# Audiometric Test Methods (ISO 8253-1 and -2)

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/perception/hearing/audiometry/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/perception/hearing/audiometry/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-10-01
Updated: 2026-11-16T21:36:13Z
Standards: ISO 8253-1:2010; ISO 8253-2:2009
Verified: 19 conformance checks (ISO 8253-1:2010; ISO 8253-2:2009) https://docs.jmrp.io/phonometry/reference/conformance/#audiometric-test-methods-iso-8253
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Audiometric rooms and thresholds: room noise and sound field (ISO 8253-1, -2), four test methods for the hearing threshold level, and its uncertainty.

---

An audiogram is a set of hearing threshold levels, and each of them rests on
three things the audiometer does not show: a room quiet enough not to mask the
faintest tone, a rule that turns a series of "heard" and "not heard" into one
number, and an uncertainty that says how far that number can be trusted.
ISO 8253-1 sets all three out for earphones and bone vibrators, and ISO 8253-2
adds what a loudspeaker needs: a room whose sound field is known.

## Is the room quiet enough? (ISO 8253-1 Clause 11)

The ambient noise must not mask the lowest test tone. Tables 2 and 4 give the
maximum permissible ambient sound pressure level $L_{S,\mathrm{max}}$ in each
one-third-octave band from 31.5 Hz to 8 kHz, for air conduction with typical
supra-aural earphones and for bone conduction, and for the lowest frequency the
audiometry tests: 125 Hz, 250 Hz or, with earphones, 500 Hz. They hold for
hearing threshold levels down to 0 dB with at most +2 dB of threshold shift
from the noise. Three things move them (11.1 and the NOTEs to the tables):

$$
L_{S,\mathrm{max}}' = L_{S,\mathrm{max}}
+ \left(A_\mathrm{earphone} - A_\mathrm{supra\text{-}aural}\right)
+ L_\mathrm{HT,min} + \Delta
$$

an earphone that excludes more noise than the supra-aural one allows the
difference of the two attenuations in Table 3; a lowest hearing threshold level
$L_\mathrm{HT,min}$ other than 0 dB is added band by band; and accepting a
threshold shift of +5 dB instead of +2 dB adds $\Delta$ = 8 dB.

```python
import numpy as np
from phonometry import hearing

# One-third-octave ambient levels of a single-walled booth, 31.5 Hz to 8 kHz.
booth = [50, 48, 45, 43, 40, 36, 30, 24, 19, 16, 14, 12, 11,
         10, 10, 10, 9, 9, 9, 8, 8, 8, 8, 8, 8]
supra_aural = hearing.check_audiometric_ambient_noise(booth)
print(supra_aural.passes)                               # False
print(supra_aural.frequencies[~supra_aural.within])     # [ 63.  80. 100. 125. 160.]
print(supra_aural.lowest_measurable_hearing_level)   # 3.0

# The same booth with insert earphones (Table 3: about 33 dB at low frequency).
insert = hearing.check_audiometric_ambient_noise(booth, earphone="ER-3A")
print(insert.passes)                                    # True
print(hearing.ambient_noise_limits("air", earphone="ER-3A")[15])  # 45.0 at 1 kHz
```

The booth's ventilation leaves too much at 63 Hz to 160 Hz for supra-aural
earphones, by up to 3 dB at 100 Hz, so with them it can only measure down to a
hearing level of 3 dB: the lowest level the rule of 11.1 allows this room.
ISO 8253-1 asks no report to state it; ISO 8253-2 11.1 f) asks a sound field
report to state the same figure when it is not 0 dB. Insert earphones exclude
that noise, and the same room qualifies. Another earphone enters with its own
attenuation per band, `earphone_attenuation=`.

