# Microphone Calibration by Comparison (IEC 61094-5/-8)

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/signals/metrology/comparison-calibration/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/signals/metrology/comparison-calibration/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-10-01
Updated: 2026-11-16T21:36:14Z
Standards: IEC 61094-5:2016; IEC 61094-8:2012; NPL Report CIRA(EXT) 010; IEC 61094-2:2009; IEC 61094-1:2000
Verified: 51 conformance checks (IEC 61094-5:2016; IEC 61094-8:2012; IEC 61094-2:2009) https://docs.jmrp.io/phonometry/reference/conformance/
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Calibration by comparison (IEC 61094-5/-8): a reference sensitivity carried to a working standard in level and phase, corrections and uncertainty.

---

A working standard microphone is not calibrated by reciprocity. It is put
beside a reference microphone whose sensitivity is already known, or in its
place, both are exposed to the same sound pressure, and the ratio of their
output voltages carries the reference's sensitivity over to it. IEC 61094-5
does this in a **pressure field**, in a coupler or a jig; IEC 61094-8 does it
in a **free field**, in an anechoic room or behind a time window. Annex D of
the first part writes the model once for both,

$$
M_\mathrm{test} = M_\mathrm{ref}\,\frac{R_V}{R_P},
\qquad
L_\mathrm{test} = L_\mathrm{ref} + 20\lg R_V - 20\lg R_P + \sum_j C_j
$$

where $R_V$ is the ratio of the output voltages of the test and reference
microphones, $R_P$ that of the effective sound pressures acting on them, which
the procedure reduces to unity, and $C_j$ the corrections the part asks for.
`metrology.comparison_calibration` computes the right-hand form, in
sensitivity levels in dB re 1 V/Pa.

This page runs a pressure calibration of a WS2P against an LS2P and a
free-field calibration of a WS2F, with their budgets, then the time-selective
processing of IEC 61094-8 Annex B, the phase of the sensitivity, the effect
of different acoustic impedances (IEC 61094-5 7.4 and 7.5) with the air
between two microphones by the circuit the part refers to, the correction
for microphones of different diameters by the model it refers to (6.5), and
the validation of a jig or a coupler (6.7); the readings are synthetic, built
from sensitivities the page knows, so every step can be checked against them.

![Two panels of sensitivity level in dB re 1 V/Pa against frequency. Left, the pressure calibration by IEC 61094-5 from 250 Hz to 20 kHz: the LS2P reference dashed, near -38.0 dB and falling to -38.2 dB at 20 kHz, and the WS2P under test solid, rising from -38.64 dB at 250 Hz to -38.34 dB at 16 kHz and -38.43 dB at 20 kHz, inside a shaded band of plus or minus its expanded uncertainty, 0.09 dB up to 8 kHz, 0.10 dB at 10 kHz, widening to 0.2 dB at 20 kHz. Right, the free-field calibration by IEC 61094-8 from 500 Hz to 20 kHz: the reference's free-field level dashed, its pressure level plus its free-field difference, rising from -37.99 dB at 500 Hz to -37.26 dB at 8 kHz and -35.75 dB at 20 kHz, and the WS2F under test solid from -38.33 dB to -38.23 dB at 8 kHz and -38.47 dB at 20 kHz, inside a band of plus or minus 0.16 dB widening to 0.51 dB at 20 kHz](https://docs.jmrp.io/phonometry/media/comparison_calibration.svg)

*A pressure calibration by simultaneous excitation (left) and a free-field
calibration by substitution (right), from the results' own `.plot()`.*

## 1. How the measurement goes

Both parts give two ways of exposing the microphones to the same sound
pressure. With **simultaneous excitation** they sit in the field at once,
each read on its own measuring channel; in a coupler or a jig they face each
other about 1 mm apart, a small fraction of the wavelength up to 20 kHz
(5.1.2). With **sequential excitation** they take the same place in turn:
either the exchange does not change the sound pressure significantly, or any
change is detected and corrected, for example with a monitor microphone near
the source (5.1.3; IEC 61094-8 5.2, 6.5 and A.2).

![The readings of a comparison calibration in two columns. Left, a pressure field with simultaneous excitation (IEC 61094-5, Annex C): 1, the reference microphone and the microphone under test face to face in a coupler, the reference on channel 1, reading L_C12; the two are then interchanged in the ports and on the preamplifiers; 2, the microphone under test on channel 1, reading L_C21; Formula (C.3) gives L_test = L_ref - half of (L_C12 - L_C21), in which the gains of the channels and the asymmetry of the field cancel. Right, a free field with sequential excitation (IEC 61094-8, Annex A): 1, a loudspeaker with a monitor microphone near it and the reference microphone at the measuring point, with output voltages V_ref and V_mon,1; 2, the microphone under test at the same point, with output voltages V_test and V_mon,2; by A.2, L_test = L_ref + 20 lg(V_test / V_mon,2) - 20 lg(V_ref / V_mon,1), the monitor taking out any drift of the source. The footer says the corrections (environment, the WS3 jig, the reference's free-field difference) and U with k = 2 follow](https://docs.jmrp.io/phonometry/media/diagram_comparison_calibration_setup.svg)

| Requirement | Value | Clause |
| :--- | :--- | :--- |
| Reference conditions | 23.0 °C, 101.325 kPa, 50 % relative humidity | 61094-5 and 61094-8, 4 |
| Simultaneous, in a coupler or a jig | Diaphragms about 1 mm apart; the microphones interchanged and the measurement repeated, the result from the two | 61094-5, 5.1.2 |
| Sequential | The exchange does not change the sound pressure significantly, or any change is detected and corrected, for example with a monitor microphone; a monitor, when used, senses the changes at the test position | 61094-5, 5.1.3; 61094-8, 5.2, 6.5 |
| Analysis bandwidth | 1/6 octave or narrower keeps the error below 0.01 dB; much narrower FFT lines can show standing waves | 61094-5, 5.1.1 NOTE |
| Output voltage | Open-circuit, by insert voltage or a high-impedance preamplifier; the method is stated on the certificate | 61094-5, 5.2 |
| Free-field source | 70 dB to 80 dB at the microphone is usually enough; plane progressive waves across the microphone | 61094-8, 6.3 |
| Free-field mounting | On a semi-infinite rod of the microphone's diameter; LS1P references without their grid | 61094-8, 6.4, 6.7 |
| Reported uncertainty | Expanded, with $k = 2$, at each frequency | 61094-5, 7.9; 61094-8, 8.8 |

## 2. A pressure calibration by simultaneous excitation

In a coupler the two microphones are read on two channels whose gains are
never exactly equal, and the two ports of a coupler never quite see the same
pressure. Annex C removes both. With the reference as microphone 1 on channel
1, the level reading difference between the channels is Formula (C.1); after
the interchange it is Formula (C.2); and their difference leaves only the two
sensitivity levels,

$$
L_\mathrm{C12} = (L_1 + L_\mathrm{m1} + L_\mathrm{d1} + L_\mathrm{WA})
- (L_2 + L_\mathrm{m2} + L_\mathrm{d1} + L_\mathrm{WB}),
\qquad
L_1 - L_2 = \tfrac12\,(L_\mathrm{C12} - L_\mathrm{C21})
\tag{C.3}
$$

so $20\lg R_V = -\tfrac12(L_\mathrm{C12} - L_\mathrm{C21})$. Here the
channels differ by 0.55 dB, port A hears up to 0.13 dB more than port B,
and each reading is repeated three times with a little noise:

```python
import numpy as np
from phonometry import metrology

f = metrology.exact_frequencies(250, 20000, fraction=3)   # 20 exact one-third octaves
x = f / 1000
l_ref = -38.0 + 0.04 * np.log10(x) - 0.25 * (x / 20) ** 2   # the LS2P's certificate, dB re 1 V/Pa
l_true = -38.6 + 0.08 * np.log10(x) + 0.2 * (x / 12) ** 2 - 0.5 * (x / 20) ** 4   # the WS2P: unknown

gain_1, gain_2 = 0.35, -0.20      # the two measuring channels, dB
field_a = 0.03 * np.sqrt(x)       # port A of the coupler hears 0.015 dB to 0.13 dB more
rng = np.random.default_rng(61094)
noise = rng.normal(0.0, 0.004, (2, 3, f.size))   # three repeats of each reading
l_c12 = (l_ref + gain_1) - (l_true + gain_2) + field_a + noise[0]   # reference on channel 1
l_c21 = (l_true + gain_1) - (l_ref + gain_2) + field_a + noise[1]   # interchanged

c = metrology.simultaneous_comparison(l_ref, l_c12, l_c21, frequencies=f)
print(c.n_determinations, c.sensitivity_level[[2, 8, 19]].round(2))   # 3 [-38.63 -38.58 -38.44]
print(np.abs(c.sensitivity_level - l_true).max().round(3))           # 0.003
```

Neither the gains nor the asymmetry reach the result; what is left, 0.003 dB
at most, is the noise of three repeats. 5.1.2 makes the interchange a
requirement for a coupler or a jig, so a pressure calibration without
`interchanged_channel_difference` is refused. The text asks for "the mean
of the two ratios", and Annex C takes it in decibels: the mean of the two
level differences, which is the geometric mean of the ratios and the only one
in which the gains cancel exactly. The two ratios read before and after the
interchange are 1.1 dB to 1.4 dB apart here, twice the difference of the
channels and of the ports, and their arithmetic mean would sit 0.018 dB to
0.027 dB high.

