# Free-Field Corrections of a Sound Level Meter (IEC 62585)

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/signals/metrology/free-field-corrections/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/signals/metrology/free-field-corrections/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-09-24
Updated: 2026-11-16T21:36:14Z
Standards: IEC 62585:2012; IEC 61672-1:2013
Verified: 66 conformance checks (IEC 62585:2012; IEC 61672-1:2013) https://docs.jmrp.io/phonometry/reference/conformance/
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: IEC 62585:2012 corrections for a sound level meter's periodic test: the adjustment value, free-field corrections, Annex I uncertainty, verdict and fiche.

---

A periodic test of a sound level meter by IEC 61672-3 does not put the meter
in an anechoic room. It drives the microphone with a sound calibrator, a
comparison coupler or an electrostatic actuator, and none of those is the
plane progressive wave the meter is specified for: the microphone and the case
respond differently to a pressure in a small cavity, or to an electrostatic
force on the diaphragm, than to a wave arriving from the front. The test
therefore needs, at each frequency, the **correction** that turns what the
meter indicates on that source into what it would indicate in a free field of
the same level, and the manufacturer has to state it in the manual. IEC
62585:2012 gives the methods for finding those corrections, the uncertainty
they may carry, and the adjustment value the manual quotes at the
calibration check frequency.

This page runs every step on a synthetic class 1 meter and on the two worked
budgets the standard prints.

![Two panels against frequency at the nine exact octaves from 63 Hz to 16 kHz. Left: the free-field correction of a synthetic meter on a multi-frequency calibrator by Formula (D.7), near 0 dB up to 1 kHz and rising to 0.6 dB at 4 kHz, 1.7 dB at 8 kHz and 5 dB at 16 kHz, a thin band around it for the range of three microphones, and the free-field correction of the reference microphone dashed, above it up to 500 Hz, meeting it at 1 kHz and below it above. Right: the correction of the same meter on an electrostatic actuator by Formula (F.13), zero at 1 kHz by construction and rising to 7.4 dB at 16 kHz, with the normalised sensitivity of the reference channel dashed](https://docs.jmrp.io/phonometry/media/free_field_correction.svg)

*The correction of the synthetic meter on its calibrator (left) and on an
electrostatic actuator (right), from the results' own `.plot()`.*

## 1. The adjustment value at the calibration check frequency

A sound level meter is adjusted in use by applying its calibrator and setting
the display. What the display should read is not the level stated for the
calibrator, $L_1$: the manufacturer first sets the sensitivity so that the
**free-field response** of the meter deviates as little as it can from the
incident level over the whole frequency range, and only then applies the
calibrator and reads $L_4$. The manual states the **adjustment value**
(Annex A)

$$
\Delta L = L_1 - L_4
$$

as a fixed number, without an uncertainty (clause 8), and it can leave a
deliberate offset at the calibration check frequency itself, because the fit
is over the whole range (NOTE 1 of clause 8).

The text asks for the sensitivity that minimises "the averaged deviation" and
says the tolerance limits of IEC 61672-1, which vary with frequency, "shall be
taken into account", but it prints no formula. `metrology.adjustment_value`
takes the least-squares reading: the adjustment $s$ minimises
$\sum_i w_i (d_i + s)^2$, $d_i$ being the deviation at frequency $i$ and
$w_i = 1/t_i^2$ for a tolerance $t_i$, so a band with a tight tolerance pulls
the fit harder than one with a loose one:

$$
s = -\frac{\sum_i w_i d_i}{\sum_i w_i},
\qquad
L_4 = L_4' + s
$$

with $L_4'$ what the meter reads on the calibrator at the sensitivity the
free-field readings were taken at. Here the tolerances are the narrower side
of the class 1 acceptance limits of IEC 61672-1 Table 3, which
`filters.weighting_class_limits` holds:

```python
import numpy as np
from phonometry import filters, metrology

f3 = metrology.exact_frequencies(63, 16000, fraction=3)   # 25 exact one-third octaves
limits = filters.weighting_class_limits(1)
nominal, lower, upper = limits.frequencies, limits.lower, limits.upper  # IEC 61672-1 Table 3, class 1
band = (nominal >= 63) & (nominal <= 16000)
tolerance = np.minimum(upper, -lower)[band]                # the narrower side at each band

x3 = f3 / 1000
response = 0.4 + 0.12 * np.cos(3 * np.log(x3)) - 0.3 * (x3 / 8) ** 2   # indication less field, dB
a = metrology.adjustment_value(
    94.0 + response,           # the meter in a 94 dB free field, before the adjustment
    94.35,                     # L4': the same meter on its calibrator
    frequencies=f3,
    calibrator_level=94.0,     # L1, stated for the calibrator
    tolerance_db=tolerance,
)
print(round(a.sensitivity_adjustment, 2), round(a.adjustment, 2))  # -0.37 0.02
print(round(a.check_frequency_offset_db, 2))                           # 0.15
a.plot()
```

The fit lowers the sensitivity by 0.37 dB, the calibrator then reads
93.98 dB, and the manual states $\Delta L = 0.02$ dB. At 1 kHz the adjusted
meter reads a free field 0.15 dB high, $L_2 - L_1$: that is the offset the
fit chose so that the response sits inside the tolerances everywhere else. A
tolerance of `np.inf` leaves a band out of the fit, and
`pressure_indicated_level=` adds the pressure response, from which
`pressure_to_free_field_correction` and $L_3$ follow.

![The deviation of the synthetic meter's free-field indication from the incident level at the 25 one-third octaves from 63 Hz to 16 kHz, before the adjustment (dashed grey, about +0.3 to +0.5 dB up to 8 kHz and falling to -0.8 dB at 16 kHz) and after it (solid, 0.37 dB lower, within 0.15 dB of zero up to 6.3 kHz and -1.2 dB at 16 kHz), inside the shaded tolerance band of plus or minus 1 dB narrowing to 0.7 dB at 1 kHz and widening to 2.5 dB at 16 kHz, with a dotted line at the calibration check frequency of 1 kHz; the title reads Adjustment value Delta L = L1 - L4 = 0.02 dB](https://docs.jmrp.io/phonometry/media/free_field_adjustment.svg)

*The free-field response before and after the adjustment, inside the class 1
tolerances it was weighed against.*

## 2. How the measurement goes

Every method compares the meter with a **laboratory standard microphone of
type LS2P**, whose free-field correction $C_\mathrm{FF,RM}$ is known from
IEC/TS 61094-7, first in a free progressive field and then on the source. The
comparison is by substitution (Annex G): the meter is measured, removed, and
the reference put in its place, so neither needs an absolute calibration.

![The readings of IEC 62585 in three columns, the free-field ones by substitution. Left, a sound calibrator (Annex D): 1, the meter in a free field reading L_ind1; 2, the reference microphone of type LS2P in its place reading L_ind2; 3, the calibrator on the meter reading L_ind3; 4, the calibrator on the reference reading L_ind4; with stable sources (NOTEs 2 and 3), Formula (D.7) gives (L_ind1 - L_ind3) - (L_ind2 - L_ind4) + C_FF,RM. Middle, a comparison coupler (Annex E): the same two free-field readings, then the reference and the meter face to face on the coupler, read together or in turn as L_ind3a and L_ind3b; with the labels of Figure E.1, a stable source and equal levels in the coupler, Formula (E.6) gives (L_ind1 - L_ind3b) - (L_ind2 - L_ind3a) + C_FF,RM. Right, an electrostatic actuator (Annex F): the two free-field readings as functions of frequency, then the actuator on the meter's diaphragm reading L_ind3(f); by Formula (F.13) every term is referred to f_0, where the correction is zero. The footer says each free-field pair is read in turn at the same place (Annex G) and that C_FF,RM comes from IEC/TS 61094-7](https://docs.jmrp.io/phonometry/media/diagram_free_field_corrections_setup.svg)