![Two panels of the same booth's one-third-octave ambient spectrum from 31.5 Hz to 8 kHz, falling from 50 dB at 31.5 Hz to about 8 dB above 2 kHz, drawn against the maximum permissible level for air conduction audiometry from 125 Hz. Left, with supra-aural earphones: the limit falls from 56 dB to 18 dB between 31.5 Hz and 500 Hz and rises to 36 dB at 4 kHz, and the spectrum crosses it at 63, 80, 100, 125 and 160 Hz, each marked with a cross; the room does not qualify. Right, with ER-3A insert earphones: the limit is raised by the difference in attenuation, to between 37 dB and 89 dB, and the spectrum stays well under it everywhere; the room qualifies](https://docs.jmrp.io/phonometry/media/audiometry_test_room.svg)

*One booth, two earphones. The limits of Table 2 are what a supra-aural
earphone lets through; an insert earphone excludes about 30 dB more below
500 Hz, and Table 3 turns that into room.*

**Show the code for this figure**

```python
import matplotlib.pyplot as plt

fig, (left, right) = plt.subplots(1, 2, figsize=(13.5, 5.6), sharey=True)
supra_aural.plot(left)   # limits, measured spectrum, crosses where it exceeds
insert.plot(right)
plt.show()
```

Bone conduction leaves both ears open, so no earphone shelters them and Table 4
is stricter than Table 2 by 10 dB at 500 Hz and by 34 dB at 4 kHz; NOTE 2 to it
warns that levels under 5 dB are hard to measure with most sound level meters.
The same
booth, for bone conduction from 250 Hz, allows only a lowest hearing level of
6 dB, set by the 4 kHz band:

```python
bone = hearing.check_audiometric_ambient_noise(
    booth, presentation="bone", lowest_test_frequency_hz=250
)
print(bone.passes, bone.lowest_measurable_hearing_level)   # False 6.0
print(bone.frequencies[np.argmax(bone.exceedance_db)])        # 4000.0

# Measuring only down to 10 dB raises every limit by 10 dB (11.1).
print(hearing.check_audiometric_ambient_noise(
    booth, presentation="bone", lowest_test_frequency_hz=250,
    lowest_hearing_level=10.0,
).passes)                                                     # True
```

### How the room measurement goes (11.1, 11.2)

A verdict on the room is worth what the spectrum set against the limits is
worth, and the last paragraph of 11.1 says how that spectrum is taken:

- at a time when conditions are representative of those in which the
  audiometric tests are carried out, with the ventilation system operating if
  it usually operates during testing;
- at the position the head of the test subject would occupy in the room, with
  the subject not present;
- with an instrument that meets the requirements for class 1 sound level
  meters of IEC 61672-1 and, for its one-third-octave bands, those of the
  filter standard IEC 61260, which the clause prints as "IEC 61620" (the
  transposed digits are registered in
  [ERRATA](https://docs.jmrp.io/phonometry/reference/errata/));
- with a noise floor at least 6 dB below the sound pressure level being
  measured.

A measurement that leaves bands out is judged on those it has and does not
pass: the room is not shown to meet the bands nobody measured. `noise_floor_levels=`
flags the bands where the reading is within 6 dB of the meter's own floor,
which 11.1 forbids; such a reading overstates the room, so a band within its
limit still is, and one over it may be the meter's.

Where the sound pressure levels cannot be measured, 11.2 allows a
psycho-acoustic check in their place: an audiometric test on at least two
subjects known to have stable audiograms and, at every frequency, hearing
threshold levels lower than the lowest hearing levels to be used in regular
testing. Thresholds that come out higher by 5 dB or more show that the noise
in the room has to be reduced. The check is made by bone conduction if bone
conduction audiometry is performed in the room, and during the time in which
the audiometry would normally be conducted. It is a procedure for the tester,
and nothing in the library runs it.

## From responses to a threshold (6.2 to 7.5)

Every method ends in a rule, and each rule has its own test of reliability: a
spread of more than 10 dB between the levels it combines.

