# Random-Incidence and Diffuse-Field Response (IEC 61183)

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/signals/metrology/random-incidence/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/signals/metrology/random-incidence/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-09-24
Updated: 2026-11-16T21:36:14Z
Standards: IEC 61183:1994
Verified: 8 conformance checks (IEC 61183:1994) https://docs.jmrp.io/phonometry/reference/conformance/#random-incidence-and-diffuse-field-sensitivity-iec-61183
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Random incidence on a sound level meter: the IEC 61183:1994 directivity factor, and the random-incidence and diffuse-field sensitivity levels.

---

A sound level meter is calibrated for sound arriving from one direction, its
**reference direction**, and most of the sound it measures arrives from all of
them: the reverberant field of a room, the traffic of a street, the machinery
of a workshop. Its microphone and its case are not transparent at high
frequency, so a meter that reads a plane wave from the front correctly reads a
field from every direction low. IEC 61183:1994 gives the two ways of finding
by how much. The **free-field method** rotates the instrument in an anechoic
room and weights the reading for each direction by the share of the sphere it
stands for; the **diffuse-field method** compares the instrument with a
reference instrument in a reverberation room.

This page runs both on a synthetic meter: a microphone whose pattern has, at
each frequency, the directivity Table B.1 of the standard prints for a type
LS2aP/LS2F laboratory standard microphone, mounted on a case that narrows it
in the plane the case sits in.

![Two panels for the synthetic meter at 8 kHz. Left: a polar plot of the level indicated for each direction of incidence relative to the reference direction, 0 dB at 0 degrees falling to about -2 dB at 90 degrees and -17 dB at 180 degrees, one curve for the X-Y plane and a slightly narrower one for the X-Z plane, with 10 lg gamma = 2.66 dB in the title. Right: the weight of each 10 degree reading in per cent of the sphere, two humps rising from 0.10 % at 0 and 180 degrees to 2.18 % at 90 and 270 degrees](https://docs.jmrp.io/phonometry/media/random_incidence_directivity.svg)

*The response of the meter at 8 kHz in the two planes of Annex A (left), and
the weight each reading takes in the sum (right), from the result's own
`.plot()`.*

## 1. The directivity factor

What the standard needs is the ratio of what the meter reads for a plane wave
from its reference direction to what it reads, on average over the sphere, for
the same wave from every direction. That ratio is the **directivity factor**
$\gamma$ of Formula (2), written in Formula (3) with the angle $\phi$ from the
reference direction and the angle $\alpha$ of the plane about it:

$$
\gamma = \frac{4\pi}{\displaystyle\int_0^{2\pi}\!\!\int_0^{\pi}
10^{-0.1\left[L_\mathrm{rd} - L(\phi,\alpha)\right]}\,
\lvert\sin\phi\rvert\,\mathrm{d}\alpha\,\mathrm{d}\phi}
$$

An omnidirectional meter has $\gamma = 1$; one that favours its reference
direction has $\gamma > 1$. The **random-incidence sensitivity level** is then
the free-field sensitivity level for the reference direction less the
directivity index (Formula (1)):

$$
G_\mathrm{RI} = G_\mathrm{F} - 10\lg\gamma,
\qquad G_\mathrm{F} = L_\mathrm{rd} - L_\mathrm{o}
$$

$G_\mathrm{F}$ depends on the individual meter, $\gamma$ only on its dimensions
and geometry, so one measurement of $\gamma$ serves every meter of a model
(4.2).