| Requirement | Value | Clause |
| :--- | :--- | :--- |
| Environment | 80 kPa to 105 kPa, 20 °C to 26 °C, 25 % to 70 % relative humidity; below 97 kPa an extra component enters the budget | 6 |
| Reference conditions | 23 °C, 101.325 kPa, 50 % relative humidity | 4 |
| Free-field mounting | On a rod of the microphone's diameter; source more than 1 m away and at least six times the largest dimension of the meter; sound in the reference direction | 6 |
| Meter support | A non-vertical rod at the rear, method 1 or 2 of Figure 1 (method 2 at less than 60°); method 3 is not used | 6 |
| Samples | Three calibrators, three microphones and one meter; five microphones for a typical response, five windscreens | 7 |
| Source level | 70 dB to 125 dB in a calibrator or coupler, preferably the reference level at 1 kHz; 80 dB to 105 dB on the display at 1 kHz with an actuator | 12, 13, 14 |
| Frequencies | Octaves from 63 Hz to 16 kHz (class 1) or 8 kHz (class 2) for a microphone; exact base-ten frequencies of Annex H | 10, B.2, C.2 |
| Determinations | Nine for a calibrator (three microphones on three calibrators), three microphones for a coupler, the combinations of clause 7 for an actuator; the correction is their mean | D.2, E.2, F.2 |

## 3. The corrections for a calibrator, a coupler and an actuator

**Annex D, a multi-frequency sound calibrator.** Four readings (Figure D.1):
the meter in a free field, $L_\mathrm{ind1}$; the reference in its place,
$L_\mathrm{ind2}$; the calibrator on the meter, $L_\mathrm{ind3}$; the same
calibrator on the reference, $L_\mathrm{ind4}$. The correction is the meter's
free-field response relative to its response on the calibrator, carried over
from the reference:

$$
C_\mathrm{FF,SLM} = (L_\mathrm{ind1} - L_\mathrm{ind3})
- (L_\mathrm{ind2} - L_\mathrm{ind4})
- (L_{p,\mathrm{F1}} - L_{p,\mathrm{F2}})
+ (L_{p,\mathrm{P1}} - L_{p,\mathrm{P2}}) + C_\mathrm{FF,RM}
\tag{D.7}
$$

where the two level differences are the drift of the free field between the
first two measurements and the difference of the calibrator's level on the
two microphones, both zero for stable sources. Each reading may be a row per
determination, and the result keeps every one:

```python
f = metrology.exact_frequencies(63, 16000, fraction=1)   # the nine exact octaves
x = f / 1000
c_ff_rm = 0.05 * x**1.3                        # C_FF,RM of the reference: illustrative
free = 0.1 * np.sin(np.log(x)) - 0.05 * x**1.2  # the meter in a free field, re the field
pressure = -0.1 * x**1.5                        # the meter on the calibrator, re its level
spread = np.array([[0.0], [0.02], [-0.03]]) * x**0.8   # three microphones of the model

c = metrology.sound_calibrator_correction(
    94.0 + free + spread,          # L_ind1: the meter in the free field
    94.0 + c_ff_rm,                # L_ind2: the reference in its place
    94.0 + pressure + spread / 2,  # L_ind3: the calibrator on the meter
    94.0,                          # L_ind4: the calibrator on the reference
    frequencies=f,
    reference_free_field_correction=c_ff_rm,
)
print(c.correction.round(2))  # [-0.04 -0.09 -0.1  -0.05  0.05  0.23  0.63  1.72  4.95]
print(c.range_db.round(2))       # [0.   0.   0.01 0.01 0.02 0.04 0.08 0.13 0.23]
print(c.clause, c.formula)       # 12 D.7
c.plot()
```

The values of `c_ff_rm` here only stand in for the free-field correction of a
real reference microphone; take yours from IEC/TS 61094-7 or from the
microphone's calibration. The correction climbs above 2 kHz because the
calibrator presses on the diaphragm while a free field also diffracts round
the microphone and the case.