**The environment.** A microphone's sensitivity depends on the static
pressure, the temperature and, for some microphones, the humidity. When the
reference and the test microphone are different models, the reference's
sensitivity has to be corrected to the conditions of the test (6.6); in a
free field, IEC 61094-8 asks for that correction whatever the models (7.6).
Neither part prints a formula. `metrology.environmental_sensitivity_correction`
takes the first-order one, a coefficient times a deviation for each
condition, with the coefficients in the units IEC 61094-2 Annex D gives them:

$$
C_\mathrm{env} = \delta_p\,(p_s - p_{s,0}) + \delta_t\,(t - t_0)
+ \delta_H\,(H - H_0)
$$

The coefficients are the microphone's own, which is why they are required:
Annex D of IEC 61094-2 puts the low-frequency static pressure coefficient of
LS2P microphones between −0.003 dB/kPa and −0.008 dB/kPa and the temperature
coefficient within ±0.005 dB/K, and both vary with frequency and from one
microphone to the next. 6.5.3 of the same part observes no influence of
humidity on a laboratory standard microphone, so that coefficient defaults
to 0.

```python
env = metrology.environmental_sensitivity_correction(
    f,
    static_pressure_kpa=99.2,        # the laboratory during the test
    temperature_c=21.5,
    relative_humidity_percent=45.0,
    static_pressure_coefficient_db_per_kpa=-0.005,   # the LS2P's own coefficients
    temperature_coefficient_db_per_k=0.002,
)   # from its certificate's 101.325 kPa, 23.0 °C and 50 %
print(env.correction[0].round(4))   # 0.0076
```

Passed as `reference_environment=`, the correction is added to the
reference's level; passed as `test_environment=`, the test microphone's own
correction is subtracted from the result, which refers it to the reference
conditions of clause 4 when reliable coefficients are known.

## 3. The uncertainty budget of Annex D

Table D.1 is a worked budget for a set-up like this one, a low-sensitivity
WS2P against an LS2P in the coupler of Figure A.1 with three repeats, at
2 kHz. That coupler is for frequencies up to 10 kHz (A.1); above, a WS2 goes
in the jig of A.2 and Figure A.3, which reaches 20 kHz. Seven of the eight
rows state a value and how it is read: the reference's certificate quotes
±0.05 dB with $k = 2$, the capacitance, non-linearity, impedance, drift and
rounding are semi-ranges of rectangular distributions, and the polarising
voltage of (200.0 ± 0.2) V gives a semi-range of
$20\lg(200.2/200) = 0.008\,68$ dB. The repeatability row states no
value, only that it was found from the standard uncertainties of a large
number of similar measurements, and prints 0.025 dB as a standard
uncertainty. `metrology.IEC61094_5_TABLE_D1` holds each row with its stated
value, its divisor and the standard uncertainty it prints, and the seven that
state a value reproduce it. `metrology.comparison_uncertainty_budget` combines them on
`metrology.uncertainty_propagation`, every component with a sensitivity of 1 in
the level model, and multiplies by $k = 2$:

```python
d1 = {
    "reference": 0.025, "capacitance": 0.006, "non_linearity": 0.017,
    "impedance": 0.003, "polarizing_voltage": 0.005, "repeatability": 0.025,
    "drift": 0.017, "rounding": 0.003,
}   # Table D.1, standard uncertainties at 2 kHz
b = metrology.comparison_uncertainty_budget(d1, frequency_hz=2000)
print(round(b.combined_uncertainty_db, 4), round(b.expanded_uncertainty_db, 3))   # 0.0437 0.087
print(round(b.linear_combined_uncertainty_db, 4))                                  # 0.0436
```

The root-sum-square of the eight printed values is 0.0437 dB, and the
expanded uncertainty 0.087 dB. D.3 prints **0.040 dB** and **0.08 dB**,
which are not the combination of its own components: 0.040 dB is what the
sum gives with one of the two 0.017 dB rows left out. The case is in the
[errata register](https://docs.jmrp.io/phonometry/reference/errata/). D.3 also notes that a
strict calculation would convert each component to linear form before
combining, with essentially the same result, and prints no value for it;
`linear_combined_uncertainty_db` does it and gives 0.043 614 dB, 0.000 055 dB
below the combination in decibels. The Spanish adoption, UNE-EN
61094-5:2017, leaves six of the eight values of Table D.1 blank and prints
0.004 dB in D.3, which is also in the register.

At other frequencies the components change. 7.9 estimates about 0.1 dB at
low and middle frequencies for microphones of the same diameter, rising to
about 0.2 dB at 20 kHz for WS2P, and 7.4 and 7.5 name the difference of the
microphones' acoustic impedances as what grows. Here that component rises
with frequency, and each budget's expanded uncertainty goes to the
calibration, which the plot draws as a band:

```python
budgets = [
    metrology.comparison_uncertainty_budget(
        {**d1, "impedance": 0.003 + 0.09 * (fx / 20000) ** 2}, frequency_hz=fx
    )
    for fx in f
]
u = [bb.expanded_uncertainty_db for bb in budgets]
c = metrology.simultaneous_comparison(
    l_ref, l_c12, l_c21, frequencies=f, reference_environment=env, expanded_uncertainty_db=u
)
print(np.round(u, 3)[[0, 16, 19]])                  # [0.087 0.101 0.205]
print(c.sensitivity_mv_per_pa[[2, 8, 19]].round(2))  # [11.72 11.78 11.98]
c.plot()
```

![Two bar charts of the standard uncertainty of each component. Left, IEC 61094-5 Table D.1 at 2000 Hz: reference microphone and repeatability 0.025 dB, non-linearity and drift of the reference 0.017 dB, capacitance 0.006 dB, polarizing voltage 0.005 dB, impedance and rounding 0.003 dB, with the combined standard uncertainty 0.0437 dB dashed and the title k = 2, U = 0.087 dB. Right, a free-field budget with the components of IEC 61094-8 Table 2 at 8000 Hz: the reference microphone 0.116 dB, the free-field quality 0.049 dB and the others 0.02 dB or less, with u_c = 0.1317 dB and U = 0.263 dB](https://docs.jmrp.io/phonometry/media/comparison_budget.svg)

*The budget of Table D.1 at 2 kHz (left) and a free-field budget at 8 kHz
(right).*

## 4. A WS3 microphone in the jig

A WS3 microphone is smaller than the LS2 reference, and in the jig of Figure
A.4 the two diaphragms face each other 0.5 mm apart: the pressure over the
smaller diaphragm is not the mean over the larger one. Table A.1 gives the
corrections to add to the WS3's sensitivity level, calculated for an LS2aP
reference and a radially symmetrical field and valid for that separation
only, with an expanded uncertainty of a tenth of each correction.
`metrology.jig_diameter_correction` reads them, and an exact base-ten
frequency reads as its row:

```python
j = metrology.jig_diameter_correction()
print(j.correction[[0, 9, 13]])          # [-0.004 -0.235 -1.443]
print(j.standard_uncertainty_db[13].round(4))   # 0.0722
```

`j.correction` goes to a calibration through `corrections=`, under a
name of its own, and `j.standard_uncertainty_db` to its budget as an
additional component, the first special case of Table D.1.

![The corrections of IEC 61094-5 Table A.1 against frequency from 1 kHz to 20 kHz, -0.004 dB at 1 kHz, -0.059 dB at 4 kHz, -0.235 dB at 8 kHz, -0.933 dB at 16 kHz and -1.443 dB at 20 kHz, with a shaded band of plus or minus a tenth of each correction around them](https://docs.jmrp.io/phonometry/media/comparison_jig_correction.svg)

*Table A.1 with its expanded uncertainty of 10 %.*

### Beyond Table A.1: microphones of different diameters

"The effect of a non-uniform pressure distribution over the surface of the
diaphragm will be significantly greater if the test and reference
microphones are of different diameters. A theoretical model which can be
used to apply corrections and assess the uncertainties in this case is given
in the literature (for example [1])" (6.5). Its reference [1], Barham,
Barrera-Figueroa and Avison (2014), is where Table A.1 comes from. The gap
between the diaphragms is a cylinder of the reference's front-cavity radius
$a$ and length $L$, closed by the reference at $z = 0$ and driven, in a
radially symmetrical field, through the annulus round the smaller
microphone at $z = L$. Its pressure is a series of the radial modes of the
cavity,

$$
p(r, z) = \sum_n C_n \cosh(\mu_n z)\, J_0(k_n r/a),
\qquad \mu_n^2 = (k_n/a)^2 - (\omega/c)^2,
\qquad J_1(k_n) = 0
$$

and each microphone reads it averaged over its diaphragm with its radial
sensitivity, the deflection of a membrane under a uniform pressure,
$S(r) = J_0(j_{01}\varpi r/a)/J_0(j_{01}\varpi) - 1$ with $\varpi$ the
frequency over the diaphragm's resonance frequency: the reference over $a$
at its face, the test microphone, with $b$ in place of $a$, over its
diaphragm across the gap. The correction is the ratio of the two averages,
$-20\lg\lvert R_P\rvert$. `metrology.diameter_sound_field_correction`
computes it from the geometry and the two resonance frequencies, at the
speed of sound of the IEC 61094-2 Annex F air:

```python
ws3 = {
    "reference_radius_m": 4.650e-3,        # an LS2: its front cavity and diaphragm
    "test_diaphragm_radius_m": 2.065e-3,   # a WS3: its diaphragm
    "test_outer_radius_m": 2.975e-3,       # and its outline
    "separation_m": 0.5e-3,                # Figure A.4
    "reference_resonance_frequency_hz": 22000,
    "test_resonance_frequency_hz": 100000,
}   # the inputs of Table 1 of [1]
model = metrology.diameter_sound_field_correction(j.frequencies, **ws3)
print(round(model.speed_of_sound, 1), model.correction[[0, 9, 13]].round(3))   # 345.9 [-0.004 -0.233 -1.434]
print(np.abs(model.correction - j.correction).max().round(4))               # 0.0093
```