### The ascending method (6.2.3.2, 6.2.4.2)

The first tone sits 10 dB below the lowest level the subject answered during
familiarization; after no response the level rises 5 dB, after a response it
drops 10 dB. An **ascent** is a run of rising tones that ends in a response,
and the series stops when three ascents end at one level, out of at most five.
The hearing threshold level is the lowest level at which responses occur in
more than half of the ascents:

```python
series = hearing.ascending_method_threshold(
    presentation_levels=[30, 20, 25, 15, 20, 25, 15, 20, 10, 15, 20, 25],
    responses=[1, 0, 1, 0, 0, 1, 0, 1, 0, 0, 0, 1],
    familiarization_level=40.0,
)
print(series.ascent_levels)   # [25. 25. 20. 25.]
print(series.threshold_level)       # 25.0
print(series.doubtful)           # False: the ascents span 5 dB, not over 10 dB

# Replayed as it happens, the same rule says what to present next.
print(hearing.ascending_method_threshold(
    presentation_levels=[30, 20], responses=[True, False],
    familiarization_level=40.0,
).next_level)                 # 25.0
```

The sequence is checked against the steps of 6.2.3.2, and a tone that breaks
them is refused with its position, which is what lets the function supply the
next level to a computer-controlled audiometer (6.4): call it with the
presentations so far and present `next_level`. The first tone, heard at once, is not an ascent: no
tone rose to it. After five ascents without three at one level the series is
exhausted and the next starts 10 dB above the last response. The response levels
of the ascents can go in alone, and `shortened=True` applies the version that
stops at two responses out of three ascents. 6.2.3.2 does not say what follows
a shortened series that ends its three ascents without two responses at one
level; the library continues it with the full method, a new call on the same
presentations with `shortened=False`.

### The bracketing method (6.2.4.3)

Ascents and descents alternate in 5 dB steps. The lowest response levels of the
ascents are averaged, those of the descents are averaged, and the mean of the
two averages, rounded to the nearest 5 dB step, is the threshold. A mean
halfway between two steps goes to the higher one, the library's reading of
"nearest" wherever a standard leaves the tie open:

```python
bracket = hearing.bracketing_method_threshold([30, 30, 35], [30, 35, 35])
print(bracket.mean_level, bracket.threshold_level)   # 32.5 35.0
print(bracket.complete, bracket.repeat_advised)  # True False
```

Three ascents and three descents make a complete series; two of each suffice
when their four levels are within 5 dB of each other.

### Automatic recording (6.3.5)

The subject holds the level down while the tone is heard and lets it rise when
it is not, and the tracing zigzags about the threshold. The first reversal after
a change of frequency is ignored, and so are both reversals of any excursion of
3 dB or less; the peaks and the valleys are averaged separately, and the mean of
the two averages, **rounded up** to the next whole decibel, is the threshold:

```python
tracing = hearing.automatic_audiometry_threshold(
    [38, 22, 31, 21, 32, 20, 30, 29, 31, 19, 30, 20, 31, 21]
)
print(tracing.peak_levels)                      # [31. 32. 30. 31.]
print(tracing.valley_levels)                    # [22. 21. 20. 19. 20. 21.]
print(tracing.mean_level, tracing.threshold_level) # 25.75 26.0
```

The recording is of doubtful reliability when the peaks or the valleys spread
over more than 10 dB or fewer than six reversals remain. NOTE 2 to 6.3.5 puts
automatic thresholds 3 dB below manual ones on average, a property of the method
this function does not correct.

![Left: twelve presentations of the ascending method as a staircase of hearing level against presentation number, starting at 30 dB, heard tones drawn filled and missed ones open, and the four ascents' ends ringed at 25, 25, 20 and 25 dB; a dashed line marks the threshold of 25 dB. Right: an automatic recording as fourteen reversals zigzagging between about 20 dB and 32 dB, the first reversal and three around a 1 dB wobble drawn grey as ignored, dotted lines at the mean of the peaks, 31 dB, and of the valleys, 20.5 dB, and a dashed line at the threshold of 26 dB](https://docs.jmrp.io/phonometry/media/audiometry_threshold_rules.svg)

*Left, the ascending method: three of the four ascents end at 25 dB, which is
more than half. Right, automatic recording: the peaks average 31 dB and the
valleys 20.5 dB, and the mean of 25.75 dB is rounded up.*