### How the measurement goes

![The two calibrations of IEC 61183 side by side. Left, the free-field method in plan: an anechoic room lined with wedges, a fixed source on the X axis, and the meter on a turntable with the microphone on the axis of rotation, its reference direction at an angle phi from the X axis, turned through 360 degrees in steps of delta phi. Right, the diffuse-field method: a reverberation room with two uncorrelated sources, and the reference meter and the meter under test placed in turn on the same circular path, not parallel to any wall, of radius at least 1 m and three times the meter, reading L_D,ref and L_D. Under each room, the readings in order: L_o with a reference microphone, L_rd with the meter facing the source, L(phi, h) in the X-Y plane, L(phi, v) after turning the meter 90 degrees about its own axis, and the weighted sum that gives gamma and G_RI; then L_D,ref, L_D, their difference by Formula (8), and G_D,ref added by Formula (9), (10) or (11)](https://docs.jmrp.io/phonometry/media/diagram_random_incidence_setup.svg)

*The meter is turned in front of a fixed source in an anechoic room (left),
or compared with a reference on the same path in a reverberation room
(right). The library takes over once the levels are read.*

| Requirement | Value | Clause |
| :--- | :--- | :--- |
| Anechoic room | Meets ISO 3745; pure tones or random noise, bands no wider than one-third octave, filters of IEC 61260 class 0 or 1 | 4.10 |
| Source | Far enough that the level varies by less than ±1 dB within 0.3 m of the microphone | A.2.2 |
| Signal | At least 20 dB above the background, and held constant during each rotation | A.2.3 |
| Turntable | The acoustical centre of the microphone on the axis of rotation, the reference direction and the source in the plane of rotation | A.2.1, A.4.2 |
| Rotations | 360° in the X-Y plane, then the meter turned 90° about its own axis and 360° again, the X-Z plane | A.4.5, A.4.6 |
| Angular step | Small enough that no element of the sphere exceeds 3 %; 10° leaves 2.2 % | A.1.6, A.1.7 |
| Pure tones | $G_\mathrm{F}$ and $\gamma$ may need the rms average of at least eight tones per one-third-octave band, spaced evenly on a logarithmic axis | 4.11 |
| Reverberation room | Meets ISO 3741; broadband or filtered random noise, bands no wider than one-third octave | 5.6 |
| Integration time | Long enough that repeated results scatter by less than 0.05 dB: 2 min from 500 Hz, 8 min from 250 Hz, 15 min from 125 Hz, longer below | 5.6, B.1.4 |
| Microphone path | Both microphones moved in turn along the same circular path, not parallel to any wall, of radius the larger of 1 m and three times the largest dimension of the meter; two uncorrelated omnidirectional sources of about equal power help the diffusivity | B.1.3, B.1.4 |
| Reference meter | A directivity factor as near unity as possible; a type LS2aP/LS2F or LS2bP microphone is recommended | B.1.1, B.1.2 |

## 2. Readings in two planes, and what each is worth

In practice the integral becomes a sum over a limited number of directions,
each reading taken to stand for the element of the sphere around it (Formula
(4)). Annex A rotates the meter through 360° in the X-Y plane, turns it 90°
about its axis, and rotates it again, which is the X-Z plane: two planes, the
angle $\phi$ in equal steps $\Delta\phi$. The elements are rings cut into
quarters by the planes, so a reading at 90° stands for far more of the sphere
than one near the axis, and Formulas (A.1) and (A.2) give the factor
$K(\phi)$ each reading is weighted with:

$$
K(\phi) = \tfrac{1}{8}\left[\cos\left(\phi - \tfrac{\Delta\phi}{2}\right) -
\cos\left(\phi + \tfrac{\Delta\phi}{2}\right)\right],
\qquad K(0) = K(180°) = \tfrac{1}{4}\left[1 - \cos\tfrac{\Delta\phi}{2}\right]
$$

```python
from phonometry import metrology

k = metrology.adjustment_factors(10.0)   # 36 factors, 0° to 350°
print(k[[0, 1, 9]].round(5))             # [0.00095 0.00378 0.02179]
print(round(2 * k.sum(), 12))            # 1.0
print(round(100 * metrology.largest_element_fraction(10.0), 2))  # 2.18
```

Those are the first, second and last rows of Table A.1, and all ten of its rows
come out to the five decimals it prints. A.1.6 asks for the largest element to
be no more than 3 % of the sphere; at 10° steps it is the one at 90°, 2.18 %,
the "approximately 2,2 %" of A.1.7. A step of 15° would leave 3.26 %, and
`adjustment_factors` then warns with a `SphereDivisionWarning`. Any step that
divides 180° works, and `n_planes=4` gives the four planes at 45° that NOTE 2 of
A.6 asks for when the reference direction is not normal to the diaphragm: every
factor halves.