At the reference air of clause 4 the model is within 0.0093 dB of every row
of Table A.1, inside the 0.144 dB the NOTE gives the largest correction as
its expanded uncertainty. The paper prints no speed of sound; at 344.8 m/s
the model gives every row to the thousandth printed. As printed, the paper's
formulas with its own inputs do not give its own table: the library reads
the radial sensitivity of its Formula (4) less 1 and the annulus of its
Formula (2) from the WS3's overall radius, the two readings that do (an
[erratum](https://docs.jmrp.io/phonometry/reference/errata/)). A.2 sets the expanded uncertainty
of Table A.1 at 10 % of each correction, "which is approximately the change
observed by doubling the distance between the microphones";
`separation_change_db` is that change by the model, of the order A.2 states
and below it:

```python
print((model.separation_change_db / np.abs(model.correction))[[0, 13]].round(3))   # [0.063 0.054]
```

Table A.1 holds for one pair of microphones at the one separation of Figure
A.4, "the only one for which the corrections specified in Table A.1 are
valid", and at fourteen frequencies. The model holds for any test microphone
that fits inside its reference's front cavity, any separation, and any
frequency below the two resonance frequencies, where the radial sensitivity
it divides by $J_0(j_{01}\varpi)$ is defined:

```python
f12 = metrology.exact_frequencies(1000, 20000, fraction=12)   # every twelfth of an octave
fine = metrology.diameter_sound_field_correction(f12, **ws3)
print(f12.size, round(f12[50]), fine.correction[50].round(3))   # 53 17783 -1.141
```

`fine.correction` goes to a calibration through `corrections=` as
Table A.1 does. The model assumes a radially symmetrical field, which is why
6.5 and A.2 ask for a coaxial source in the far field or a diffuse field.

![The correction of a WS3 microphone against an LS2 by the model of IEC 61094-5 6.5, at every twelfth of an octave from 1 kHz to 20 kHz, from -0.004 dB at 1 kHz to -0.233 dB at 8 kHz and -1.427 dB just under 20 kHz, with the fourteen rows of Table A.1 as diamonds on it, and dashed above it the correction at twice the separation, 1 mm, about 6 % smaller. The title gives a = 4.650 mm, b = 2.065 mm, an outer radius of 2.975 mm and L = 0.500 mm](https://docs.jmrp.io/phonometry/media/comparison_diameter_correction.svg)

*The model of reference [1] for the WS3 in the jig of Figure A.4, with the
rows of Table A.1 and the correction at twice the separation.*

## 5. A free-field calibration by substitution

IEC 61094-8 A.2 describes the basic method: the reference is put at a point
of the free field, then the microphone under test in its place, each read
against a monitor microphone near the source, and "the quotient of these two
gives the ratio of the microphone under test output voltage to the reference
microphone output voltage, corrected for any variation in the sound pressure
generated by the source". With $V_\mathrm{ref}$ and $V_\mathrm{mon,1}$ the
output voltages of the reference and the monitor in the first reading, and
$V_\mathrm{test}$ and $V_\mathrm{mon,2}$ those of the second, the monitor's
two readings going together as a `metrology.MonitorReadings`,

$$
20\lg R_V = 20\lg\frac{V_\mathrm{test}}{V_\mathrm{mon,2}} - 20\lg\frac{V_\mathrm{ref}}{V_\mathrm{mon,1}}
$$

The reference here is an LS2P calibrated in a pressure field, so its
free-field sensitivity is its pressure sensitivity plus the free-field to
pressure difference of IEC/TS 61094-7 (Table 1, 8.2). The values below stand
in for that difference; take yours from the specification or the
microphone's calibration. Between the two readings the source drifts by
0.25 dB:

```python
ff = metrology.exact_frequencies(500, 20000, fraction=3)
xf = ff / 1000
c_ff = 0.05 * xf**1.3            # the LS2P's free-field to pressure difference: illustrative
l_ref_p = -38.0 + 0.04 * np.log10(xf) - 0.25 * (xf / 20) ** 2   # its pressure calibration
l_ws2f = -38.3 + 0.1 * np.log10(xf) - 0.3 * (xf / 20) ** 3      # the WS2F: unknown
field_1 = 74.0 + 0.05 * np.sin(xf)   # the free field while the reference is in place
field_2 = field_1 + 0.25             # the source has drifted by the time the WS2F is
monitor = -40.0                      # the monitor microphone near the source, dB re 1 V/Pa


def free_field_budget(fx):
    x20 = fx / 20000
    return metrology.comparison_uncertainty_budget(
        {
            "reference": 0.06 + 0.14 * x20,    # pressure calibration and IEC/TS 61094-7
            "source_stability": 0.01,
            "positioning": 0.02,
            "alignment": 0.01 + 0.04 * x20**2,
            "free_field": 0.03 + 0.12 * x20**2,
            "non_linearity": 0.017,
            "rounding": 0.003,
            "repeatability": 0.02,
        },
        frequency_hz=fx,
        field="free",
    )


free = metrology.sequential_comparison(
    l_ref_p,
    l_ref_p + c_ff + field_1,     # the reference's output, dB re 1 V
    l_ws2f + field_2,             # the WS2F's output in its place
    frequencies=ff,
    monitor=metrology.MonitorReadings(
        reference_level=monitor + field_1, test_level=monitor + field_2
    ),
    field="free",
    reference_free_field_difference=c_ff,
    expanded_uncertainty_db=[free_field_budget(fx).expanded_uncertainty_db for fx in ff],
)
print(bool(np.abs(free.sensitivity_level - l_ws2f).max() < 1e-9))   # True
print(free.expanded_uncertainty_db[[0, 12, 16]].round(2))              # [0.16 0.26 0.51]
```

The monitor takes the drift out exactly; without it the result would be
0.25 dB high at every frequency. Table 2 of IEC 61094-8 lists the typical
components with the subclause that discusses each, and prints no values:
the ones above are illustrative, chosen so that the expanded uncertainty is
near the 0.2 dB 8.8 estimates at low and middle frequencies and the 0.5 dB
it estimates at the upper frequency limit of an LS2 reference.
`metrology.IEC61094_8_TABLE_2` holds the list, and
`metrology.IEC61094_8_TABLE_1` the typical expanded uncertainty of each way
the reference can be calibrated:

```python
row = metrology.IEC61094_8_TABLE_1["primary_pressure"]
print(row.expanded_uncertainty_1khz_db, row.expanded_uncertainty_10khz_db)   # 0.12 0.4
```

## 6. The effective free-field region of a time window

Where the room is not anechoic enough, a time window can keep the direct
sound and drop the reflections (6.2.3, Annex B). The window defines the region
it simulates a free field in: a prolate spheroid with the source and the
microphone at its foci, $d$ apart, and the major diameter

$$
A = d + \tau c
\tag{B.1}
$$

where $\tau$ runs from the arrival of the sound at the microphone to the end
of the window and $c$ is the speed of sound at the prevailing conditions,
here from the humid air of IEC 61094-2 Annex F. A reflection from any point
of its surface arrives $\tau$ after the direct sound, so every reflecting
surface has to be outside it, and so does the far end of the mounting rod:

```python
r = metrology.free_field_region(1.0, 0.005)   # 1 m, a 5 ms window, at 23.0 °C
print(round(r.speed_of_sound, 1), round(r.major_axis_m, 2))     # 345.9 2.73
print(round(r.semi_minor_axis_m, 2), round(r.rod_clearance_m, 2))   # 1.27 0.86
```

A 5 ms window at 1 m needs 1.27 m of clearance round the axis mid-way between
source and microphone, and a rod that runs more than 0.86 m behind the
microphone.

![The effective free-field region of IEC 61094-8 B.1 in a plane through its axis: an ellipse 2.73 m long and 2.54 m across, the sound source at x = -0.5 m and the microphone at x = +0.5 m on its axis, joined by the dashed direct path of 1.00 m, and the mounting rod running from the microphone out past the right end of the ellipse; the title gives A = d + tau c = 2.73 m, tau = 5.0 ms, c = 345.9 m/s, a clearance of 1.27 m across and 0.86 m behind the microphone](https://docs.jmrp.io/phonometry/media/free_field_region.svg)

*The region of a 5 ms window at 1 m, with the rod that has to leave it.*

## 7. Time-selective processing

Where the room is not anechoic enough, Annex B keeps the direct sound and
drops the reflections: an impulse response separates the two in time, a
window keeps the first, and a Fourier transform gives the frequency response
of the direct sound alone (B.1.1). The impulse response can come from a
stepped sine taken to the time domain by the inverse transform,

$$
H(f) = \int_{-\infty}^{\infty} h(t)\,\mathrm{e}^{-\mathrm{j}2\pi f t}\,\mathrm{d}t,
\qquad
h(t) = \int_{-\infty}^{\infty} H(f)\,\mathrm{e}^{\mathrm{j}2\pi f t}\,\mathrm{d}f
\tag{B.2, B.3}
$$

(`metrology.stepped_sine_impulse_response`), from a sweep or a maximum length
sequence (`room.impulse_response`, `room.mls_impulse_response`), or from a
direct impulse. `metrology.time_selective_response` weights it with a time
window and evaluates (B.2) at any frequency. "In order not to produce
artefacts in the transformed data, the time window normally has 'tapered'
edges" (B.1.2); B.1.3 lists Hann, Hamming, Tukey, Butterworth cosine and
Gaussian windows and does not recommend a rectangular one, "because it
usually leads to spectral leakage". The default is a Tukey window, flat in
the middle with an eighth of it tapered at each edge, a proportion the
library chose, as no clause prints one.