The three rows here are three microphones on one calibrator. D.2 step 6 asks
for nine determinations, three microphones on three calibrators, and clause 12
judges "the range of correction values at each frequency measured using a
sample of three microphones", not the spread between the calibrators. Given
nine rows, `microphones=[1, 1, 1, 2, 2, 2, 3, 3, 3]` says which rows share a
microphone, and `range_db` is then the range of the three microphones' mean
corrections, `microphone_corrections`. Without it every row counts as a
microphone of its own, as the three rows above do.

**Annex E, a comparison coupler.** The meter and the reference face each other
in the two openings of one cavity, so the third measurement reads both at
once. `metrology.comparison_coupler_correction` takes the same free-field
readings and the two readings in the coupler, `slm_coupler_level` and
`reference_coupler_level`, named by what each is of. That naming is on
purpose: Figure E.1, its list of symbols and Table I.1 call the reference's
reading $L_\mathrm{ind3a}$ and the meter's $L_\mathrm{ind3b}$, while Formulas
(E.4) to (E.6) are written the other way round, as (D.5) to (D.7) are for a
calibrator. Read with the figure's labels, the printed (E.6) comes out
$2(\Delta L_\mathrm{P,SLM} - \Delta L_\mathrm{P,RM})$ away from the
correction, the deviations of the two channels' indications from the level in
the coupler. The reference channel reads the output voltage of its microphone
and the method needs no absolute calibration (E.1, E.2), so that difference
carries the two channels' absolute sensitivities and is there at every
frequency; only when both channels read sound pressure level does it shrink to
twice the difference of the two pressure responses. The defect is in the
[errata register](https://docs.jmrp.io/phonometry/reference/errata/).

**Annex F, an electrostatic actuator.** An actuator is not an absolute source,
so its correction is **normalised** to the calibration check frequency
$f_0$, where it is zero, and the absolute response there comes from a sound
calibrator (NOTE 2 of F.2). Every term is referred to $f_0$,
$X_\mathrm{N} = X(f) - X(f_0)$, and the reference contributes its free-field
sensitivity $S_\mathrm{RM}$ and the gain of its channel $G_\mathrm{RC}$:

$$
C_\mathrm{N,FF,SLM} = R_\mathrm{N,ind1} - R_\mathrm{N,ind2}
- (R_\mathrm{N,p,F1} - R_\mathrm{N,p,F2}) + S_\mathrm{N,RM}
+ G_\mathrm{N,RC} - (R_\mathrm{N,ind3} - R_\mathrm{N,EA})
\tag{F.13}
$$

```python
e = metrology.electrostatic_actuator_correction(
    94.0 + free + spread,                  # L_ind1(f)
    94.0 - 26.0 + c_ff_rm,                 # L_ind2(f): the reference channel, dB re 1 V
    94.0 - 0.08 * x**1.7 + spread / 2,     # L_ind3(f): the actuator on the meter
    frequencies=f,
    reference_sensitivity_level=-26.0 + c_ff_rm,   # S_RM(f), dB re 1 V/Pa
)
print(e.correction.round(2))  # [-0.07 -0.12 -0.13 -0.09  0.    0.18  0.64  2.16  7.39]
```

The drive voltage of the actuator is taken to be independent of frequency, so
its simulated level is the same at $f$ and $f_0$ (NOTE 4); pass
`actuator_level=` when it is not, and `reference_channel_gain_db=` for a
reference channel that is not flat.

