# Errata in ISO 8041

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/reference/errata/iso-8041/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/reference/errata/iso-8041/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-10-28
Updated: 2026-11-16T21:36:14Z
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Errata in ISO 8041: 4 defects in the printed text, with the evidence and what phonometry does about each.

---

These are the 4 entries of the [errata registry](https://docs.jmrp.io/phonometry/reference/errata/) on ISO 8041-1:2017 and ISO 8041-2:2021, in the order the registry lists them. The registry's introduction says how each claim is checked against the printed page and what the status of an entry means.

## ISO 8041-1:2017, clause 12.7 ("the appropriate weighting factor (see Table 1)" for `Wf`)

- **Location:** clause 12.7, printed folio 30 (PDF page 30 of the ISO release,
  PDF page 38 of the copy read here), fourth paragraph. The clause opens on
  printed folio 29.
- **The print:** "For each frequency weighting provided, a steady sinusoidal
  electrical signal shall be applied to the electrical input facility at the
  appropriate reference frequency. With an input signal adjusted to indicate
  the reference vibration value on the reference measurement range with
  band-limiting frequency weighting, the indicated frequency-weighted
  vibration values shall equal the indicated band-limited weighted vibration
  value multiplied by the appropriate weighting factor (see Table 1) within
  the tolerance limits of Table 2."
- **The problem:** the pointer to Table 1 names a quantity the test cannot be
  satisfied with. The test fixes the input so that the *band-limited*
  indication reads the reference value, so the frequency-weighted indication a
  conforming meter shows is
  $a_\mathrm{ref}\,|H(f_\mathrm{ref})| / |H_\mathrm{BL}(f_\mathrm{ref})|$: the
  factor that closes the identity is the
  **ratio** of the two responses at the reference frequency, not the overall
  weighting Table 1 prints. For eight of the nine weightings the distinction is
  invisible, because their band-limiting weighting sits between 0.999 68 and
  0.999 97 at their own reference frequency and the two readings agree to
  0.03 %. `Wf` is the exception: its reference frequency of 2.5 rad/s =
  0.397 887 Hz falls inside its own band-limiting skirt, whose corners Table 3
  puts at 0.08 Hz and 0.63 Hz. There the band-limiting weighting is 0.928 078
  and the overall weighting 0.388 848, which Table B.5 prints as 0.927 9 and
  0.388 4 at the neighbouring 0.398 1 Hz band centre. Read as the 0.388 8 of
  Table 1, the row demands a value that a conforming `Wf` meter's indication
  exceeds by 7.76 % of that demanded value ($0.418\,982 / 0.388\,8 - 1$),
  against the ±5 % Table 2 allows low-frequency whole-body vibration: half
  again over the limit, on an instrument with no defect. Read as the ratio
  0.418 982, the row is true by construction.
- **Evidence:** the printed clause against Table 1 (printed folio 9), Table 2
  (printed folio 12), Table 3 (printed folios 12 to 13) and Table B.5. The two
  responses at 2.5 rad/s are evaluated from the Formula (1) to (5) cascade the
  same Table 3 parameters define, and they reproduce the two Table B.5 columns
  at the neighbouring band centre to four figures. Verified on PDF page 38
  (printed p. 30) of ISO 8041-1:2017(E).
- **Consequence for the standard's own tables:** none. Annex B tabulates the
  band-limiting weighting and the overall weighting in separate columns, so
  both readings can be recovered from it; only clause 12.7's one-line
  instruction is ambiguous.
- **Library behaviour:**
  [`band_limited_weighting_factor`](https://github.com/jmrplens/phonometry/blob/main/src/phonometry/vibration/human/instrumentation.py)
  returns the ratio, which is the reading that makes the test satisfiable, and
  its docstring tabulates the two readings side by side for all nine
  weightings so a report can say which one it used. `reference_indication`
  returns the Table 1 product, which is the other quantity and the one the
  reference-conditions row of Table 1 is about.
- **Status:** unreported.

## ISO 8041-2:2021, clause 12.7 ("the appropriate weighting factor (see Table 1)" for `Wf`, carried over from Part 1)

- **Location:** clause 12.7, printed folio 26 (PDF page 34), the second
  paragraph on that folio. The clause opens on printed folio 25.
- **The print:** the sentence of ISO 8041-1:2017 12.7 recorded in the entry
  above, word for word: "With an input signal adjusted to indicate the
  reference vibration value on the reference measurement range with
  band-limiting frequency weighting, the indicated frequency-weighted
  vibration values shall equal the indicated band-limited weighted vibration
  value multiplied by the appropriate weighting factor (see Table 1) within
  the tolerance limits of Table 2." Table 1 of this document prints the same
  weighting factor for `Wf` at 2,5 rad/s, 0,388 8, and Table 3 the same `Wf`
  parameters, with its band-limiting corners at 0,08 Hz and 0,63 Hz.
- **The problem:** the ISO 8041-1:2017 defect, carried into the personal
  vibration exposure meter. Part 2 keeps `Wf` among the weightings a PVEM may
  provide, with the same reference frequency inside the same band-limiting
  skirt, so the arithmetic of the entry above holds unchanged: read as the
  0.388 8 of Table 1, the paragraph demands a value that a conforming `Wf`
  meter's indication exceeds by 7.76 % of that demanded value
  ($0.418\,982 / 0.388\,8 - 1$), outside the ±3 % that the second row of
  Table 2 allows the difference and outside the ±5 % its first row allows a
  low-frequency whole-body indication. Read as the ratio of the overall
  weighting to the band-limiting one at the reference frequency, 0.418 982,
  the paragraph is true by construction.
- **Evidence:** the printed clause against Table 1 (printed folio 5) and
  Tables 2 and 3 (printed folio 8), whose `Wf` row carries the parameters
  ISO 8041-1:2017 prints. Verified on PDF pages 13, 16 and 34 (printed pp. 5,
  8 and 26) of ISO 8041-2:2021(E).
- **Consequence for the standard's own tables:** none, as in Part 1.
- **Library behaviour:** unchanged by this entry.
  [`band_limited_weighting_factor`](https://github.com/jmrplens/phonometry/blob/main/src/phonometry/vibration/human/instrumentation.py)
  returns the ratio, and it serves both parts because Part 2 prints the same
  weightings. The conformance row "ISO 8041-2:2021 Table 2" records the
  tolerances this paragraph is judged against.
- **Status:** unreported.