Here an LS2P reference and a WS2F stand in turn 1 m from the source, and a
reflection off the rig arrives along a path of 1.48 m. The WS2F hears it off
axis, so it does not cancel in the ratio of the two outputs as the direct
sound does: "it cannot be assumed that the reference microphone and
microphone under test will deviate in the same way" (8.5). Each is measured
by a stepped sine from 0 Hz to 100 kHz in 50 Hz steps, past three times the
resonance as B.2.1 asks; the microphones are illustrative:

```python
c0 = metrology.free_field_region(1.0, 0.005).speed_of_sound   # 345.9 m/s at 23.0 °C
f_step = np.arange(0.0, 100000.0 + 1.0, 50.0)   # 0 Hz to 100 kHz in 50 Hz steps


def mic(fx, level, f0, d):   # a microphone's sensitivity, V/Pa: illustrative
    return 10 ** (level / 20) / (1 - (fx / f0) ** 2 + 1j * d * fx / f0)


direct, rig = 1.0 / c0, 1.48 / c0   # the direct sound over 1 m, the rig's reflection over 1.48 m


def heard(m, off_axis):   # each step: the direct sound and the rig's reflection
    return m * (
        np.exp(-2j * np.pi * f_step * direct)
        + 0.25 * off_axis * np.exp(-2j * np.pi * f_step * rig)
    )


ir_ref = metrology.stepped_sine_impulse_response(heard(mic(f_step, -38.0, 22000, 1.1), 1.0), frequencies=f_step)
ir_ws2f = metrology.stepped_sine_impulse_response(
    heard(mic(f_step, -26.4, 18000, 0.9), 1 / (1 + 1j * f_step / 6000)), frequencies=f_step
)   # the WS2F hears the reflection off axis
print(ir_ref.duration_s, round(ir_ref.fs))   # 0.02 200050
```

The impulse response lasts the inverse of the frequency step, 20 ms. B.2.2
sizes the step by the room: 120 Hz is enough in a small anechoic room
"because the primary reflections all occur before 8 ms", about 30 Hz in a
large one. (B.2.1 says the increment "will determine the time domain
resolution"; it is the length it sets, as B.2.2 says, and the
[errata register](https://docs.jmrp.io/phonometry/reference/errata/) records the sentence.)
`ir_ws2f.plot()` draws the whole record:

![The WS2F's impulse response from the stepped sine of 0 Hz to 100 kHz in 50 Hz steps, normalised to its peak, over the whole 20 ms record: the direct sound at 2.89 ms with a short ringing, the rig's reflection at 4.28 ms at about a tenth of the peak, and nothing after them up to 20 ms. The title gives the 50 Hz step and the 20.00 ms length](https://docs.jmrp.io/phonometry/media/stepped_sine_impulse_response.svg)

*The WS2F's impulse response from the stepped sine, the whole $1/\Delta f$ of
it, from `ir_ws2f.plot()`.*

The window has to end before the reflection arrives. A point on
the boundary of the region of Formula (B.1) reflects along a path of the
major diameter $A = d + \tau c$, so a reflection whose path is $L$ stays
outside while

$$
\tau \le \frac{L - d}{c}
$$

which `metrology.reflection_free_window_s` gives, at the speed of sound of
the IEC 61094-2 Annex F air:

```python
tau = metrology.reflection_free_window_s(1.0, 1.48)
window = {"window_start_s": direct - 0.0005, "window_end_s": direct + tau}
ref = metrology.time_selective_response(ir_ref.impulse_response, ir_ref.fs, **window)
ws2f = metrology.time_selective_response(ir_ws2f.impulse_response, ir_ws2f.fs, **window)
print(round(1000 * tau, 2), round(ref.frequency_resolution_hz))   # 1.39 530
```

A window of 1.89 ms tells frequencies apart no closer than 530 Hz, the
low-frequency limit 6.2.3 warns time-selective techniques have; lower
frequencies need another method. The levels and phases through the window
are the two outputs of a substitution:

```python
fw = metrology.exact_frequencies(1900, 20000, fraction=3)   # 2 kHz to 20 kHz
m_ref = mic(fw, -38.0, 22000, 1.1)   # the reference's free-field sensitivity, from its certificate
tsel = metrology.sequential_comparison(
    20 * np.log10(np.abs(m_ref)),
    ref.level_at(fw),
    ws2f.level_at(fw),
    frequencies=fw,
    field="free",
    phase=metrology.SequentialComparisonPhase(
        reference_sensitivity_phase_deg=np.degrees(np.angle(m_ref)),
        reference_output_phase_deg=ref.phase_deg_at(fw),
        test_output_phase_deg=ws2f.phase_deg_at(fw),
    ),
)
m_ws2f = mic(fw, -26.4, 18000, 0.9)   # what the page knows the WS2F to be
print(tsel.sensitivity_level[[0, 6, 10]].round(2))   # [-26.34 -25.46 -26.6 ]
print(bool(np.abs(tsel.sensitivity_level - 20 * np.log10(np.abs(m_ws2f))).max() < 1e-4))   # True
record = 20 * np.log10(np.abs(ws2f.record_response_at(fw) / ref.record_response_at(fw)))
print(np.abs(record - 20 * np.log10(np.abs(m_ws2f / m_ref))).max().round(2))   # 1.91
```

Through the window the WS2F's sensitivity comes back to within 0.0001 dB;
from the whole record, the reflection included, it would be up to 1.91 dB
off.

![Left, the WS2F's impulse response from a stepped-sine measurement, its direct sound at 2.89 ms and the rig's reflection at 4.28 ms, with a Tukey window 1.89 ms long from 2.39 ms that ends as the reflection arrives. Right, its frequency response from 1 kHz to 40 kHz: through the window it is the smooth response of the direct sound, and from the whole record it ripples round it, from 2.45 dB below to 1.89 dB above up to 16 kHz](https://docs.jmrp.io/phonometry/media/time_selective_response.svg)

*The WS2F's impulse response with the window over it (left), and its
response through the window and from the whole record (right).*

**The direct impulse method** (B.6), which B.6.2 calls "largely superseded",
drives the source with a short rectangular pulse of amplitude $a$. Its
spectrum is

$$
X(f) = \frac{2ab\sin(2\pi f b)}{2\pi f b}
\tag{B.10}
$$

whose first zero, $1/(2b)$, "must be approximately an order of magnitude
higher than the upper limit of the frequency range of interest". The formula
and its first zero are those of a pulse lasting $2b$, although the text calls
$b$ the duration (it is in the [errata register](https://docs.jmrp.io/phonometry/reference/errata/));
`metrology.rectangular_pulse` takes the whole duration $T = 2b$:

```python
T = metrology.rectangular_pulse_duration_s(20000)   # a first zero ten times 20 kHz
pulse = metrology.rectangular_pulse(T, amplitude_v=10.0)
print(T, pulse.half_duration_s, round(pulse.first_zero_hz))   # 5e-06 2.5e-06 200000
print(pulse.level_at([20000.0]).round(2))                  # [-0.14]
```

The responses to several pulses, averaged synchronously
(`signals.time_synchronous_average`; B.6.2 counts on $10\lg n$ dB of
signal-to-noise ratio from $n$ of them), go to `time_selective_response` with
`excitation=pulse`, which divides the pulse's spectrum out, as a pulse that
begins at the start of the record.