**Show the code for this figure**

```python
fig, (left, right) = plt.subplots(1, 2, figsize=(13.5, 5.6))
series.plot(left)    # the staircase, heard and missed, ascents ringed
tracing.plot(right)  # the tracing, its ignored reversals and the averages
plt.show()
```

### Sweep-frequency audiometry (7.5)

The frequency sweeps, normally at 0.5 to 2 octaves per minute, while the
subject tracks the threshold. At a specified frequency the three peaks and the
three valleys closest to it are averaged, closest on the logarithmic axis the
sweep runs on, and the mean is rounded to the nearest decibel. Run along the
tracing, six consecutive reversals give the threshold at the geometric mean of
their frequencies:

```python
reversal_frequencies = np.geomspace(500.0, 4000.0, 18)
reversal_levels = [34, 24, 33, 23, 35, 25, 36, 24, 38, 27, 40, 29,
                   43, 31, 46, 34, 48, 37]
sweep = hearing.sweep_audiometry_threshold(reversal_levels,
                                           reversal_frequencies=reversal_frequencies)
print(sweep.frequencies)     # [ 500.  750. 1000. 1500. 2000. 3000. 4000.]
print(sweep.threshold_level)    # [29. 29. 30. 32. 36. 40. 40.]
print(np.round(sweep.running_frequencies[:2]))  # [679. 767.]
```

### The repeat at 1 kHz and the cautions (6.2.3.2, 8.4)

Three more rules judge the levels once they are found. When every frequency of
the first ear is done, Step 3 of 6.2.3.2 repeats the measurement at 1 kHz: an
agreement to 5 dB or less moves the test to the other ear, and an improvement
or a worsening of 10 dB or more sends it back to retest the further
frequencies, in the same order, until the agreement returns. A hearing level
of 40 dB or more in either ear calls for caution because the tone may be heard
by the other ear, and contralateral masking can then be necessary (6.2.3.2).
And a bone vibrator on the mastoid is felt, on average, at a hearing level of
about 40 dB at 250 Hz, 60 dB at 500 Hz and 70 dB at 1 kHz (8.4, about 10 dB
lower when the audiometer is calibrated for the forehead), so a
bone-conduction response there may be touch rather than hearing:

```python
repeat = hearing.check_retest_agreement(25.0, 35.0)   # first and repeat at 1 kHz
print(repeat.passes, repeat.retest_further_frequencies)   # False True

cautions = hearing.audiogram_cautions(
    [250, 500, 1000, 2000, 4000],
    air_conduction_threshold_levels=[20, 25, 30, 45, 55],
    bone_conduction_threshold_levels=[40, 20, 25, 40, 50],
)
print(cautions.cross_hearing)   # [False False False  True  True]
print(cautions.vibrotactile)    # [ True False False False False]
```

The text is written for the 5 dB steps of a manual audiometer, where a repeat
either agrees to 5 dB or differs by 10 dB. A finer control can give 7 dB, which
is not agreement, and for which Step 3 names no action; the check then neither
passes nor asks for the retest. 8.4 gives levels at three frequencies only and
warns that individuals vary widely about them, so a flag there is a reason to
look again, not a verdict; bone conduction gets no cross-hearing flag, because
8.1 already asks for the other ear to be masked at every level.
`cautions.plot()` draws the audiogram with the flagged levels crossed.

## The uncertainty of a threshold (Annex A)

Annex A models the hearing threshold level as the determined value plus seven
zero-mean input quantities, for the audiometer, the transducer and its fitting,
the ambient noise, the masking, the tester, the subject and an unusually
difficult measurement. None is correlated with another and each has a
sensitivity coefficient of 1, so

