Summary: Humid air by IEC 61094-2:2009 Annex F (CIPM-2007): density, speed of sound, ratio of specific heats, viscosity and thermal diffusivity.

---

A [reciprocity calibration](https://docs.jmrp.io/phonometry/signals/metrology/reciprocity-calibration/)
needs the air it is done in to five or six figures, which is why the standard that defines that calibration also carries the most
carefully stated model of humid air in the acoustic literature. Annex F is
that model, and this page is what the library does with it.

## The shortest useful call

```python
from phonometry import fluids

f = fluids.air(
    temperature_c=23.0,
    static_pressure_pa=101325.0,
    relative_humidity_percent=50.0,
)
print(f.density)                    # 1.1860847889882964 kg/m3
print(f.speed_of_sound)             # 345.86651725321 m/s
print(f.characteristic_impedance)   # 410.2270151343906 Pa.s/m
```

Those are the conditions of the first row of Table F.1, and the numbers are
the ones the annex prints. Every one of the ten figures it tabulates comes
back inside the rounding of its last printed digit.

![Flow diagram of the humid-air model of the annex, read from top to bottom. Four input boxes run across the top: the temperature t in degrees Celsius, valid from 15 to 27 degrees; the static pressure ps in pascals, valid from 60 to 110 kilopascals; the relative humidity H, valid from 10 to 90 per cent; and, in a dashed box, the carbon dioxide mole fraction xc, 0.000 4 unless it is measured, with a line under it saying that it enters the density, the speed of sound and the ratio of specific heats. Each measured condition reaches only the steps its own expression names, on a line of its own: the temperature drops into the saturation vapour pressure psv on the left, an exponential in the kelvin temperature with four Table F.2 coefficients, and runs across into the enhancement factor f on the right, with three; the static pressure runs across into f as well and drops into the water vapour mole fraction xw between them; and the relative humidity drops into xw alone. psv and f point into xw from either side. From xw a line branches down into the rows below, labelled with the seven quantities it enters. In the left column the compressibility factor Z, with nine coefficients, points across into the density rho in the middle, and under it the thermal conductivity ka points across into the thermal diffusivity alpha t, which also takes the density from above and the specific heat capacity Cp from below. The right column holds the zero-frequency speed of sound c0 and the ratio of specific heats kappa, with sixteen coefficients each, and the viscosity eta, with six. The five properties Table F.1 tabulates have green borders and carry its values for 23 degrees, 101 325 pascals and 50 per cent: a density of 1.186 084 8 kilograms per cubic metre, a speed of sound of 345.866 52 metres per second, a ratio of 1.400 757 3, a viscosity of 1.826 566 times ten to the minus five pascal seconds and a diffusivity of 2.115 317 times ten to the minus five square metres per second. The conductivity and the specific heat have dashed red borders: they come back from the same call, but Clause F.6 only writes them out and no column of Table F.1 prints a value for them. A key names the three borders. A box at the foot writes out the density equation, the compressibility factor, the water vapour mole fraction and alpha t as ka over rho Cp. Three lines under it say which boxes take the temperature in kelvin; give the relative standard uncertainties the annex quotes on its equations, 22 parts per million for the density, three parts in ten thousand for c0 and 3.2 parts in ten thousand for kappa; and note that dispersion moves the speed of sound by less than that uncertainty](https://docs.jmrp.io/phonometry/media/diagram_humid_air_chain.svg)

*The chain the call above runs: the water vapour comes first and every
quantity after it depends on it, and the five with a printed value carry the
one Table F.1 gives for those conditions.*

## The three conditions are not equally important

The library asks for the temperature and assumes the other two. That is not
laziness in one direction and rigour in the other; it follows from how much
each is worth. Sweeping each condition across its plausible range, with the
annex's own equations:

| Condition | Swept over | Moves the density by |
| :--- | :--- | ---: |
| Static pressure | 80 to 105 kPa | 25 % |
| Temperature | 15 to 27 °C | 4.5 % |
| Relative humidity | 0 to 100 % | 1.05 % |

The pressure dominates, and it is also the one people believe they already
know. A test site 1000 metres up sits at roughly 90 kPa. That is 11 % below
the standard atmosphere the library would otherwise assume, and it puts the
density about 13 % high, which is half a decibel on a sound power level with
nothing on the page to say so.

Humidity is the opposite case. It moves the density least, and almost nobody
knows theirs without measuring it.

So both are assumed and both are announced, once, in the same warning:

```python
import warnings

from phonometry import fluids

with warnings.catch_warnings(record=True) as caught:
    warnings.simplefilter("always")
    fluids.air(temperature_c=20.0)

print(caught[0].category.__name__)              # FluidAssumptionWarning
print(str(caught[0].message))
# air() assumed 101325 Pa and 50 % relative humidity. The pressure is the one
# worth measuring: a site 1000 m up sits near 90 kPa, and taking it for one
# standard atmosphere puts the density about 13 % high. The whole span of
# humidity, 0 % to 100 %, is worth about 1 % of the density. Pass the
# conditions to silence this.
```

One warning, not two, because a caller who supplied neither has one thing to
fix rather than two.

Supplying both is silent, which is the property somebody who measured their
air should get. Python shows the warning once per call site, so exploring at a
prompt is not drowned in it, and it derives from `PhonometryWarning`, so one
rule escalates it to an error for anyone who wants their runs to fail rather
than warn:

```python
import warnings

from phonometry import PhonometryWarning

warnings.filterwarnings("error", category=PhonometryWarning)
```

The temperature has no default at all. It is the one condition a caller has
actually measured, and there is no value that would be defensible to invent
for it.

The carbon dioxide fraction defaults quietly, to 0.000 4. That is not a guess
about the caller's air: Clause F.2 names it as the value to use for laboratory
conditions in the absence of a measurement, so the library is quoting the
annex rather than filling a hole.