## 4. The uncertainty budget of Annex I

Table I.1 lists fifteen likely components of the uncertainty of a correction
measured with a comparison coupler, each with the distribution that sets its
divisor: the half-width of a rectangular distribution over $\sqrt{3}$ for most,
the expanded uncertainty of $C_\mathrm{FF,RM}$ from IEC/TS 61094-7 over 2, and
the repeatability from repeat measurements over 1, with its degrees of freedom.
Tables I.2 and I.3 fill it in at 1 kHz and 8 kHz.
`metrology.correction_uncertainty_budget` builds the budget on
`metrology.uncertainty_propagation`, every component a correction entering
Formula (E.6) with a sensitivity of $\pm 1$, and takes the coverage factor
from the Welch-Satterthwaite effective degrees of freedom at a level of
confidence of 95 % (clause 5):

$$
u_\mathrm{c}^2 = \sum_{i=1}^{15} u_i^2,
\qquad
\nu_\mathrm{eff} = \frac{u_\mathrm{c}^4}{\sum_i u_i^4/\nu_i},
\qquad
U = t_{0.975}(\nu_\mathrm{eff})\, u_\mathrm{c}
$$

```python
table_i2 = {
    "a1": 0.005, "a2": 0.005, "a3": 0.005, "a4": 0.005,   # the four level readings
    "a5": 0.05, "a6": 0.0,                                  # the two level differences
    "a7": 0.06,                                             # U of C_FF,RM (k = 2)
    "a8": 0.025, "a9": 0.025, "a10": 0.029,                 # gains and distance
    "a11": 0.013, "a12": 0.013, "a13": 0.0, "a14": 0.005,   # field, mountings, diameters, rounding
    "a15": 0.03,                                            # repeatability (k = 1)
}
b1 = metrology.correction_uncertainty_budget(table_i2, repeatability_dof=2, frequency_hz=1000)
print(round(b1.combined_uncertainty_db, 4), round(b1.effective_dof, 2))   # 0.059 29.98
print(round(b1.coverage_factor, 2), round(b1.expanded_uncertainty_db, 2))  # 2.04 0.12

table_i3 = {**table_i2, "a7": 0.17, "a11": 0.104, "a12": 0.104, "a15": 0.06}
b8 = metrology.correction_uncertainty_budget(table_i3, repeatability_dof=2, frequency_hz=8000)
print(round(b8.combined_uncertainty_db, 3), round(b8.coverage_factor, 2),
      round(b8.expanded_uncertainty_db, 2))                                # 0.14 2.0 0.28
```

Both tables reproduce: 0.0590 dB and 29.98 degrees of freedom at 1 kHz,
0.140 dB, $k = 2$ and 0.28 dB at 8 kHz. Table I.2 prints $k = 2.11$, though,
and that is not the factor its own 29.98 degrees of freedom give. At 95 % the
Student factor for 29.98 is 2.042; 2.11 is the one for about 17. The expanded
uncertainty the page's numbers give is $2.042 \times 0.059\,03 = 0.121$
dB, printed 0.12 either way, but the table's guard digit, 0.12(4), is the
misprinted $k$ at work; the case is in the [errata
register](https://docs.jmrp.io/phonometry/reference/errata/). Below 97 kPa,
`static_pressure_kpa=` adds the component of clause 6 (0.15 dB up to 3 kHz,
0.25 dB above, $k = 2$), and `additional_components=` takes any component a
laboratory's own method needs, such as the drive of an actuator.

![Two bar charts of the standard uncertainty of each of the fifteen components of IEC 62585 Table I.1, labelled a1 to a15. Left, at 1000 Hz: the largest bars are a7, the free-field correction of the reference microphone, and a15, the repeatability, both 0.030 dB, then a5 at 0.029 dB; the combined standard uncertainty 0.0590 dB is marked dashed, and the title reads nu_eff = 29.98, k = 2.04, U = 0.121 dB. Right, at 8000 Hz: a7 grows to 0.085 dB and a11 and a12, the free field and the mountings, to 0.060 dB each, with the repeatability 0.060 dB; the combined standard uncertainty is 0.1400 dB and the title reads nu_eff = 59.33, k = 2.00, U = 0.280 dB. The repeatability bar is drawn in a second colour as the one Type A component](https://docs.jmrp.io/phonometry/media/free_field_uncertainty.svg)

*The budgets of Tables I.2 and I.3, one bar per component, the statistical one
apart.*