![The spectrum of a 5 microsecond rectangular pulse relative to its value at 0 Hz, from 0 to 500 kHz: flat to within 0.14 dB up to the 20 kHz upper limit, falling to its first zero at 200 kHz and a second at 400 kHz](https://docs.jmrp.io/phonometry/media/rectangular_pulse.svg)

*Formula (B.10) for a 5 µs pulse: 0.14 dB down at 20 kHz, the first zero at
200 kHz.*

## 8. The phase of the sensitivity

Both parts carry the phase over with the modulus: IEC 61094-5 5.1.1
calculates "the sensitivity (both modulus and phase) of the test
microphone", and IEC 61094-8 5.1 "both the modulus and phase of the
free-field sensitivity of the microphone under test". D.2 holds for the complex
ratios, so the phase follows the level term by term, and the interchange of
Annex C cancels the phase shifts of the two channels and of the coupler as
(C.3) cancels their gains:

$$
\varphi_\mathrm{test} = \varphi_\mathrm{ref} + \arg R_V - \arg R_P,
\qquad
\varphi_\mathrm{ref} - \varphi_\mathrm{test} = \tfrac12\left(\Phi_\mathrm{C12} - \Phi_\mathrm{C21}\right)
$$

with $\Phi_\mathrm{C12}$ the phase of channel 1's reading re channel 2's.
Neither part prints this form; it is (C.3) written for the complex ratio. The
pressure calibration of section 2, with the phase of each reading:

```python
phi_ref = np.degrees(np.angle(1 / (1 - (f / 22000) ** 2 + 1.1j * f / 22000)))   # the LS2P's certificate: illustrative
phi_true = np.degrees(np.angle(1 / (1 - (f / 18000) ** 2 + 0.9j * f / 18000)))   # the WS2P: unknown
shift_1, shift_2 = 1.5 * x, -0.8 * x   # the two channels' phase shifts, degrees
late_a = 0.4 * x                       # port A hears the field later, degrees
jitter = np.random.default_rng(5).normal(0.0, 0.02, (2, 3, f.size))
p12 = (phi_ref + shift_1) - (phi_true + shift_2) + late_a + jitter[0]   # reference on channel 1
p21 = (phi_true + shift_1) - (phi_ref + shift_2) + late_a + jitter[1]   # interchanged
cp = metrology.simultaneous_comparison(
    l_ref, l_c12, l_c21, frequencies=f,
    phase=metrology.SimultaneousComparisonPhase(
        reference_sensitivity_phase_deg=phi_ref,
        channel_phase_difference_deg=p12,
        interchanged_channel_phase_difference_deg=p21,
    ),
)
print(cp.sensitivity_phase_deg[[2, 8, 16, 19]].round(1))   # [  -1.1   -4.6  -35.9 -102.9]
print(np.abs(cp.sensitivity_phase_deg - phi_true).max().round(3))   # 0.016
```

The difference of the two readings is twice the phase difference of the
microphones, so it is taken into one turn before it is halved: the result is
unique while they differ by less than 90°. The phase inputs go together, as a
`metrology.SimultaneousComparisonPhase` here. In a sequential calibration
they are a `metrology.SequentialComparisonPhase`, the phases of the two
outputs are read against the monitor's, which go with its levels in its
`metrology.MonitorReadings`, or against the signal driving the source, and
their quotient is $\arg R_V$. In a free field
"determination of the phase of the free-field sensitivity requires the
definition of consistent reference phases at the acoustic centres of the
microphones" (IEC 61094-8 5.1). 7.3 positions the centres "at the
measurement points"; in a sequential substitution each goes in turn to the
same point, where $\arg R_P$ is 0, and a delay common to the two readings,
like the travel time from the source in section 7, cancels in their quotient.

![The phase of the sensitivity against frequency. Left, the WS2P calibrated in a coupler through the interchange, from 250 Hz to 20 kHz: near the LS2P reference's phase up to 4 kHz, then falling faster, to -35.9 degrees at 10 kHz and -102.9 degrees at 20 kHz against the reference's -32.2 and -79.9. Right, the WS2F calibrated through the time window from 2 kHz to 20 kHz, the same shape](https://docs.jmrp.io/phonometry/media/comparison_phase.svg)

*The phase of the WS2P through the interchange (left) and of the WS2F through
the window of section 7 (right), from `.plot(quantity="phase")`.*

## 9. Microphones of different acoustic impedance

"Differences in the acoustic impedance between the test and reference
microphones can cause the sound pressure at the test and reference
microphones to differ" (7.4), most where a pressure and a free-field
microphone meet above 10 kHz, "because of differing volume velocities at the
diaphragms" (7.5). Neither clause prints a model: 7.4 sends the reader to the
literature, and 7.5 asks for the error to be estimated and added to the
budget when no reference of similar impedance is available.
`metrology.impedance_pressure_ratio` writes the effect for two circuits with
each microphone's equivalent volume, which IEC 61094-1 6.2.2
defines as $V_\mathrm{e} = \kappa_\mathrm{r} p_{s,\mathrm{r}} / (\mathrm{j}\omega
Z_\mathrm{a})$ with $\kappa_\mathrm{r} = 1.40$:

$$
R_P = \frac{V_x + V_\mathrm{e,ref}}{V_x + V_\mathrm{e,test}}
$$

In a closed coupler small against the wavelength the source drives the sum
of the admittances of the gas and of every microphone in it, IEC 61094-2
Formula (3), so putting the test microphone in the place of the reference
changes the pressure by the ratio of the two sums, $V_x$ being the rest of
the load: the gas, the source, a monitor. For a simultaneous excitation,
Table D.1 puts "the acoustical impedance of the microphone [...] in series
with that of the air in the space between the two microphones", so that the
two "see slightly different pressures [...] (see 7.4 and [2])": each diaphragm
takes the pressure of a common source behind the same series impedance, and
$V_x$ is the equivalent volume of that impedance, which the circuit of [2]
gives (below). `metrology.ReciprocityMicrophone`,
the microphone of
[the reciprocity calibration](https://docs.jmrp.io/phonometry/signals/metrology/reciprocity-calibration/),
holds the lumped parameters of IEC 61094-2 E.4, the low-frequency volume, the
resonance frequency and the loss factor, and its
`complex_equivalent_volume_m3` gives $V_\mathrm{e}$ from them. Its ratio to
the low-frequency volume holds no $\kappa_\mathrm{r}$, so with the 10 mm³
that Table 3 of IEC 61094-1 gives an LS2P, $V_\mathrm{e}$ is the volume of
6.2.2 with its 1.40. E.4 finds the representation "of sufficient accuracy" up
to "about 1,3 times the resonance frequency of the microphone": 20.8 kHz for
the WS2F below, so the example stays within it. The front cavity the
microphone also holds enters the transfer impedance of a coupler, not
$V_\mathrm{e}$.

```python
fi = metrology.exact_frequencies(1000, 20000, fraction=3)
front = {"front_cavity_volume_m3": 34e-9, "front_cavity_depth_m": 0.5e-3,
         "front_cavity_diameter_m": 9.3e-3}   # an LS2aP's (Table C.1 of IEC 61094-2)
lumped_ls2p = metrology.ReciprocityMicrophone(
    equivalent_volume_m3=10e-9, resonance_frequency_hz=22000, loss_factor=1.1, **front
)   # an LS2P: 10 mm³ (IEC 61094-1 Table 3); the resonance and loss: illustrative
lumped_ws2f = metrology.ReciprocityMicrophone(
    equivalent_volume_m3=30e-9, resonance_frequency_hz=16000, loss_factor=0.5, **front
)   # a more compliant WS2F: illustrative
rp = metrology.impedance_pressure_ratio(
    fi,
    reference_equivalent_volume_m3=lumped_ls2p.complex_equivalent_volume_m3(fi),
    test_equivalent_volume_m3=lumped_ws2f.complex_equivalent_volume_m3(fi),
    coupling_equivalent_volume_m3=600e-9,   # the coupler's gas, source and monitor: illustrative
)
print(rp.level_difference[[0, 10, 13]].round(3))      # [-0.281 -0.421  0.37 ]
print(rp.standard_uncertainty_db[[0, 10, 13]].round(3))  # [0.163 0.243 0.213]
```

A monitor that "shall accurately sense changes in the sound pressure at the
test/reference microphone position" (5.1.3) takes this ratio out. Without
one, `rp.level_difference` and `rp.phase_difference_deg` go to the
comparison as `pressure_level_difference` and as the
`pressure_phase_difference_deg` of its phase inputs. Where the ratio is estimated rather than
corrected for, `rp.standard_uncertainty_db` is the `"impedance"` component of
the budget: the level difference read as the semi-range of a rectangular
distribution, as Table D.1 reads its 0.005 dB at 2 kHz, "equivalent to a
standard uncertainty of 0,005/√3 dB = 0,003 dB". The same row warns about
exactly this pairing: "When microphones have significantly differing
impedances (for example WS2F microphone compared against LS2P at frequencies
above 10 kHz), the measurement uncertainty can be considerably larger and
should be established experimentally." Above 10 kHz the value here is the
model's estimate, not that experiment; where the ratio crosses 0 dB, near
16 kHz, it gives almost nothing at all.

![The pressure ratio of a WS2F in the place of an LS2P in a closed coupler, from 1 kHz to 20 kHz. Left, its level: -0.28 dB at 1 kHz, falling to -0.44 dB at 12.5 kHz, crossing 0 dB at 16 kHz past the WS2F's resonance and reaching +0.37 dB at 20 kHz, with its standard uncertainty dashed from 0.16 dB to 0.25 dB, dropping to almost nothing at 16 kHz where the level crosses 0 dB. Right, its phase, about 0.7 degrees at 8 kHz and peaking at 4.86 degrees at 16 kHz](https://docs.jmrp.io/phonometry/media/comparison_impedance.svg)

*A WS2F in the place of an LS2P in a closed coupler: the level of $R_P$ with
its semi-range read as a standard uncertainty (left), which above 10 kHz
Table D.1 asks to be established experimentally, and its phase (right).*

### The air between the two microphones

The "Microphone impedance" row of Table D.1 refers the simultaneous
excitation to its reference [2], Jarvis (1996), NPL Report CIRA(EXT) 010.
Its Appendix B models a symmetric coupler as a ladder from a source in its
middle: on each side a mass $Z_L$, the compliance $Z_c$ of that side's half
of the volume, a second $Z_L$ and the microphone, for a tube of length $L$
and radius $r$,

$$
Z_c = \frac{\kappa p_0}{\mathrm{j}\omega V/2},
\qquad
Z_L = \frac{\mathrm{j}\omega}{3}\,\frac{\rho L/4}{\pi r^2},
\qquad V = \pi r^2 L
$$