$$
u = \sqrt{\sum_{i=1}^{8} u_i^2}, \qquad U = 2u
$$

and A.3 gives typical values: 2.5 dB for repeated air-conduction
determinations up to 4 kHz, the audiometer's ±3 dB and its 5 dB steps as two
rectangular distributions, $\sqrt{1.5^2 + 2.5^2}$ = 2.9 dB for the transducer,
and 2 dB for a room that meets Clause 11. That is the example of Table A.2:

```python
budget = hearing.audiometric_uncertainty(1000.0)
print(np.round(budget.standard_uncertainties_db[:4], 1))           # [2.5 2.3 2.9 2. ]
print(round(budget.combined_uncertainty_db, 1), round(budget.expanded_uncertainty_db))  # 4.9 10

# Bone conduction above 4 kHz with masking noise: every component grows.
masked_bone = hearing.audiometric_uncertainty(6000.0, conduction="bone", masked=True)
print(round(masked_bone.expanded_uncertainty_db, 1))               # 15.3
```

An expanded uncertainty of 10 dB is two audiometer steps either way, which is
what a single audiogram can claim. `environment_uncertainty_db=` replaces the 2 dB of A.3.5
for a room that exceeds Clause 11, or drops it for thresholds well above 0 dB,
and `tester_uncertainty_db=`, `subject_uncertainty_db=` and `special_uncertainty_db=` carry the exceptional
components of A.3.7 to A.3.9; `budget.plot()` draws the eight as bars.

## The sound field (ISO 8253-2)

With loudspeakers the room is part of the instrument, so ISO 8253-2 qualifies
it, with the subject and the chair absent and the audiometry's own signals, as
one of three fields. Every level is read on a sound level meter that complies
with IEC 61672-1, type 1, except the readings a directional microphone takes
in the diffuse field (5.1):

- a **free field** (5.2): the loudspeaker at least 1 m away, the level 0.15 m
  to the left, right, above and below the reference point within ±1 dB of the
  level at it up to 4 kHz and ±2 dB above, the ear-side positions within 3 dB
  of each other above 4 kHz, and the points 0.15 m in front of and behind it on
  the loudspeaker's axis within ±1 dB of the inverse distance law,
  $20 \lg\left((r + d)/(r - d)\right)$. Only an anechoic room meets it, and pure
  tones need it (4.2);
- a **quasi-free field** (5.4): the same with ±2 dB at the lateral positions and
  the axial points 0.10 m out. Its usable frequency range is the range where the
  requirements hold;
- a **diffuse field** (5.3): six positions 0.15 m out within ±2.5 dB and the ear
  sides within 3 dB, read with an omnidirectional microphone kept in the same
  orientation at every position, and from 500 Hz up the levels a directional
  microphone reads in the directions of the most and the least incident energy
  within the variation Table 1 allows for its front-to-random index: 5 dB for
  an index of 5 dB or more, 4.5 dB for 4.5 dB, 4 dB for 4 dB. The notes to the
  table add that the microphone should be turned through enough directions to
  include the two planes where the largest and the smallest levels may be
  expected, and that the field takes more than one loudspeaker, fed if
  necessary with non-coherent signals to reduce standing waves.

![Two rooms, each drawn in plan and in elevation with the listener and the chair dashed because both are out of the room. A, one loudspeaker on its axis, for the free and the quasi-free field of ISO 8253-2: the loudspeaker stands at the head height of a seated listener, its reference axis through the reference point and a distance r of at least 1 m from it; four points lie 0.15 m to the left, right, above and below the reference point, and two more on the axis a distance d in front of it and behind it. B, sound from every side, for the diffuse field of ISO 8253-2 and the test sites of ISO 4869: more than one loudspeaker around the reference point, six points on the three axes around it, and a directional microphone turned at the reference point. Under each drawing, what each clause asks. Free field, 5.2: the four lateral points within plus or minus 1 dB up to 4 kHz and 2 dB above, left and right within 3 dB of each other above 4 kHz, and with d equal to 0.15 m the level in front less the level behind within plus or minus 1 dB of the inverse distance law. Quasi-free field, 5.4: plus or minus 2 dB and d equal to 0.10 m. Plane progressive wave, ISO 4869-3 5.2.3: the two points the centres of the fixture's end faces will occupy, 145 mm apart, within 2 dB, and from 500 Hz a directional microphone reading at least 10 dB more facing the source than turned away. Diffuse field, 5.3: six points at 0.15 m within plus or minus 2.5 dB with an omnidirectional microphone kept in one orientation, and from 500 Hz the loudest and quietest directions within 5 dB for a front-to-random index of 5 dB or more. Hearing protector test site, ISO 4869-1 4.2.2: six points at 15 cm and the microphone turned through 360 degrees in the horizontal plane. Random-incidence field, ISO 4869-3 5.2.2: six positions at 150 mm with the fixture removed](https://docs.jmrp.io/phonometry/media/diagram_soundfield_qualification.svg)

