# Errata in DIN 45669

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/reference/errata/din-45669/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/reference/errata/din-45669/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-10-28
Updated: 2026-11-16T21:36:14Z
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Errata in DIN 45669: 2 defects in the printed text, with the evidence and what phonometry does about each.

---

These are the 2 entries of the [errata registry](https://docs.jmrp.io/phonometry/reference/errata/) on DIN 45669-1:2010-09 and DIN 45669-2:2005-06, in the order the registry lists them. The registry's introduction says how each claim is checked against the printed page and what the status of an entry means.

## DIN 45669-1:2010-09, Table 9 (a peak-velocity row that contradicts Formula (5), and the $KB_\mathrm{F}$ row beside it)

- **Location:** Table 9, printed folio 35 (PDF page 35 of the copy read here,
  which prints its folio numbers without an offset), rows "|v|max in mm/s bei
  f_u = 1 Hz und f_o = 80 Hz" and "$KB_\mathrm{F}(t)$ ± 2 % Schwankung".
- **The print:** for a sinusoidal input at the test frequencies, the peak row
  reads 0,852 at 1 Hz, 1,000 at 5,6 Hz, 1,000 at 31,5 Hz, 0,843 at 80 Hz and
  0,249 at 315 Hz; the $KB_\mathrm{F}$ row of the same five columns reads 0,103, 0,500,
  0,693, 0,594 and 0,071.
- **The problem:** the two rows are computed on different band limitations, and
  the peak row does not follow the standard's own Formula (5). With
  $f_u = 1$ Hz and $f_o = 80$ Hz that formula gives $|H_{u\mathrm{Soll}}|$ =
  0.995 at 31.5 Hz and 0.100 at 315 Hz, against the 1.000 and 0.249 printed. The
  $KB_\mathrm{F}$ row settles which of the two is the intended reading: $KB_\mathrm{F}$ is
  $|H_{B\mathrm{Soll}}|/\sqrt{2}$ for a 1 mm/s sine, and at 31.5 Hz that is
  0.6928 from 0.995 and 0.6962 from 1.000, so the printed 0.693 is the first;
  at 315 Hz it is 0.0709 from 0.100 and 0.176 from 0.249, so the printed 0.071
  is again the first, by a factor of two and a half. The 0.852 at 1 Hz is the
  same kind of departure at the other end, against the 0.842 of Formula (5).
- **Evidence:** the printed table against Formulae (5) and (6) on printed folio
  18 and the note under Formula (3) that puts the two corners at 0.8 Hz and
  100 Hz. Verified on PDF page 35 (printed p. 35) of DIN 45669-1:2010-09; the
  two rows are adjacent cells of one column, so no
  offset or transcription question arises. What the two anomalous cells do
  match is the maximum a max-hold display shows when the switch-on transient of
  the band limitation is included: a 1 mm/s sine started at a zero crossing
  gives 0.852 at 1 Hz and 0.248 at 315 Hz through the same filter. That
  reading, though, is not what the other three cells of the row show, so the
  row is not consistently one convention or the other.
- **Consequence for the standard's own tables:** confined to that row.
  Berichtigung 1:2012-12 rewrites Table 8 and does not touch Table 9, and the
  reference indications of 6.2.3.12, which are the values the same test signal
  produces at 16 Hz, are reproduced exactly by the formulas.
- **Library behaviour:**
  [`KB_TEST_INDICATIONS`](https://github.com/jmrplens/phonometry/blob/main/src/phonometry/vibration/immission/vibration_meter.py)
  publishes the three rows of Table 9 that follow the formulas, and the
  conformance report runs a 1 mm/s sine through the whole chain and reproduces
  all fifteen of those values to the three decimals they are printed with. The
  peak row is not published and not checked; the module docstring says why, and
  the peak a record shows is computed from the record rather than from a table.
- **Status:** unreported.

## DIN 45669-2:2005-06, Clause 5.1.4 (a coupling clause cited by the wrong number)

- **Location:** Clause 5.1.4, last paragraph of printed page 6 (PDF page 6 of
  the copy read here, which prints its folios without an offset).
- **The print:** "Bei Messungen am Erdreich sollten für die Schwingungsaufnehmer
  die Ankopplungsverfahren nach 5.3.3 angewandt und müssen die durch die
  Ankopplung verursachten Messabweichungen nach 8.2.3 beachtet werden": for
  measurements on the ground, the coupling methods of 5.3.3 apply and the
  deviations of 8.2.3 are to be observed.
- **The problem:** 5.3.3 is "Ankopplung bei weichen Unterlagen", the coupling
  on soft floor coverings, whose method is the spiked device of Figure 1 a)
  pressed through a carpet. The coupling to the ground is 5.3.4, "Ankopplung
  an das Erdreich", and its methods, the stake, the buried transducer, the
  borehole and the bedded plate, are 5.3.4.2 with Table 2. The other reference
  in the same sentence, 8.2.3, is "Ankopplung an das Erdreich" and points back
  to 5.3.4, which is what the first reference was meant to be.
- **Evidence:** the sentence on printed page 6, the heading of 5.3.3 on
  printed page 8 and the heading of 5.3.4 on printed page 9. Verified on PDF
  page 6 (printed p. 6), PDF page 8 (printed p. 8) and PDF page 9 (printed
  p. 9) of DIN 45669-2:2005-06.
- **Consequence for the standard's own tables:** none; a reader who follows
  the number lands on the carpet device instead of on Table 2.
- **Library behaviour:**
  [`vibration.immission.coupling`](https://github.com/jmrplens/phonometry/blob/main/src/phonometry/vibration/immission/coupling.py)
  carries the loose-mounting limits of 5.3.2 and 5.3.3 and the ground
  deviation of 5.3.4.1, and its docstring names each by the clause that
  prints it.
- **Status:** not reported.