## 5. The maximum permitted uncertainty, and the verdict

Each clause states the largest expanded uncertainty a correction may carry
and still be used for the manual (clause 5); a measurement that exceeds one is
not used. The range is inclusive: "shall not exceed" lets a value equal to its
maximum pass, and one that reaches it through floating-point arithmetic too,
by the same rule as the IEC TC 29 conformance verdict.

| Clause | Correction for | Maximum expanded uncertainty |
| :--- | :--- | :--- |
| 9 | Reflections from the case, diffraction round the microphone | 0.25 dB up to and including 4 kHz; 0.35 dB above |
| 10 | The microphone's deviation from a uniform response | 0.25 dB from 63 Hz to 4 kHz; 0.35 dB to 8 kHz; 0.45 dB above |
| 11 | Windscreens and similar accessories | 0.20 dB up to and including 4 kHz; 0.30 dB above |
| 12, 13, 14 | A calibrator, a coupler, an actuator | 0.25 dB up to and including 4 kHz; 0.35 dB below 10 kHz; 0.50 dB from 10 kHz |

Clauses 10 and 11 exclude the variability of the samples of microphone or
accessory. For a calibrator, a coupler or an actuator, clauses 12 to 14 add a
second requirement on the microphone: if the range of the corrections measured
with three microphones exceeds the maximum at a frequency, the microphone is
unsuitable for that source unless more samples show otherwise.
`metrology.verify_correction_uncertainty` judges both, and a frequency within
2 % of a boundary reads as it, so the exact 3 981 Hz of the 4 kHz band is "up
to and including 4 kHz". Here each octave of the calibrator's correction gets a
budget like Table I.2 below 1 kHz, with a reference, a field and a
repeatability that grow above it:

```python
print(metrology.max_expanded_uncertainty(f, clause=c.clause))
# [0.25 0.25 0.25 0.25 0.25 0.25 0.25 0.35 0.5 ]

a7 = [0.06, 0.06, 0.06, 0.06, 0.06, 0.08, 0.10, 0.17, 0.30]     # U of C_FF,RM, k = 2
a11 = [0.013, 0.013, 0.013, 0.013, 0.013, 0.03, 0.06, 0.104, 0.18]  # field and mountings
a15 = [0.03, 0.03, 0.03, 0.03, 0.03, 0.04, 0.05, 0.06, 0.09]    # repeatability
budgets = [
    metrology.correction_uncertainty_budget(
        {**table_i2, "a7": u7, "a11": u11, "a12": u11, "a15": u15},
        repeatability_dof=2,
        frequency_hz=fx,
        correction=cx,
    )
    for fx, u7, u11, u15, cx in zip(f, a7, a11, a15, c.correction, strict=True)
]
v = metrology.verify_correction_uncertainty(
    [b.expanded_uncertainty_db for b in budgets],
    frequencies=f,
    clause=c.clause,
    correction=c.correction,
    coverage_factor=[b.coverage_factor for b in budgets],
    correction_range_db=c.range_db,
)
print(v.passes)              # True
print(v.margin_db.round(2))  # [0.13 0.13 0.13 0.13 0.13 0.1  0.06 0.07 0.04]
v.plot()
v.report("free-field-corrections.pdf")
```

The verdict is the object, not a boolean: `v.passes` is the answer,
`failing_frequencies_hz` says where it fails, and `bool(v)` raises rather than
let `if verify_correction_uncertainty(...):` pass every measurement. The
tightest margin is at 16 kHz, 0.04 dB under the 0.50 dB of clause 12.

