# Errata in ISO 9053

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/reference/errata/iso-9053/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/reference/errata/iso-9053/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-10-28
Updated: 2026-11-16T21:36:14Z
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Errata in ISO 9053: 1 defect in the printed text, with the evidence and what phonometry does about it.

---

This is the entry of the [errata registry](https://docs.jmrp.io/phonometry/reference/errata/) on ISO 9053-2:2020. The registry's introduction says how each claim is checked against the printed page and what the status of an entry means.

## ISO 9053-2:2020, Annex A.3 (two air properties credited to a document that does not print them)

- **Location:** Annex A.3, printed folio 13 (PDF page 17) for the first four
  values and printed folio 14 (PDF page 18) for the fifth.
- **The print:** "The following physical properties for air, valid at 23 °C,
  101,325 kPa and 50 % RH, are used for the calculation (values from
  IEC 61094-2:2009):", followed by $c_0 = 345{,}9$ m/s, $\rho_0 = 1{,}186$
  $\text{kg/m}^3$, $\kappa = 1{,}400\,8$, $k_\mathrm{a} = 0{,}023\,55$ J/(s·m·K) and,
  overleaf, $C_\mathrm{P} = 938{,}7$ J/(kg·K).
- **The problem:** two of the five are not IEC 61094-2:2009 values. Table F.1 of
  that standard (printed folio 40) tabulates exactly five quantities at this
  state: $\rho$, $c_0$, $\kappa$, $\eta$ and the thermal **diffusivity**
  $\alpha_t = 2.115\,317 \times 10^{-5}\ \text{m}^2\text{/s}$. It does not tabulate the
  thermal conductivity or the specific heat capacity; those appear in Annex F
  only as the two expressions under Clause F.6, which print no values. The three
  Annex A.3 values that do match are precisely the three Table F.1 cells rounded
  to four figures ($345.866\,52 \to 345.9$; $1.186\,084\,8 \to
  1.186$; $1.400\,757\,3 \to 1.400\,8$). The two that do not match are
  precisely the two quantities Table F.1 does not print: evaluated at the same
  state, Clause F.6 gives $k_\mathrm{a} = 0.025\,434\,1$ J/(s·m·K) and
  $C_\mathrm{P} = 1013.74$ J/(kg·K), each larger than the printed pair by the
  same factor 1.0800.

  The common factor is not a coincidence and not a unit difference. The pair is
  locked to the tabulated diffusivity: $0.023\,55 / (1.186 \times
  2.115\,317 \times 10^{-5}) = 938.708\,5$, which prints as 938.7. So one
  of the two came from elsewhere and the other was computed back through
  Formula (F.5) to keep $\alpha_t$ right. Which one is foreign is settled by
  thermodynamics rather than by preference: $C_\mathrm{P} = 938.7$ J/(kg·K) is
  27.19 J/(mol·K), below the rigid-rotor diatomic floor $(7/2)R = 29.10$
  J/(mol·K), so it is not air at any temperature, in any unit, per mass or per
  mole, and the Annex F expression for $C_\mathrm{P}$ never falls below about
  1013 J/(kg·K) anywhere from 200 K to 400 K. The conductivity 0.023 55
  J/(s·m·K), by contrast, is a real conductivity of air: it is what the Annex F
  expression gives near −1.4 °C, outside the 15 °C to 27 °C domain Annex F
  prints for itself.
- **Consequence for the annex's own example:** none. Formula (A.5) uses
  $k_\mathrm{a}$ and $C_\mathrm{P}$ only through the combination
  $k_\mathrm{a}/(\rho_0 c_0 C_\mathrm{P})$, and the common factor cancels
  there, so both pairs give the printed $b = 1.83 \times 10^{-3}$ m and
  $\kappa' = 1.370$. The defect is invisible inside Annex A.3 and appears only
  when either constant is read out on its own, as a document credited with
  publishing it.
- **Evidence:** the two printed pages against IEC 61094-2:2009 Table F.1
  (printed folio 40) and Clause F.6 (printed folio 39); the Clause F.6
  expressions evaluated at 23 °C, 101 325 Pa and 50 % RH, which reproduce the
  printed $\alpha_t$ to $1.0 \times 10^{-7}$ relative; the molar heat
  capacity implied by 938.7 J/(kg·K) against the diatomic floor. IEC 61094-2:2009
  is not a normative reference of ISO 9053-2:2020; it appears only as
  Bibliography item [4]. Verified on PDF page 17 (printed p. 13) and PDF page 18
  (printed p. 14) of ISO 9053-2:2020, and on PDF page 42 (printed p. 40) and PDF
  page 41 (printed p. 39) of BS EN 61094-2:2009.
- **Library behaviour:** the conformance rows that reproduce Annex A.3 pass the
  five values the annex prints, so they reproduce the standard rather than merely
  agree with it. The defaults a caller receives are the same air state computed
  from IEC 61094-2:2009 Annex F, which is what the annex says it is using; both
  land on the printed $b$ and $\kappa'$ ([`tests/reference_data/`](https://github.com/jmrplens/phonometry/blob/main/tests/reference_data),
  conformance checks "ISO 9053-2:2020 Annex A.3").
- **Status:** unreported.