Seen from each diaphragm, the middle of the ladder is a pressure common to
both behind one impedance,

$$
Z_x = Z_L + \frac{Z_L Z_c}{Z_L + Z_c}
$$

so each diaphragm takes $p\,Z_\mathrm{a}/(Z_\mathrm{a} + Z_x)$, the divider
above, and the ratio of the two is the one Appendix B prints.
`metrology.air_gap_series_impedance_pa_s_m3` gives $Z_x$ for a coupler of
known length and radius, at the air of the conditions given. Appendix B's
example is an LS2P against a high sensitivity WS2P in a tube 2 mm long and
12.7 mm across, each microphone a resistance, a mass and a compliance in
series:

```python
fj = np.arange(100.0, 20000.0 + 1.0, 100.0)   # Appendix B's frequencies
w = 2 * np.pi * fj


def z_mic(r, c, m):   # Z_m = r + j w m + 1/(j w c), Appendix B
    return r + 1j * w * m + 1 / (1j * w * c)


def v_e(z):           # the equivalent volume of IEC 61094-1 6.2.2
    return 1.40 * 101325 / (1j * w * z)


z_x = metrology.air_gap_series_impedance_pa_s_m3(fj, gap_length_m=2e-3, gap_radius_m=6.35e-3)
gap = metrology.impedance_pressure_ratio(
    fj,
    reference_equivalent_volume_m3=v_e(z_mic(330e6, 48e-15, 1210)),   # the LS2P of Appendix B
    test_equivalent_volume_m3=v_e(z_mic(7e7, 2.6e-13, 820)),           # its high sensitivity WS2P
    coupling_impedance_pa_s_m3=z_x,
)
print(gap.level_difference[[69, 199]].round(4))   # [ 0.0064 -0.025 ]
print(gap.phase_difference_deg[124].round(3))         # -0.158
```

The WS2P hears 0.0064 dB more than the LS2P at 7 kHz and 0.025 dB less at
20 kHz, and lags it by up to 0.158° at 12.5 kHz. Appendix B draws the same
curves for $P_1/P_2$, the LS2P re the WS2P, so the other way up, with its
1.21 kg/m³ for the density of the air; with that density the conformance
report reproduces them. In the published scan its impedance of the LS2P
shows no imaginary unit on the mass term, where blank space stands; the
second microphone's has it, and the graph needs it (an
[erratum](https://docs.jmrp.io/phonometry/reference/errata/)).

![The pressure ratio of the high sensitivity WS2P re the LS2P of Jarvis (1996) Appendix B in a coupler 2 mm long and 12.7 mm across, from 100 Hz to 20 kHz. Left, its level: nothing up to 1 kHz, rising to +0.0064 dB at 7 kHz, crossing 0 dB near 10 kHz and falling to -0.025 dB at 20 kHz, with its standard uncertainty dashed. Right, its phase: nothing up to 2 kHz, falling to -0.158 degrees at 12.5 kHz and back to -0.09 degrees at 20 kHz](https://docs.jmrp.io/phonometry/media/comparison_air_gap.svg)

*The air between an LS2P and a high sensitivity WS2P in series with each,
by the circuit of reference [2]: the level of $R_P$ with its semi-range read
as a standard uncertainty (left), and its phase (right).*

## 10. The diffuse-field comparison of IEC 61183

The diffuse-field method of IEC 61183 clause 5 is a sequential comparison
without a monitor: a sound level meter and a reference meter read in turn at
the same positions of a reverberation room, the difference of their readings
added to the reference's diffuse-field sensitivity level. The library computes
it through the same model, so the two cannot drift apart:

```python
d = metrology.diffuse_field_sensitivity(
    [94.3, 94.6], [94.0, 94.1], frequencies=[1000.0, 2000.0],
    reference_random_incidence_level=[-26.0, -26.2],
)
s = metrology.sequential_comparison([-26.0, -26.2], [94.0, 94.1], [94.3, 94.6], frequencies=[1000.0, 2000.0])
print(d.diffuse_field_level, s.sensitivity_level)   # [-25.7 -25.7] [-25.7 -25.7]
```

## 11. Validating a jig or a coupler

"Calibrations performed in any particular jig or coupler shall be validated
by comparison with calibrations performed in other jigs and couplers and
alternative sound sources. A separate validation is necessary for each
different type of microphone. If the test microphone is a laboratory
standard microphone, then the jig or coupler can be validated by comparing
a comparison calibration with a reciprocity calibration. For some
microphones, it can be necessary to use more than one jig and/or coupler to
cover a full frequency range with low uncertainty" (6.7).
`metrology.verify_jig_or_coupler` sets a calibration made in the jig or
coupler beside another calibration of the same microphone, a comparison made
elsewhere or, for a laboratory standard microphone, the
`ReciprocityCalibration` of [the reciprocity
calibration](https://docs.jmrp.io/phonometry/signals/metrology/reciprocity-calibration/) with the
index of that microphone, and compares them at the frequencies they share.
The clause prints no criterion. The library reads "validated" as the two
agreeing within the expanded uncertainty of their difference, which for two
independent calibrations is the root-sum-square of the two,

$$
\lvert L_\mathrm{cal} - L_\mathrm{val} \rvert \le U_\Delta =
\sqrt{U_\mathrm{cal}^2 + U_\mathrm{val}^2}
$$

Two calibrations that share a component are correlated, and the part they
share comes out of $U_\Delta$ (below).

Here the jig of A.2 is validated against the coupler of A.1 for the WS2P of
the first example, the coupler used up to the 10 kHz it is for. Near
6.3 kHz the jig has a standing wave close to the microphones, the anomaly
Barham et al. found in a test box at that frequency, and the field moves
between the reading and the interchange, so (C.3) no longer cancels it:

```python
fv = f[6:]   # the jig of A.2, 1 kHz to 20 kHz
mode = 0.5 * np.exp(-(((fv / 1000 - 6.3) / 0.6) ** 2))   # the field moves near 6.3 kHz
jitter = np.random.default_rng(42).normal(0.0, 0.01, (2, 3, fv.size))
j12 = (l_ref[6:] + gain_1) - (l_true[6:] + gain_2) + field_a[6:] + mode + jitter[0]
j21 = (l_true[6:] + gain_1) - (l_ref[6:] + gain_2) + field_a[6:] + jitter[1]
jig = metrology.simultaneous_comparison(l_ref[6:], j12, j21, frequencies=fv, expanded_uncertainty_db=u[6:])
coupler = metrology.simultaneous_comparison(
    l_ref[:17], l_c12[:, :17], l_c21[:, :17], frequencies=f[:17], expanded_uncertainty_db=u[:17]
)   # the coupler of A.1, up to 10 kHz
check = metrology.verify_jig_or_coupler(jig, coupler)
print(check.passes, check.failing_frequencies_hz.round())    # False [6310.]
print(check.unvalidated_frequencies_hz.round())              # [12589. 15849. 19953.]
print(check.difference[8].round(3), check.expanded_uncertainty_db[8].round(3))   # -0.247 0.128
```

The jig agrees with the coupler at every frequency they share but 6.3 kHz,
where they differ by 0.247 dB against 0.128 dB, and above 10 kHz it is not
validated at all, the coupler not reaching there: a second set-up has to
validate it there, as 6.7 allows. `check.normalized_difference` is the
difference in units of its expanded uncertainty, -1.93 at 6.3 kHz.

### Calibrations that share a reference

Both calibrations here take the same LS2P as their reference, so the first
row of Table D.1, the calibration of that reference (0.025 dB, the
"±0,05 dB with a coverage factor of k = 2" of its certificate), is in both
budgets. Whatever error that calibration carries moves the two sensitivity
levels together and cancels in their difference, while the root-sum-square
counts it twice. In the terms of the GUM, two levels that each depend on a
common quantity with a sensitivity of 1 have its variance as their
covariance (JCGM 100:2008 F.1.2.3, Formula (F.2), whose sensitivity
coefficients are then both 1), and the law of propagation for correlated inputs
(5.2.2, Formula (13)) gives for the difference

$$
u_\Delta^2 = u_\mathrm{cal}^2 + u_\mathrm{val}^2 - 2u_\mathrm{sh}^2
$$

with $u_\mathrm{sh}$ the standard uncertainty of what the two share.
`shared_standard_uncertainty_db` takes it, one value or one per frequency of
the calibration, and the band becomes, with the $k = 2$ of 7.9,

$$
U_\Delta = \sqrt{U_\mathrm{cal}^2 + U_\mathrm{val}^2 - 2(k\,u_\mathrm{sh})^2}
$$

```python
shared = metrology.verify_jig_or_coupler(jig, coupler, shared_standard_uncertainty_db=0.025)
print(shared.passes, shared.failing_frequencies_hz.round())                                # False [6310.]
print(shared.expanded_uncertainty_db[8].round(3), shared.normalized_difference[8].round(2))   # 0.106 -2.32
print(shared.correlation_coefficient[8].round(2))                                           # 0.31
```

The band narrows from 0.124 dB to 0.101 dB at the low frequencies and from
0.128 dB to 0.106 dB at 6.3 kHz, where the two calibrations are correlated
with a coefficient of 0.31, the reference's variance over the product of
their standard uncertainties (Formula (14)). The verdict does not change
here: the jig still fails at 6.3 kHz alone, now by 2.32 times the
uncertainty of the difference. A pair that agrees within the root-sum-square
can fail once the shared part is out of it. Which components the two share
is the laboratory's to say: the calibration of a reference both use, its
drift since then (the row "Drift in reference microphone sensitivity since
last calibration", 0.017 dB) for two calibrations made close together, a
measuring chain they both go through. Only a component that enters both
levels with the same sign and size belongs here. A reciprocity calibration
made in the same set of three microphones as the reference's is not such a
case: in IEC 61094-2:2009 Formula (7) (folio 12) the electrical transfer
impedances of the two pairs that go through the third microphone are in the
numerator of one microphone's sensitivity and in the denominator of the
other's, so those measurements enter the two levels with opposite signs.
Their covariance is negative and widens the uncertainty of the difference
beyond the root-sum-square, which `shared_standard_uncertainty_db` cannot
express; only the factors common to both levels with the same sign may be
given there. A shared part larger than either calibration's standard
uncertainty is refused, since nothing can be shared beyond the whole of it.