![The expanded uncertainty of the calibrator's corrections at the nine exact octaves from 63 Hz to 16 kHz, 0.12 dB up to 1 kHz and rising to 0.28 dB at 8 kHz and 0.46 dB at 16 kHz, under the stepped maximum of clause 12, 0.25 dB up to 4 kHz, 0.35 dB above and 0.50 dB from 10 kHz, with the range of the corrections over the three microphones below both, from 0.003 dB to 0.23 dB; the title reads Clause 12: within the maximum at every frequency](https://docs.jmrp.io/phonometry/media/free_field_verification.svg)

*The expanded uncertainty and the range over the microphones, under the
maximum of clause 12.*

`.report()` renders what clause 15 n) and o) ask the laboratory to issue: the
corrections with their expanded uncertainties and coverage factors, and a
statement of whether the uncertainties are within the maximum permitted
values. The example fiche is regenerated with `make reports` and kept rendered
in the repository; click the preview to open the PDF.

**Report:** [IEC 62585 free-field corrections example report (PDF)](https://docs.jmrp.io/phonometry/media/iec62585_free_field_correction_example.pdf): One-page fiche for the corrections of a class 1 sound level meter on its multi-frequency calibrator: a metadata header, the basis line citing IEC 62585:2012 clause 12, a table of the nine exact octaves from 63.1 Hz to 15 848.9 Hz with the correction, the expanded uncertainty, the coverage factor, the maximum, the range over three microphones and the verdict at each, the verification plot beside it, the boxed statement Within the maximum permitted values of clause 12 at all 9 frequencies with the largest U / Umax of 0.92, and a PASS verdict.

Free-field correction fiche (CorrectionUncertaintyVerification.report): the corrections with their uncertainty against the maximum of clause 12.

## 6. Exact frequencies

The corrections are measured and reported at **exact** base-ten frequencies,
not at the nominal ones (clause 10, B.2, C.2). Annex H gives them by Formula
(H.1),

$$
f_x = f_\mathrm{r}\, 10^{3x/(10b)},
\qquad f_\mathrm{r} = 1000\ \mathrm{Hz}
$$

for every integer $x$, with $b = 12$ for one-twelfth-octave steps; Table H.1
prints the decade from 1 kHz to 10 kHz to seven significant digits.
`metrology.exact_frequencies` evaluates it for any designator, and the octaves
and thirds above came from it:

```python
h = metrology.exact_frequencies(1000, 10000)   # the 41 rows of Table H.1
print(h.size, (h[[1, 31, 40]] / 1000).round(6))   # 41 [ 1.059254  5.956621 10.      ]
```

All 41 values reproduce. The exponent column prints $10^{31/80}$ for index 31,
where its own value is $10^{31/40}$; the library evaluates the formula and never
reads that column.

**Show the code for these figures**

```python
import matplotlib.pyplot as plt

fig, (ax_calibrator, ax_actuator) = plt.subplots(1, 2, figsize=(13.5, 5.6))
c.plot(ax_calibrator)
e.plot(ax_actuator)
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
a.plot(ax)
fig.tight_layout()

fig, (ax_1k, ax_8k) = plt.subplots(1, 2, figsize=(13.5, 6.4))
b1.plot(ax_1k)
b8.plot(ax_8k)
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
v.plot(ax)
fig.tight_layout()
plt.show()
```

## What this guide covers

- **Covered.** **IEC 62585:2012, the calculations of Annexes A, D, E, F, H and I:** the
  adjustment value of Annex A, with the pressure-to-free-field correction and
  the levels $L_1$ to $L_4$ of Figure A.1; the free-field corrections for a
  multi-frequency sound calibrator (Formula (D.7)), a comparison coupler
  (Formula (E.6), with the readings named by what they are of) and an
  electrostatic actuator (Formula (F.13), normalised to the calibration check
  frequency), averaged over the determinations with their range over the
  microphones kept; the uncertainty budget of Annex I with the fifteen
  components of Table I.1, the static-pressure component of clause 6 and the
  Welch-Satterthwaite coverage factor; the maxima of clauses 9 to 14 and the
  range requirement of clauses 12 to 14 as a verdict and a fiche; and the exact
  frequencies of Annex H. Tables I.2, I.3 and H.1 are reproduced.

- **Covered, by a reading of the text.** **The fit of Annex A.** The standard asks for the sensitivity that minimises
  the averaged deviation with the tolerances taken into account, and prints no
  formula; the least-squares fit weighted by $1/t^2$ is this library's reading,
  and a manufacturer optimising another criterion states another adjustment
  value.

