: Formulas (6) to (9) and (D.1), the acoustic centre of 6.5, the arrangement of 6.4 and 7.3, the attenuation of Annex B, Table B.1 and the uncertainty components of Table 1. Read from Edition 2.0 (2016-06), English-French, with its corrigendum IEC 61094-3:2016/COR1:2016 (2016-12), which makes Formula (9) the complex sensitivity.
- **International Organization for Standardization. (1993).** [*Acoustics — Attenuation of sound during propagation outdoors — Part 1: Calculation of the absorption of sound by the atmosphere*](https://www.iso.org/standard/17426.html) (ISO 9613-1:1993).
  The method Table B.1 of IEC 61094-3 is computed by, and which B.2 adjusts.
- **Joint Committee for Guides in Metrology. (2008).** [*Evaluation of measurement data — Guide to the expression of uncertainty in measurement*](https://www.bipm.org/documents/20126/2071204/JCGM_100_2008_E.pdf) (JCGM 100:2008, the GUM). BIPM. https://doi.org/10.59161/JCGM100-2008E
  The combination both budgets ask for; also published as ISO/IEC Guide 98-3:2008.