## What the model fixes, and what it will not invent

Table F.1 prints five quantities. Clause F.6 gives expressions for two more,
and three follow by identity from the ones above:

```python
f.density                    # printed by Table F.1
f.speed_of_sound             # printed, at zero frequency
f.heat_capacity_ratio        # printed
f.viscosity                  # printed
f.thermal_diffusivity        # printed
f.thermal_conductivity       # Clause F.6 expression
f.specific_heat_capacity     # Clause F.6 expression
f.characteristic_impedance   # rho c
f.prandtl_number             # eta / (rho alpha_t)
f.kinematic_viscosity        # eta / rho
```

A quantity this model did not determine raises rather than returning a number
nobody printed. The error names the model, what it does determine, and the two
ways out:

```python
from phonometry.fluids import Fluid, FluidPropertyUnavailableError

measured = Fluid(
    temperature_c=10.0,
    static_pressure_pa=1.1e6,
    composition={"salinity_psu": 35.0},
    model="a density and a sound speed, measured",
    validity="",
    properties={"density": 1027.0, "speed_of_sound": 1491.0},
)
try:
    measured.heat_capacity_ratio
except FluidPropertyUnavailableError as exc:
    print(exc)
# 'a density and a sound speed, measured' does not determine
# 'heat_capacity_ratio'; it determines density, speed_of_sound. Supply the
# quantity yourself, or use a model that prints it.
```

That looks like pedantry with one fluid in the library, and stops looking like
it with two: sea water has no ratio of specific heats to give, and a
`Fluid` that invented one would be handing a caller a number with the shape of
a measurement and none of the substance.

Three attributes are not quantities and never raise. They are the record of
where the numbers came from:

```python
f.properties          # exactly what this model determined
f.model               # 'IEC 61094-2:2009 Annex F (CIPM-2007)'
f.composition         # the humidity and CO2 it was computed with
```

## The printed domain, and the difference between a fit and a fact

Annex F states, on its own page, where its equations were validated: 15 °C to
27 °C, 60 kPa to 110 kPa, 10 % to 90 % relative humidity. It states no range
for the carbon dioxide fraction.

Outside that box the library warns and still answers:

```python
import warnings

from phonometry import fluids

with warnings.catch_warnings(record=True) as caught:
    warnings.simplefilter("always")
    hot = fluids.air(temperature_c=60.0, static_pressure_pa=101325.0,
                     relative_humidity_percent=50.0)

print(caught[0].category.__name__)                    # FluidWarning
print("temperature 60 degC (stated 15 to 27)" in str(caught[0].message))  # True
print(round(hot.speed_of_sound, 2))                   # 371.52
```

The distinction matters and it is the rule the whole library follows. Air at
60 °C in a duct exists; a fit past its validated range is still arithmetic, and
refusing it would refuse a real measurement. What is refused is a state that
cannot be, and that is the whole list: a temperature at or below absolute zero,
a pressure that is not positive, a relative humidity outside 0 % to 100 %, a
carbon dioxide mole fraction outside 0 to 1, and a non-finite value for any of
the four. Each raises `ValueError` naming the argument.

## How the numbers are pinned

The oracle is Table F.1: five quantities at two condition sets, ten figures in
all. The tolerance is not chosen. A value is stated to within half a unit of
the last place it was printed to, so that is what reproducing it means, and the
test derives that bound from the printed string rather than carrying a constant
somebody picked. It is the last *place* that sets the bound, not a count of
significant figures: `1.186 084 8` carries eight of them and `2.115 317 x 10^-5`
seven, and the same rule reads both:

**The tolerance, derived from the print**

```python
from decimal import Decimal

def half_of_the_last_printed_figure(printed: str) -> float:
    """Half a unit in the last place the annex printed."""
    exponent = int(Decimal(printed).as_tuple().exponent)
    return float(Decimal(5) * Decimal(10) ** (exponent - 1))

half_of_the_last_printed_figure("1.1860848")    # 5e-08
half_of_the_last_printed_figure("2.115317e-5")  # 5e-12
```

Nobody can loosen it by accident, and a future edition quoting more figures
tightens it by itself.

Three conformance rows record how much of that allowance each set uses: 55 %
for the first, 96 % for the second, and none at all for the identity that ties
the thermal conductivity and the specific heat capacity to the diffusivity
printed between them. The 96 % is the density of the second set, and it is the
annex's own rounding rather than any slack of the library's. Reporting it as a
percentage means that if a change ever pushes it out, it is visible coming.

## Where this air is, and is not, used

Passing a `Fluid` to a measurement function is the subject of the packages
that consume it, and it comes with a rule worth stating here: where a standard
*fixes* the air of its own procedure, using better air is a departure from the
standard, and the library says so rather than letting a measurement quietly
stop reproducing the document it cites.

That is why the simplified air formulas of ISO 10534-2, ASTM E2611 and
ISO 17497-1 stay in their own modules with their own clauses, and why the
Johnson-Champoux-Allard Prandtl number of 0.71 stays frozen at 0.71 even
though the air at that state has 0.728.

Two states of that air are published as constants, because a caller has no
business writing them out twice. `materials.ANNEX_A_AIR` is the reference state
ISO 9053-2:2020 Annex A.3 fixes for an airflow-resistance measurement, 23 °C
and 50 % relative humidity at one standard atmosphere, evaluated through the
same CIPM-2007 model as any other air. `simulation.SIMULATION_AIR` is the
solver's default medium, and it is deliberately not that: 343 m/s and
1.2 kg/m³ flat, the round pair a finite-difference grid is set up with, carried
as a `Fluid` so that a run says which air it was made in rather than leaving
two floats standing side by side in a call.