![The difference between the WS2P calibrated in the jig and in the coupler at the eleven frequencies they share, 1 kHz to 10 kHz, inside a shaded band of plus or minus the expanded uncertainty of the difference with the reference they share taken out, 0.101 dB widening to 0.124 dB at 10 kHz, and dotted around it the band of two independent calibrations, 0.124 dB widening to 0.143 dB: within 0.008 dB of zero everywhere but 6.3 kHz, where it falls to -0.247 dB, outside both bands and marked with a cross; three dashed vertical lines at 12.6 kHz, 15.8 kHz and 20 kHz mark the frequencies of the jig the coupler does not cover](https://docs.jmrp.io/phonometry/media/comparison_jig_validation.svg)

*The jig against the coupler: the difference of the two calibrations inside
the expanded uncertainty of the difference once the reference they share is
taken out, the root-sum-square of two independent calibrations dotted around
it, the one frequency where the jig fails, and dashed, the three above
10 kHz the coupler does not reach.*

The clause prints neither a criterion nor an example, so the conformance
report checks the verdict against its own definition. Jarvis and Watkins
(1997), NPL Report CIRA(EXT) 022, who compared jig calibrations of working
standard microphones at NPL with reciprocity calibrations of the same
microphones, print no numeric criterion either: their Figures 6 and 7
(folio 4) plot the differences, and the text judges them against "the
typical repeatability" of the measurements (folio 2), one standard
deviation drawn as a band in Figure 7 and never printed as a number, with
no number a criterion could be checked against. The residual differences
"of less than 0.1 dB" of folio 5 report a result rather than set a test, and
the report says nothing about what the two calibrations share. The shared
part has printed numbers, in the GUM: two items
calibrated against the same standard, of relative standard uncertainty
$10^{-4}$, are correlated with coefficients of about 0.5, 0.990 and 1.000
for comparisons of $100 \times 10^{-6}$, $10 \times 10^{-6}$ and
$1 \times 10^{-6}$ (F.1.2.3, Example 2). The conformance report checks
`correlation_coefficient` against them, and the difference of two
calibrations with the budget of Table D.1 against the seven rows left once
the reference's cancels.

**Show the code for these figures**

```python
import matplotlib.pyplot as plt

fig, (ax_pressure, ax_free) = plt.subplots(1, 2, figsize=(13.5, 5.6))
c.plot(ax_pressure)
free.plot(ax_free)
fig.tight_layout()

fig, (ax_d1, ax_ff) = plt.subplots(1, 2, figsize=(13.5, 5.6))
b.plot(ax_d1)
free_field_budget(8000).plot(ax_ff)
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
j.plot(ax)
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
fine.plot(ax, marker="none")
ax.plot(j.frequencies, j.correction, ls="none", marker="D", label="Table A.1")
ax.legend()
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
r.plot(ax)
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
ir_ws2f.plot(ax)
fig.tight_layout()

fig, (ax_time, ax_response) = plt.subplots(1, 2, figsize=(13.5, 5.6))
ws2f.plot(ax_time)
ax_time.set_xlim(1000 * (direct - 0.0015), 1000 * (direct + 0.0035))
ws2f.plot(ax_response, quantity="response", frequencies=np.geomspace(1000, 40000, 300))
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
pulse.plot(ax, upper_frequency_hz=20000)
fig.tight_layout()

fig, (ax_pressure, ax_free) = plt.subplots(1, 2, figsize=(13.5, 5.6))
cp.plot(ax_pressure, quantity="phase")
tsel.plot(ax_free, quantity="phase")
fig.tight_layout()

fig, (ax_level, ax_phase) = plt.subplots(1, 2, figsize=(13.5, 5.6))
rp.plot(ax_level)
rp.plot(ax_phase, quantity="phase")
fig.tight_layout()

fig, (ax_level, ax_phase) = plt.subplots(1, 2, figsize=(13.5, 5.6))
gap.plot(ax_level, marker="none")
gap.plot(ax_phase, quantity="phase", marker="none")
fig.tight_layout()

fig, ax = plt.subplots(figsize=(10, 6))
shared.plot(ax)
fig.tight_layout()
plt.show()
```

## What this guide covers

- **Covered.** **IEC 61094-5:2016 and IEC 61094-8:2012, the calculations:** the level
  model of D.2; the simultaneous excitation with the interchange of Annex C
  (Formulas (C.1) to (C.3)), required for a pressure calibration, and without
  it for a free-field one (5.3); the sequential excitation against a monitor
  microphone (61094-5 5.1.3, 61094-8 A.2); the average over the
  determinations; the reference's free-field to pressure difference in a
  free-field calibration (Table 1, 8.2); the WS3 corrections of Table A.1 with
  their uncertainty; the budgets of Table D.1 and Table 2 on the GUM with
  $k = 2$ (7.9, D.2, D.3, 8.8), and the combination in linear form D.3
  mentions; Tables A.1, D.1, 1 and 2 as published data; and the effective
  free-field region of Formula (B.1), with the speed of sound of IEC 61094-2
  Annex F, and the longest window it allows for a reflection. The phase of the
  sensitivity (61094-5 5.1.1, 61094-8 5.1) through the interchange and the
  monitor ratios. **IEC 61094-8 Annex B:** the time window of B.1.3 (Tukey,
  Hann, Hamming, and the rectangular one it does not recommend) with the
  tapered edges of B.1.2, Formula (B.2)
  through it, the stepped sine of B.2 by Formula (B.3), and the rectangular
  pulse of the direct impulse method (B.6, Formula (B.10)). **IEC 61094-5 7.4
  and 7.5:** the pressure ratio of microphones of different acoustic impedance
  in a closed coupler (IEC 61094-2 Formula (3)), with the lumped equivalent
  volume of IEC 61094-2 E.4 and IEC 61094-1 6.2.2, and in series with the air
  between two microphones, whose impedance is that of the circuit of Appendix B
  of the part's reference [2], Jarvis (1996), with the ratio it prints and the
  extremes of its graph reproduced. **IEC 61094-5 6.5:** the correction for a
  test microphone smaller than its reference by the model of the part's
  reference [1], Barham et al. (2014), from any geometry and separation, with
  its change on doubling the separation that A.2 names; it reproduces every
  row of Table A.1 from the inputs of the paper's Table 1. **IEC 61094-5 6.7:**
  the comparison of a calibration made in a jig or a coupler with another of
  the same microphone, by comparison elsewhere or by reciprocity, at the
  frequencies they share, with the part of their uncertainty the two share
  taken out of the uncertainty of the difference by the GUM law of
  propagation for correlated inputs (JCGM 100:2008 5.2.2 and F.1.2.3), whose
  correlation coefficients of F.1.2.3 Example 2 are reproduced. The IEC 61183 clause 5 comparison computes through
  the same model. Table D.1 and Table A.1 are reproduced.

- **Covered, by a reading of the text.** **The environmental correction.** 6.6 and 7.6 ask for the reference to be
  corrected to the conditions of the test, or the result to the reference
  conditions, and print no formula; the first-order sum of a coefficient times
  a deviation for each condition is this library's reading, with the
  coefficients the microphone's own. **The mean of the two ratios** of 5.1.2
  is taken in decibels, as Annex C does. **The coverage factor of the WS3
  corrections.** A.2, the NOTE to Table A.1 and the special case of Table D.1
  give their expanded uncertainty as 10 % of each correction without a coverage
  factor; it is read as the $k = 2$ that 7.9 and D.2 report every expanded
  uncertainty with, so the standard uncertainty is 5 % of the correction.
  **The phase.** Neither part prints a phase form of (C.3) or of the monitor
  ratios; they are the level forms written for the complex ratio that 5.1.1
  calls the sensitivity "both modulus and phase", the half-difference of the
  interchange taken in one turn. **The impedance ratio.** 7.4 and 7.5 leave the
  model to the literature. The coupler is the printed Formula (3) of
  IEC 61094-2; for a simultaneous excitation the divider, each diaphragm behind
  the same series impedance from a common source, is the circuit of Appendix B
  of Jarvis (1996) seen from each microphone, read with the imaginary unit its
  first microphone's mass term has lost in the published scan. The standard uncertainty
  reads the level difference as a semi-range, as that row does; above 10 kHz,
  for a WS2F against an LS2P, the row asks for it to be established
  experimentally. **The model of 6.5.** As printed, the formulas of Barham et
  al. (2014) with their own inputs do not give their own Table 2; the library
  reads the radial sensitivity of Formula (4) less 1 and the annulus of
  Formula (2) from the overall radius, which give it, at a speed of sound the
  paper does not print. Its uncertainty is the caller's to estimate: the model
  gives the change on doubling the separation that A.2 names as approximately
  the 10 % of Table A.1. **The validation of 6.7.** The clause prints no
  criterion, and Jarvis and Watkins (1997), which compared jig calibrations
  with reciprocity ones, print no numeric one either, judging the differences
  in words against one standard deviation of repeatability drawn in their
  Figure 7; agreement within the expanded uncertainty of the difference, the
  root-sum-square of the two less the part they share, is this library's
  reading, and the conformance report checks it against its own definition.
  Which components two calibrations share is the caller's to give, and only
  one that enters both levels with the same sign and size can be given. **Formula (B.10)** is read as the spectrum of a
  pulse lasting $2b$, as the formula and its first zero say and the text does
  not.