The 72 factors sum to one only when the readings at 0° and 180° enter both
sums of Formula (A.3), and the paragraph under it could be read otherwise: it
says those readings are the same in the two planes and "have only to be taken
into account once". They have to be measured once. Each plane's pole factor
covers half the polar cap, so counted in one sum only, the factors add up to
0.998 097: the sum loses $K(0)$ times the energy read at each pole, and
$10\lg\gamma$ comes out high by
$-10\lg\left(1 - \gamma K(0)\left[10^{-0.1[L_\mathrm{rd} - L(0°)]} + 10^{-0.1[L_\mathrm{rd} - L(180°)]}\right]\right)$.
That is 0.008 dB for a meter that reads the same in every direction, and it
grows with the directivity: about 0.02 dB at $10\lg\gamma = 7$ dB. The library
counts the poles in both sums, and the omnidirectional meter has $\gamma = 1$
exactly.

The directivity factor is Formula (A.3), one row of readings per plane:

```python
import numpy as np

def pattern_db(phi_deg, n, floor=0.02):
    """L(phi) - L_rd of a meter whose squared pressure is (1 - b)[(1 + cos phi)/2]^n + b."""
    lobe = ((1 + np.cos(np.radians(phi_deg))) / 2) ** n
    return 10 * np.log10((1 - floor) * lobe + floor)

phi = np.arange(0, 360, 10)
levels = 94.0 + np.vstack((pattern_db(phi, 0.785),    # X-Y plane (h), 8 kHz
                           pattern_db(phi, 0.98)))    # X-Z plane (v), the case narrows it
d = metrology.directivity_factor(levels)
print(round(d.gamma, 3), round(d.directivity_index, 2))   # 1.845 2.66
d.plot()                  # the polar response, one curve per plane
d.plot(view="weights")    # K(phi) of each reading
```

The reference level $L_\mathrm{rd}$ defaults to the reading at 0° in the first
plane, which A.4.4 takes with the meter in that very position; pass
`reference_level=` when it was measured separately.

## 3. One plane, or 38 equal elements

A meter that is rotationally symmetric about its reference direction needs one
rotation only (NOTE 1 of A.4.7), and Formula (A.4) counts that plane twice:

```python
a = metrology.axisymmetric_directivity_factor(levels[0])
print(round(a.directivity_index, 2))   # 2.45
```

That is the X-Y plane alone, whose pattern was built to have the 2.45 dB Table
B.1 prints for the microphone at 8 kHz; the case of the synthetic meter is what
adds the other 0.21 dB in the two-plane result.

The note to A.1.8 offers the other way round: place the directions so that the
elements are equal, 38 of them at 2.6 % of the sphere each, and every reading
weighs the same (Formula (A.5)). Each direction here halves its element's
area in polar angle, which reproduces the list the note prints:

```python
angles = metrology.equal_area_incidence_angles()
horizontal, vertical = angles.horizontal_deg, angles.vertical_deg
print(horizontal[:6].round(1))   # [ 0.  32.6 50.8 65.1 77.8 90. ]
e = metrology.equal_area_directivity_factor(
    94.0 + pattern_db(horizontal, 0.785), 94.0 + pattern_db(vertical, 0.98)
)
print(round(e.directivity_index, 2))   # 2.66
```

The note prints 77.9° and 282.1° where the construction gives 77.85° and
282.15°. All the other pairs of the printed list sum to 180.0° about the grazing
direction, and 77.9° + 102.2° is 180.1°: the two are in the [errata
register](https://docs.jmrp.io/phonometry/reference/errata/). A tenth of a degree does not move a
directivity factor, but the list is not symmetric as printed.