*Where the microphone goes, in plan and in elevation. A is the room of 5.2 and
5.4: one loudspeaker, four points off its axis and two on it. B is the room of
5.3, whose six points and turned microphone ISO 4869-1 and ISO 4869-3 ask of
their own test sites.*

```python
# A treated room, the loudspeaker 1.2 m away: deviations from the level at
# the reference point, 125 Hz to 8 kHz, at the four lateral positions.
lateral = {
    "left": [2.6, -2.2, 0.8, 0.6, -0.4, 0.5, 0.9, -0.7, 1.2, -1.1, 1.4],
    "right": [-2.4, 1.9, -0.6, 0.3, 0.7, -0.8, 0.4, 1.1, -1.3, 1.5, -0.9],
    "up": [1.2, 2.3, 0.5, -0.9, 0.2, 0.6, -0.5, 0.8, 0.4, -1.2, 1.6],
    "down": [-1.5, -1.8, -0.9, 0.4, -0.6, -0.3, 0.7, -0.4, -0.8, 0.9, -1.7],
}
law = 20 * np.log10(1.3 / 1.1)   # the axial difference at 0.10 m, 1.45 dB
front = law + np.array([1.8, 1.2, 0.4, -0.3, 0.2, -0.4, 0.3, 0.5, -0.2, 0.6, -0.5])
quasi = hearing.check_quasi_free_sound_field(
    lateral, np.zeros(11), front_levels=front, back_levels=np.zeros(11),
    loudspeaker_distance_m=1.2,
)
print(quasi.passes)              # False: the room's modes spoil 125 Hz and 250 Hz
print(quasi.usable_frequencies)  # [ 500.  750. 1000. 1500. 2000. 3000. 4000. 6000. 8000.]

# Four loudspeakers with uncorrelated narrow bands of noise: the six positions,
# and a directional microphone of index 5 dB read in two directions.
six = {
    "front": [1.4, 0.9, 0.6, -0.8, 0.5, 1.1, -0.4, 0.7, -0.6, 0.9, 1.2],
    "back": [-1.2, -0.6, 0.8, 0.5, -0.9, -0.7, 0.6, -0.5, 0.8, -1.0, -0.8],
    "left": [0.8, 1.3, -0.5, 0.9, 0.4, -0.6, 1.0, 0.6, -0.9, 0.5, 1.4],
    "right": [-0.9, -1.1, 0.7, -0.4, -0.8, 0.9, -0.7, -0.8, 0.7, -0.6, -1.1],
    "up": [2.1, 1.6, -1.1, 0.6, 1.2, -0.9, 0.8, -1.2, 0.5, 1.3, -0.7],
    "down": [-1.8, -1.4, 0.9, -1.0, -0.5, 0.8, -1.1, 0.9, -0.4, -0.9, 0.6],
}
readings = np.zeros((2, 11))
readings[1] = [np.nan, np.nan, 3.1, 2.7, 3.4, 2.9, 3.8, 3.2, 3.6, 4.1, 4.3]
diffuse = hearing.check_diffuse_sound_field(
    six, np.zeros(11), directional_levels=readings, front_to_random_index=5.0
)
print(diffuse.passes)            # True
```