- **Not covered.** The measurements themselves, and the choice of the set-ups a jig or a
  coupler is validated against, for each type of microphone (6.7). The sweeps, maximum length sequences
  and random noise of B.3 to B.5 are `room.impulse_response`,
  `room.mls_impulse_response` and `electroacoustics.transfer_function`, and the
  synchronous averaging of B.6.2 is `signals.time_synchronous_average`, on
  their own pages; the qualification of the free field by ISO 26101 is on
  [its own page](https://docs.jmrp.io/phonometry/devices/emission/free-field-qualification/). The
  values of IEC/TS 61094-7, which are inputs here; the reciprocity calibration
  of the reference (IEC 61094-2, -3) is the subject of
  [Microphone calibration by reciprocity](https://docs.jmrp.io/phonometry/signals/metrology/reciprocity-calibration/). The second special case of Table
  D.1, a microphone calibrated as a system with its preamplifier (0.002 dB,
  and for the polarising voltage and the capacitance "frequency dependent" and
  under 0.02 dB above 200 Hz), is not held as data; a budget that needs it
  takes it through `additional_components`. A covariance that widens the
  uncertainty of the difference, as between two microphones of the same
  reciprocity set through the pairs with the third (IEC 61094-2:2009
  Formula (7)), is not taken by `verify_jig_or_coupler`.

## See also

- [Microphone calibration by reciprocity](https://docs.jmrp.io/phonometry/signals/metrology/reciprocity-calibration/):
  IEC 61094-2 and IEC 61094-3, which calibrate the laboratory standard
  microphone a comparison takes as its reference.
- [Free-field corrections of a sound level meter](https://docs.jmrp.io/phonometry/signals/metrology/free-field-corrections/):
  IEC 62585, which compares a meter with an LS2P reference in the same way.
- [Random-incidence and diffuse-field response](https://docs.jmrp.io/phonometry/signals/metrology/random-incidence/):
  IEC 61183, whose diffuse-field method is a comparison through the same
  model.
- [Measurement uncertainty](https://docs.jmrp.io/phonometry/signals/metrology/gum-uncertainty/):
  the GUM law of propagation the budgets are combined by.
- [Humid air](https://docs.jmrp.io/phonometry/fluids/humid-air/): the IEC 61094-2 Annex F air
  whose speed of sound sets the free-field region.
- [Calibration and dBFS](https://docs.jmrp.io/phonometry/signals/metrology/calibration/): the
  calibrator tone that turns a recording into pascals once a microphone's
  sensitivity is known.
- API reference: [`metrology.comparison_calibration`](https://docs.jmrp.io/phonometry/reference/api/metrology/comparison-calibration/).

## References

- **Asociación Española de Normalización. (2017).** *Electroacústica — Micrófonos de medida — Parte 5: Métodos para la calibración en presión de los micrófonos patrones de trabajo por comparación* (UNE-EN 61094-5:2017).
  The Spanish adoption of IEC 61094-5:2016, read beside it. Its Annex D leaves six standard uncertainties of Table D.1 empty and misprints the combined uncertainty of D.3, defects of the translation that the errata registry records; the library implements the English text.
- **Barham, R., Barrera-Figueroa, S., & Avison, J. E. M. (2014).** Secondary pressure calibration of measurement microphones. *Metrologia*, *51*(3), 129-138. https://doi.org/10.1088/0026-1394/51/3/129
  Reference \[1\] of IEC 61094-5:2016 and the origin of its Table A.1: the model of Formulas (1) to (5) for microphones of different diameters, with the inputs of Table 1, that diameter_sound_field_correction computes. Formula (4) is read less 1 and the annulus of Formula (2) from the overall radius, the readings that give its Table 2 (an erratum).
- **International Electrotechnical Commission. (2000).** *Measurement microphones – Part 1: Specifications for laboratory standard microphones* (IEC 61094-1:2000).
  6\.2.2 defines the equivalent volume of a microphone from its acoustic impedance, with the ratio of specific heats at the reference conditions taken as 1.40; Table 3 gives the 10 mm³ ± 5 mm³ of an LS2P. Read from BS EN 61094-1:2001, the English text of EN 61094-1:2000.
- **International Electrotechnical Commission. (2006).** *Measurement microphones — Part 7: Values for the difference between free-field and pressure sensitivity levels of laboratory standard microphones* (IEC/TS 61094-7:2006).
  The free-field to pressure sensitivity level difference a pressure-calibrated reference takes in a free-field calibration (IEC 61094-8 Table 1 and 8.2). The values on this page are illustrative, not read from it.
- **International Electrotechnical Commission. (2009).** [*Electroacoustics — Measurement microphones — Part 2: Primary method for pressure calibration of laboratory standard microphones by the reciprocity technique*](https://webstore.iec.ch/en/publication/4486) (IEC 61094-2:2009).
  Annex D gives the static pressure and temperature coefficients of laboratory standard microphones in dB/kPa and dB/K, the units the environmental correction takes them in; Annex F is the humid air whose speed of sound sets the free-field region of a time window; Formula (3) is the closed coupler and E.4 the lumped acoustic impedance of a microphone that the impedance ratio is written with.
- **International Electrotechnical Commission. (2012).** *Electroacoustics – Measurement microphones – Part 8: Methods for determining the free-field sensitivity of working standard microphones by comparison* (IEC 61094-8:2012).
  The free-field counterpart: the sequential substitution with a monitor microphone of 5.2 and A.2, the simultaneous excitation of 5.3, the phase of 5.1, the reference calibration options of Table 1, the uncertainty components of Table 2 with the expanded uncertainty of 8.8, and the time-selective processing of Annex B: the effective free-field region of Formula (B.1), the time window of B.1.3, the stepped sine of B.2 (Formulas (B.2) and (B.3)) and the pulse of the direct impulse method of B.6 (Formula (B.10)). Read from BS EN 61094-8:2012, the English text of EN 61094-8:2012, which is IEC 61094-8:2012 unchanged.
- **International Electrotechnical Commission. (2016).** *Electroacoustics – Measurement microphones – Part 5: Methods for pressure calibration of working standard microphones by comparison* (IEC 61094-5:2016).
  The implemented methods: the level model of D.2 and the phase of 5.1.1, the simultaneous excitation with the interchange of 5.1.2 and Annex C (Formulas (C.1) to (C.3)), the sequential excitation with a monitor of 5.1.3 and Annex B, the environmental correction of 6.6, the correction for microphones of different diameters of 6.5 by the model of its reference \[1\], the validation of a jig or a coupler of 6.7, the effect of different acoustic impedances of 7.4 and 7.5 with the air between two microphones of Table D.1 by the circuit of its reference \[2\], the WS3 corrections of Table A.1 and the example budget of Table D.1 combined as D.3 asks. Read from Edition 2.0 (2016-05), English-French.
- **Jarvis, D. R. (1996).** [*Methods for determining the pressure sensitivity of working standard microphones: a report on Euromet project A311*](https://eprintspublications.npl.co.uk/439/) (NPL Report CIRA(EXT) 010). National Physical Laboratory.
  Reference \[2\] of IEC 61094-5:2016: the circuit of Appendix B for the air between two microphones in a symmetric coupler, whose series impedance air_gap_series_impedance_pa_s_m3 gives, and its example of an LS2P against a high sensitivity WS2P. The published scan has lost the imaginary unit of the first microphone's mass term (an erratum).
- **Jarvis, D. R., & Watkins, S. (1997).** *Methods for determining the pressure sensitivity of IEC type WS3 measurement microphones* (NPL Report CIRA(EXT) 022). National Physical Laboratory.
  Jig calibrations of working standard microphones compared with reciprocity calibrations of the same microphones (Figures 6 and 7) and judged in words against the typical repeatability of the measurements, one standard deviation drawn as a band in Figure 7; it prints no numeric criterion for the validation of 6.7, no numbers one could be checked against, and nothing on what the two calibrations share.
- **Joint Committee for Guides in Metrology. (2008).** [*Evaluation of measurement data — Guide to the expression of uncertainty in measurement*](https://www.bipm.org/documents/20126/2071204/JCGM_100_2008_E.pdf) (JCGM 100:2008, the GUM). BIPM. https://doi.org/10.59161/JCGM100-2008E
  The law of propagation of uncertainty both budgets are combined by, and the coverage factor k = 2 both parts report with; its form for correlated inputs (5.2.2) with the covariance of two estimates of a common quantity (F.1.2.3) takes what two calibrations share out of the uncertainty of their difference, and the correlation coefficients of F.1.2.3 Example 2 are reproduced. Also published as ISO/IEC Guide 98-3:2008.