## 4. The random-incidence sensitivity level, band by band

With $\gamma$ measured at each band and $G_\mathrm{F}$ from the free-field
calibration of A.3 ($L_\mathrm{rd}$ with the meter in place, $L_\mathrm{o}$
with a free-field reference microphone in its stead), Formula (1) gives the
random-incidence sensitivity level. Here the synthetic meter is measured at
every preferred frequency Table B.1 prints, its X-Y plane given the
$10\lg\gamma$ of the microphone at that band, and its free-field sensitivity
level rolls off above 10 kHz:

```python
bands = sorted(metrology.IEC61183_TABLE_B1)   # the 30 preferred frequencies, 25 Hz to 20 kHz

def meter(frequency_hz):
    """The synthetic meter at one band: the X-Y plane has the 10 lg gamma Table B.1
    prints for the microphone, and the case narrows the X-Z plane."""
    target = metrology.IEC61183_TABLE_B1[frequency_hz].directivity_index
    n = 0.98 / (10 ** (-target / 10) - 0.02) - 1   # the exponent with that index
    return metrology.directivity_factor(
        94.0 + np.vstack((pattern_db(phi, n), pattern_db(phi, 1.25 * n))))

index = [meter(f).directivity_index for f in bands]
g_f = -10 * np.log10(1 + (np.array(bands) / 25000) ** 4)   # L_rd - L_o at each band
r = metrology.random_incidence_sensitivity(g_f, index, frequencies=bands)
shown = np.isin(r.frequencies, [1000, 4000, 8000, 16000])
print(r.correction[shown].round(2))              # [-0.06 -0.95 -2.66 -5.62]
print(r.random_incidence_level[shown].round(2))  # [-0.06 -0.95 -2.71 -6.29]
r.plot()                      # G_F and G_RI
r.plot(view="correction")     # G_RI - G_F = -10 lg gamma
```

![Two panels against frequency from 25 Hz to 20 kHz for the synthetic meter. Left: G_F, flat at 0 dB up to 8 kHz and rolling off to -0.7 dB at 16 kHz and -1.5 dB at 20 kHz, and G_RI drawn dashed with open squares, on top of G_F up to 800 Hz and then falling below it to about -0.9 dB at 4 kHz, -2.7 dB at 8 kHz and -8.5 dB at 20 kHz. Right: the correction G_RI - G_F = -10 lg gamma, from 0 dB up to 800 Hz to -2.7 dB at 8 kHz and -7 dB at 20 kHz](https://docs.jmrp.io/phonometry/media/random_incidence_correction.svg)

*$G_\mathrm{F}$ and $G_\mathrm{RI}$ of the synthetic meter at every preferred
frequency Table B.1 prints (left), and the correction between them (right).*

The correction is what a meter calibrated in a free field reads low in a
random-incidence field: negligible up to about 1 kHz, where the wavelength is
large against the microphone, and several decibels by 10 kHz. It is why a
meter is specified for one field or the other. IEC 61672-1:2013 applies its
frequency-weighting limits to the free-field response or to the
random-incidence response, as applicable (5.5.4), and has the random-incidence
response determined by the free-field method of IEC 61183 (5.5.5), the method
of this page. A meter flat for a plane wave from the front reads a
random-incidence field low by this correction, and one flat in a
random-incidence field reads that plane wave high by it: it cannot be flat for
both.