## ISO 8041-2:2021, clause 12.7 (time weightings graded against a row Table 2 no longer prints)

- **Location:** clause 12.7, printed folio 26 (PDF page 34), the third
  paragraph on that folio; read against Table 2 on printed folio 8, clause
  5.13 on printed folio 15, clause 5.1.2 on printed folio 6 and Table 8 on
  printed folios 13 and 14.
- **The print:** "For an instrument where time weightings are provided, a
  steady sinusoidal electrical signal shall be applied to the electrical input
  facility at the reference frequency. [...] With the same input signal, the
  indicated vibration values on each time weighting shall equal the indicated
  reference vibration value within the tolerance limits of Table 2." Table 2
  prints two rows: the tolerance of indication at the reference frequency
  (±4 %, and ±5 % for low-frequency whole-body vibration) and the difference
  between a frequency-weighted indication and the band-limited one times the
  weighting factor (±3 %). Clause 5.13, "Running RMS acceleration", reads in
  full: "Not applicable for PVEM."
- **The problem:** the paragraph is the one ISO 8041-1:2017 prints in its own
  12.7 (printed folio 30 of that document), with "the vibration meter" turned
  into "the PVEM", and there the limit it points at is the third row of the
  Part 1 Table 2: the running r.m.s. indication against the linear
  time-averaged one, ±2 %. The time weighting in this family of standards is
  that running r.m.s.; Part 1 titles its Tables 10 and 11 "Time-weighting
  decay rates". Part 2 dropped the row, since its 5.13 declares the running
  r.m.s. not applicable, and kept the paragraph that grades it, so the
  paragraph now sends the reader to a table in which no row states a limit for
  a time-weighted indication. And the document itself lets a PVEM carry one:
  its 5.1.2 (printed folio 6) allows a whole-body PVEM to "optionally, measure
  exposure characteristics based on maximum transient vibration value (MTVV)",
  which ISO 8041-1:2017 3.1.2.4, adopted by clause 3, defines as the "maximum
  value of the running r.m.s. vibration acceleration value when the
  integration time is equal to 1 s", and its Table 8 (printed folios 13 and
  14) grades the "MTVV linear" and "MTVV exponential" of that running r.m.s.
  in the burst. For a PVEM that provides no time weighting the paragraph is
  empty; for one that reports the MTVV, the document does not say what limit
  the indication on each time weighting is held to.
- **Evidence:** the three printed passages against each other, against 5.1.2
  and Table 8, and against ISO 8041-1:2017 12.7 and its Table 2, which print
  the running r.m.s. row this paragraph was written for. Verified on PDF
  pages 14, 16, 21, 22, 23 and 34 (printed pp. 6, 8, 13, 14, 15 and 26) of
  ISO 8041-2:2021(E), and on PDF pages 13, 20, 28 and 38 (printed pp. 5, 12,
  20 and 30) of ISO 8041-1:2017(E) for the definition of the MTVV, the Part 1
  Table 2, the title of its Table 10 and its 12.7.
- **Consequence for the standard's own tables:** none. The defect is a
  cross-reference that outlived its row.
- **Library behaviour:**
  [`PVEM_INDICATION_TOLERANCES_PERCENT`](https://github.com/jmrplens/phonometry/blob/main/src/phonometry/vibration/human/instrumentation.py)
  publishes the two rows Part 2 prints and no running r.m.s. row, so nothing
  in the library grades a PVEM on this paragraph; the Part 1 row remains
  `RUNNING_RMS_CONSISTENCY_TOLERANCE_PERCENT`, documented as Part 1 only. The
  conformance row "ISO 8041-2:2021 Table 2" records the two rows.
- **Status:** unreported.

## ISO 8041-2:2021, clause 12.22 ("exited" for "excited")

- **Location:** clause 12.22, "Logging capabilities", printed folio 36 (PDF
  page 44), the first paragraph of the clause.
- **The print:** "Part A of the PVEM shall be placed on a shaker and be exited
  2 times for at least 300 s each."
- **The problem:** "exited" where the sense is "excited": the part is put on a
  shaker to be vibrated twice, once at each end of the 12 h run, and the
  paragraph that follows on printed folio 37 counts the 600 logged samples
  that "correspond to the vibration magnitude", which is those two 300 s
  excitations at one sample per second. As printed, the verb says the part
  leaves the shaker twice.
- **Evidence:** PDF page 44 (printed p. 36) of ISO 8041-2:2021(E), and PDF
  page 45 (printed p. 37) for the 600 samples.
- **Consequence for the standard's own tables:** none.
- **Library behaviour:** unaffected. The 12 h logging test of 12.22 is a test
  on a physical meter and is not implemented.
- **Status:** unreported (typographic, no numerical consequence).