The directional test of 5.3 b) is a requirement, not an extra: without the
readings, `diffuse.directionality_judged` is false and so is `passes`, while
`uniform` and `balanced` still give the verdict of a) band by band. A
microphone of index 4 dB would allow only 4 dB and fail the 6 kHz and 8 kHz
bands of the same readings. The sound-field room also has its own ambient limits,
ISO 8253-2 Table 2 up to 12.5 kHz, derived from ISO 8253-1 for binaural
listening (footnote a); laid side by side, they are its bone-conduction limits
less 3 dB: `check_audiometric_ambient_noise(levels, presentation="sound field")`.
With narrow bands of noise as the test signal, as in this diffuse field,
footnote b asks for lower limits than the table's and prints no figure, so a
room that passes the check may still be too loud for them.

![Two panels against frequency from 125 Hz to 8 kHz. Left, a quasi-free field: the largest lateral deviation falls from 2.6 dB at 125 Hz and 2.3 dB at 250 Hz, both crossed as outside the field, to within plus or minus 2 dB from 500 Hz up, between dashed tolerance lines at plus and minus 2 dB; the axial deviation from the inverse distance law falls from 1.8 dB at 125 Hz and 1.2 dB at 250 Hz, both crossed too, to within the dotted plus or minus 1 dB lines from 500 Hz; the title reads usable from 500 Hz to 8000 Hz. Right, a diffuse field: the largest deviation of the six positions stays within plus or minus 2.5 dB, the right-left difference under 3 dB, and the directional variation from 500 Hz between 2.7 dB and 4.3 dB, under the Table 1 limit of 5 dB drawn as dashes; the title reads qualifies](https://docs.jmrp.io/phonometry/media/audiometry_sound_field.svg)

*Left, 5.4: the usable frequency range of a quasi-free field is where it meets
its requirements, here 500 Hz to 8 kHz. Right, 5.3: a diffuse field made of four
loudspeakers, qualified with a directional microphone of index 5 dB.*

**Show the code for this figure**

```python
fig, (left, right) = plt.subplots(1, 2, figsize=(13.5, 5.6))
quasi.plot(left)     # lateral and axial deviations, bands outside crossed
diffuse.plot(right)  # positions, ear sides and directional variation
plt.show()
```

No standardized reference threshold exists for a loudspeaker away from 0°
(4.8, NOTE 1). Annex B prints how much higher the level is at the ear nearer
the loudspeaker at 45° and 90°, rounded to 0.5 dB, which is what a report can
state beside an off-axis threshold:

```python
print(hearing.incidence_correction([500, 1000, 4000], incidence_angle_deg=90))
# [ 4.5  5.5 -0.5]
```

The paragraph above Table B.1 announces it from 200 Hz; the table prints rows
from 125 Hz, and the library keeps every printed row. The discrepancy is
registered in [ERRATA](https://docs.jmrp.io/phonometry/reference/errata/).

## What this guide covers

- **Covered.** ISO 8253-1:2010's maximum permissible ambient levels of Clause 11 (Tables 2 and
  4) with the earphone allowance of Table 3, the lowest hearing threshold level
  and the +5 dB relaxation, and the noise-floor margin of 11.1; the threshold
  rules of the ascending method (6.2.3.2 and 6.2.4.2, with its full and shortened
  series and the replay of a presentation sequence), the bracketing method
  (6.2.4.3), automatic recording (6.3.5) and sweep-frequency audiometry (7.5),
  each with its reliability test; the repeat at 1 kHz of 6.2.3.2 Step 3, the
  cross-hearing caution of 6.2.3.2 and the vibrotactile levels of 8.4; and the
  uncertainty budget of Annex A (A.3 to A.5), via
  `hearing.check_audiometric_ambient_noise`, `hearing.ambient_noise_limits`,
  `hearing.ascending_method_threshold`, `hearing.bracketing_method_threshold`,
  `hearing.automatic_audiometry_threshold`, `hearing.sweep_audiometry_threshold`,
  `hearing.check_retest_agreement`, `hearing.audiogram_cautions` and
  `hearing.audiometric_uncertainty`. Every cell of Tables 2 to 4, the levels of
  8.4, the components of A.3.3 and A.3.4 and the budget of Table A.2 (its four
  components, $u$ = 4.9 dB and $U$ = 10 dB) are reproduced in the test suite and
  in the conformance report.