## 5. The diffuse-field method

The second method needs no anechoic room. The meter under test and a
reference meter are placed in turn at the same positions in a reverberation
room, and the difference of what they indicate (Formula (8)) is added to the
diffuse-field sensitivity level of the reference, known in one of three ways:

$$
\Delta G_\mathrm{D} = L_\mathrm{D} - L_\mathrm{D,ref},
\qquad
G_\mathrm{D} = \Delta G_\mathrm{D} +
\begin{cases}
G_\mathrm{RI,ref} & \text{(9), calibrated by clause 4}\\
G_\mathrm{F,ref} - 10\lg\gamma_\mathrm{ref} & \text{(10), free-field calibrated}\\
G_\mathrm{P,ref} + \Delta_\mathrm{DP} & \text{(11), pressure calibrated}
\end{cases}
$$

Annex B recommends a type LS2aP/LS2F or LS2bP laboratory standard
microphone as the reference, and Table B.1 prints, for the first, both its $10\lg\gamma$ and $\Delta_\mathrm{DP}$ from 25 Hz to 20 kHz.
`metrology.IEC61183_TABLE_B1` holds the table, and Formulas (10) and (11) take
their correction from it unless told otherwise:

In an 80 dB diffuse field each instrument indicates the field plus its own
diffuse-field sensitivity level: the meter its random-incidence level (1.2),
and a reference LS2aP whose pressure sensitivity level is 0 dB its
$\Delta_\mathrm{DP}$:

```python
print(metrology.IEC61183_TABLE_B1[8000.0])
# ReferenceMicrophoneRow(directivity_index=2.45, diffuse_pressure_difference_db=1.2)

delta_dp = np.array([metrology.IEC61183_TABLE_B1[f].diffuse_pressure_difference_db
                     for f in bands])
dd = metrology.diffuse_field_sensitivity(
    80.0 + r.random_incidence_level,  # L_D, the meter under test
    80.0 + delta_dp,                  # L_D,ref, a pressure-calibrated LS2aP
    frequencies=bands,
    reference_pressure_level=0.0,     # G_P,ref; Delta_DP from Table B.1
)
print(dd.reference_correction[shown])            # [0.   0.25 1.2  3.05]
print(dd.diffuse_field_level[shown].round(2))    # [-0.06 -0.95 -2.71 -6.29]
dd.plot()
```

![The diffuse-field sensitivity level of the synthetic meter against a pressure-calibrated LS2aP microphone from 25 Hz to 20 kHz, three curves that lie together at 0 dB up to 800 Hz: the reference's diffuse-field level, Delta_DP of Table B.1, rising to 3.05 dB at 16 kHz and 2.2 dB at 20 kHz; the level difference of Formula (8) falling to about -3.9 dB at 8 kHz and -10.7 dB at 20 kHz; and their sum, G_D, which retraces the random-incidence level of the previous figure, -8.5 dB at 20 kHz](https://docs.jmrp.io/phonometry/media/diffuse_field_sensitivity.svg)

*Formula (11) on the synthetic meter: the pressure-calibrated reference reads
the diffuse field high by $\Delta_\mathrm{DP}$, and adding it back to the
level difference returns the meter's own diffuse-field level, which is its
random-incidence level (1.2).*

The two methods meet here: for the purpose of the standard the diffuse-field
sensitivity level may be used interchangeably with the random-incidence one
(1.2), and the example returns, from a comparison in a room, the level the
first method measured in an anechoic room. `reference_free_field_level`
selects Formula (10) and `reference_random_incidence_level` Formula (9);
`reference_directivity_index` and
`reference_diffuse_pressure_difference_db` replace the Table B.1 defaults for a
reference microphone of another type.