- **Not covered.** The measurements themselves: the free field, the mounting of the meter and the
  microphone, the time-selective and steady-state techniques of Annex G, and the
  free-field correction of the reference microphone, which comes from IEC/TS
  61094-7 or its calibration and is an input here. The corrections of clauses 9
  to 11 are judged against their maxima but not computed: the case and
  diffraction corrections of Annex B and the windscreen and accessory
  corrections of Annex C, each the difference between two set-ups; the typical
  microphone response of clause 10, the mean over at least five microphones
  with its maximum deviation; and the mean over at least four rotational
  positions that clauses 10 and 11 ask for when the reference direction is not
  the principal axis. The comparison of the supplied corrections with the
  acceptance limits of IEC 61672-1, which clause 15 p) asks for separately, and
  the periodic test of IEC 61672-3 that uses the corrections, are not
  implemented on this page; the verdict on that periodic test is in [Periodic
  tests of a sound level
  meter](https://docs.jmrp.io/phonometry/signals/metrology/sound-level-meter-periodic-tests/).

## See also

- [Random-incidence and diffuse-field response](https://docs.jmrp.io/phonometry/signals/metrology/random-incidence/):
  IEC 61183, the response of the meter to sound from every direction, which
  this standard leaves to it for a random-incidence microphone.
- [Measurement uncertainty](https://docs.jmrp.io/phonometry/signals/metrology/gum-uncertainty/):
  the GUM law of propagation and the Welch-Satterthwaite degrees of freedom
  the budget is built on.
- [Compliance and verification](https://docs.jmrp.io/phonometry/signals/metrology/compliance-verification/):
  what the IEC 61672-1 class of a meter asserts, and which of its tests the
  library runs.
- [Calibration and dBFS](https://docs.jmrp.io/phonometry/signals/metrology/calibration/): the
  calibrator tone that sets the pressure sensitivity every level of a meter
  starts from.
- API reference: [`metrology.free_field_corrections`](https://docs.jmrp.io/phonometry/reference/api/metrology/free-field-corrections/).

## References

- **International Electrotechnical Commission. (2006).** *Measurement microphones — Part 7: Values for the difference between free-field and pressure sensitivity levels of laboratory standard microphones* (IEC/TS 61094-7:2006).
  The free-field correction of the reference microphone, C_FF,RM, every method takes as an input. The values on this page are illustrative, not read from it.
- **International Electrotechnical Commission. (2012).** *Electroacoustics — Methods to determine corrections to obtain the free-field response of a sound level meter* (IEC 62585:2012).
  The implemented methods: the adjustment value of Annex A, the corrections of Formulas (D.7), (E.6) and (F.13), the uncertainty budget of Annex I (Tables I.1 to I.3), the maximum expanded uncertainties of clauses 9 to 14, the static-pressure component of clause 6 and the exact frequencies of Annex H (Table H.1). Read from BS EN 62585:2012, the English text of EN 62585:2012, which is IEC 62585:2012 unchanged.
- **International Electrotechnical Commission. (2013).** [*Electroacoustics — Sound level meters — Part 1: Specifications*](https://webstore.iec.ch/en/publication/5708) (IEC 61672-1:2013).
  The acceptance limits of Table 3 the example weighs the Annex A fit with, and the frequency weightings the corrections serve.
- **Joint Committee for Guides in Metrology. (2008).** [*Evaluation of measurement data — Guide to the expression of uncertainty in measurement*](https://www.bipm.org/documents/20126/2071204/JCGM_100_2008_E.pdf) (JCGM 100:2008, the GUM). BIPM. https://doi.org/10.59161/JCGM100-2008E
  The law of propagation of uncertainty and the Welch-Satterthwaite effective degrees of freedom (Annex G.4) the budget of Annex I is built on. Also published as ISO/IEC Guide 98-3:2008.