- **Covered.** ISO 8253-2:2009's free, quasi-free and diffuse sound fields (5.2 to 5.4, with
  Table 1), the ambient limits of Clause 6 and Table 2, and the off-axis increases
  of Annex B (Table B.1), via `hearing.check_free_sound_field`,
  `hearing.check_quasi_free_sound_field`, `hearing.check_diffuse_sound_field`,
  `hearing.check_audiometric_ambient_noise` and `hearing.incidence_correction`.
  Every cell of Tables 2 and B.1 is reproduced in the test suite and in the
  conformance report, and so is the derivation of Table 2 from ISO 8253-1
  Table 4. The diffuse-field check is the one ISO 4869-3 applies to the
  random-incidence field of its test site, on the
  [hearing protectors](https://docs.jmrp.io/phonometry/perception/hearing/hearing-protectors/) page.

- **Not covered.** Nothing here presents a tone, drives an audiometer's hardware or judges a
  subject. The preparation and instruction of the subject (Clause 5), the
  familiarization of 6.2.2, the masking procedures of 6.2.3.3 and 8.5, the
  screening test of Clause 9, the audiogram symbols of Clause 10 and the
  psycho-acoustic check of 11.2 are procedures a tester follows, and the
  calibration scheme of Clause 12 belongs to IEC 60645-1 and the ISO 389 series.
  The test signal characteristics of ISO 8253-2 Clause 4, its monaural and
  binaural testing (8.2, 8.3) and testing with a hearing aid (Clause 9) are
  likewise taken as given, and so are the lower ambient limits footnote b to its
  Table 2 asks for narrow-band noise, for which it prints no figure. Speech
  audiometry is ISO 8253-3, which the library does not implement. The reference zero the thresholds are levels above is on
  the [hearing threshold](https://docs.jmrp.io/phonometry/perception/hearing/hearing-threshold/) page.

## See also

- [Hearing threshold (age and reference zero)](https://docs.jmrp.io/phonometry/perception/hearing/hearing-threshold/):
  the audiometric zero the hearing levels are measured from, and the threshold
  an otologically normal ear of a given age is expected to show.
- [Hearing Protectors (ISO 4869-1, -2, -3 and -6)](https://docs.jmrp.io/phonometry/perception/hearing/hearing-protectors/):
  the same diffuse-field test, applied to the site where an earmuff is screened
  on its test fixture.
- [Noise-induced hearing loss (ISO 1999)](https://docs.jmrp.io/phonometry/perception/hearing/noise-induced-hearing-loss/):
  what the threshold levels an audiogram records are compared against.
- API reference: [`hearing.audiometry`](https://docs.jmrp.io/phonometry/reference/api/hearing/audiometry/)
  and [`hearing.sound_field_audiometry`](https://docs.jmrp.io/phonometry/reference/api/hearing/sound-field-audiometry/).

## References

- **International Organization for Standardization. (2009).** *Acoustics — Audiometric test methods — Part 2: Sound field audiometry with pure-tone and narrow-band test signals* (ISO 8253-2:2009).
  The free, diffuse and quasi-free sound fields of 5.2 to 5.4 with Table 1, the ambient limits of Clause 6 and Table 2, and the off-axis increases of Annex B, validated against every cell of Tables 2 and B.1.
- **International Organization for Standardization. (2010).** *Acoustics — Audiometric test methods — Part 1: Pure-tone air and bone conduction audiometry* (ISO 8253-1:2010).
  The implemented standard: the maximum permissible ambient levels of Clause 11 with Tables 2 to 4, the threshold rules of 6.2.4, 6.3.5 and 7.5, and the uncertainty budget of Annex A, validated against every cell of Tables 2 to 4 and against Table A.2. Read in its European adoption BS EN ISO 8253-1:2010, which adopts it without modification.