**Show the code for these figures**

```python
import matplotlib.pyplot as plt

fig = plt.figure(figsize=(13.5, 6.0))
d.plot(fig.add_subplot(1, 2, 1, projection="polar"))
d.plot(fig.add_subplot(1, 2, 2), view="weights")
fig.tight_layout()

fig, (ax_levels, ax_correction) = plt.subplots(1, 2, figsize=(13.5, 5.4))
r.plot(ax_levels)
r.plot(ax_correction, view="correction")
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
dd.plot(ax)
fig.tight_layout()
plt.show()
```

## What this guide covers

- **Covered.** **IEC 61183:1994, the calculations of both methods:** the adjustment factors
  of Formulas (6), (7), (A.1) and (A.2) for any angular step that divides the
  half circle and any number of planes, with the halving of NOTE 2 of A.6 and
  the doubling of Formula (A.4); the largest element of the division and the 3 %
  criterion of A.1.6, as a warning; the directivity factor from two or more
  planes (Formula (A.3)), from one plane under rotational symmetry (Formula
  (A.4)) and from 38 equal-area elements (Formula (A.5)), with the directions of
  the note to A.1.8; the random-incidence sensitivity level of Formulas (1) and
  (A.6); the diffuse-field sensitivity level by the three routes of Formulas (8)
  to (11); and Table B.1, which supplies the reference corrections. Table A.1 is
  reproduced to its five decimals, and the 2.2 % of A.1.7 and the 2.6 % of the
  note to A.1.8 are reproduced.

- **Not covered.** The measurements themselves: the anechoic room of ISO 3745 and the
  reverberation room of ISO 3741 the standard requires, the placement of the
  source and of the meter on its turntable, the signal-to-noise ratio of A.2.3
  and the integration times of B.1.4. The averaging of at least eight tones
  within a one-third-octave band of 4.11 is left to the caller, who passes the
  averaged level of each band. The corrections of IEC 62585 that bring a
  measurement made with a sound calibrator, a comparison coupler or an
  electrostatic actuator to the meter's free-field response are on [their own
  page](https://docs.jmrp.io/phonometry/signals/metrology/free-field-corrections/); its corrections
  for the case, the microphone and the windscreen (clauses 9 to 11) are not
  implemented.

## See also

- [Free-field corrections](https://docs.jmrp.io/phonometry/signals/metrology/free-field-corrections/):
  IEC 62585, the corrections that bring a meter on a calibrator, a coupler or
  an actuator to its free-field response, for a meter with a free-field
  microphone.
- [Calibration and dBFS](https://docs.jmrp.io/phonometry/signals/metrology/calibration/): the
  calibrator tone that sets the pressure sensitivity every level of a meter
  starts from.
- [Compliance and verification](https://docs.jmrp.io/phonometry/signals/metrology/compliance-verification/):
  what the IEC 61672-1 class of a meter asserts, and which of its tests the
  library runs.
- [Microphone characterisation](https://docs.jmrp.io/phonometry/devices/electroacoustics/microphones/):
  the IEC 60268-4 directivity index of a microphone, the same $10\lg\gamma$ for
  a transducer on its own.
- [Build a sound level meter](https://docs.jmrp.io/phonometry/signals/sound-level-meter/): the
  whole measurement chain on one runnable page.
- API reference: [`metrology.random_incidence`](https://docs.jmrp.io/phonometry/reference/api/metrology/random-incidence/).

## References

- **International Electrotechnical Commission. (1992).** *Measurement microphones — Part 1: Specifications for laboratory standard microphones* (IEC 61094-1:1992).
  The type LS2aP/LS2F and LS2bP laboratory standard microphones Annex B recommends for the reference instrument of the diffuse-field method, whose characteristics Table B.1 prints.
- **International Electrotechnical Commission. (1994).** *Electroacoustics — Random-incidence and diffuse-field calibration of sound level meters* (IEC 61183:1994).
  The implemented method: the directivity factor of Formulas (2) to (5) and (A.3) to (A.5), the adjustment factors of Formulas (6), (7), (A.1) and (A.2) and their halving in NOTE 2 of A.6, the 3 % criterion of A.1.6, the random-incidence sensitivity level of Formulas (1) and (A.6), the diffuse-field sensitivity level of Formulas (8) to (11), and Table B.1. Read from BS EN 61183:1995, the English text of EN 61183:1994, which is IEC 1183:1994 unchanged.