# Sound Power

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/devices/emission/sound-power/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/devices/emission/sound-power/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-07-07
Updated: 2026-11-16T21:36:14Z
Standards: ISO 4871:1996
Verified: 2 conformance checks (ISO 4871:1996) https://docs.jmrp.io/phonometry/reference/conformance/#intensity-sound-power
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Sound power level by ten standardized routes (ISO 3741 to 3747, ISO 9614, ISO/TS 7849): how to choose, their grades and the ISO 4871 declaration.

---

Sound *pressure* depends on where you stand and on the room you stand in;
sound **power** does not. The sound power level $L_W$ is the total acoustic
energy per second a source radiates, referenced to $P_0 = 1\ \text{pW}$, and it is
the device-independent **emission** descriptor that goes on a datasheet,
feeds a room prediction (EN 12354) or is checked against a noise-emission
limit. phonometry implements ten standardised routes to it, split across
seven guides: an enveloping *pressure* surface in the field
(ISO 3744/3746) and the precision grade in an *anechoic room* (ISO 3745),
covered in
[Sound Power by Pressure Methods](https://docs.jmrp.io/phonometry/devices/emission/sound-power-pressure/);
the diffuse field of a *reverberation room* (ISO 3741), covered in
[Sound Power in the Reverberation Room](https://docs.jmrp.io/phonometry/devices/emission/sound-power-reverberation/);
a *small test room* for a small movable source, hard-walled with a
reference sound source (ISO 3743-1) or specially reverberant (ISO 3743-2),
covered in
[Sound Power in Small Test Rooms](https://docs.jmrp.io/phonometry/devices/emission/sound-power-test-rooms/);
the same comparison against a reference sound source taken to the room the
machine works in (ISO 3747), covered in
[Sound Power in Situ by Comparison](https://docs.jmrp.io/phonometry/devices/emission/sound-power-in-situ/);
*intensity* read at discrete points over a surface (ISO 9614-1), covered in
[Sound Intensity](https://docs.jmrp.io/phonometry/devices/emission/intensity/); the same intensity
*scanned* over that surface (ISO 9614-2), with its precision
counterpart (ISO 9614-3), covered in
[Sound Power by Intensity Scanning](https://docs.jmrp.io/phonometry/devices/emission/sound-power-intensity/);
and the *surface velocity* of the machine's own casing (ISO/TS 7849-1 and -2),
the one route that needs no acoustic measurement at all, covered in
[Sound power from surface vibration](https://docs.jmrp.io/phonometry/devices/emission/vibration-sound-power/).
This page is the front door: how to choose among them, what the accuracy
grades actually promise, and how a measured $L_W$ becomes the ISO 4871
noise-emission declaration a datasheet prints.

**Video:** [Same source, two rooms, one sound power](https://docs.jmrp.io/phonometry/media/anim_power_two_rooms.webm): The same source in an anechoic room and in a reverberation room produces different microphone pressures, and the free-field and diffuse-field formulas converge to the same sound power level L_W.

## Choosing a method

All deliver the same quantity for a source that runs steadily, a per-band
$L_W$ and an A-weighted total $L_{W\mathrm{A}}$, but under different
environments, accuracy grades and practical constraints. A source that emits in
bursts has no steady power to report, and the in situ route and the
hard-walled room of ISO 3743-1 determine its sound energy level $L_J$
instead (ISO 3747 clause 8.5, ISO 3743-1 clause 8.2).

| Method | Standard | Measured quantity | Environment | Accuracy grade | Use when |
| :--- | :--- | :--- | :--- | :--- | :--- |
| Enveloping surface | **ISO 3744** (engineering) / **ISO 3746** (survey) | Sound pressure on a hemisphere or box | Essentially free field over one or more reflecting planes | Grade 2 ($\sigma_{R0} \approx 1.5\ \text{dB}$) / grade 3 ($\approx 3.0\ \text{dB}$) | In situ or a large room; no special test facility available |
| Reverberation room | **ISO 3741** | Sound pressure in the diffuse field | Qualified hard-walled reverberation room | Grade 1 (precision) | Highest accuracy for steady, broadband sources in a lab |
| Hard-walled test room | **ISO 3743-1** | Sound pressure of the source and of a calibrated reference sound source at the same positions | An ordinary hard-walled room of at least 40 m³, no surface absorbing more than 0.20 | Grade 2 ($\sigma_{R0} \approx 1.5\ \text{dB}$ A-weighted) | A small movable source, with no qualified laboratory room at hand |
| Special reverberation room | **ISO 3743-2** | Sound pressure in a room whose reverberation time follows a prescribed curve | A special reverberation room of at least 70 m³, at most 300 m³ for the direct method with the 4 kHz and 8 kHz octaves | Grade 2 ($\sigma_{R0} \approx 2.0\ \text{dB}$ A-weighted) | Small sources tested one after another, read off the room alone or against a reference source |
| In situ comparison | **ISO 3747** | Sound pressure of the source and of a calibrated reference sound source at the same positions | The reverberant part of the room the machine works in ($\Delta L_f \ge 7$ dB) | Grade 2 ($\sigma_{R0} \approx 1.5\ \text{dB}$) / grade 3 ($\approx 4.0\ \text{dB}$) | A machine that cannot leave its installation, in a room too reverberant for an enveloping surface |
| Intensity at discrete points | **ISO 9614-1** | Normal sound intensity held at each of $N$ points, one per segment | Almost any, tolerant of steady extraneous noise | Grade 1 or 2 per band from the Annex B criteria; grade 3 on the A-weighted total | On-site where the probe stands still at each point rather than sweeping the surface |
| Intensity scanning | **ISO 9614-2** | Normal sound intensity scanned over a surface | Almost any, tolerant of steady extraneous noise | Grade 2 / 3 (from per-band field indicators) | On-site with background noise, or one machine among many |
| Anechoic room | **ISO 3745** | Sound pressure on a fixed microphone array | Qualified anechoic or hemi-anechoic room | Grade 1 (precision) | Reference-grade emission in a free-field laboratory |
| Precision intensity scanning | **ISO 9614-3** | Scanned normal intensity, tighter criteria | Almost any, tolerant of steady extraneous noise | Grade 1 (precision) | Precision on-site, with the ISO 9614-3 field-indicator checks |
| Surface vibration | **ISO/TS 7849-1** (survey) / **-2** (engineering) | Surface-averaged velocity level and a radiation factor | Any; no acoustic measurement | Upper limit ($\varepsilon = 1$) / engineering | The machine cannot be quietened, enclosed or approached with a microphone |

The pressure methods correct the surface level for the room ($K_2$) and for
background noise ($K_1$); the reverberation method needs a *qualified* room
but reaches precision grade; intensity rejects steady background energy at
the cost of a two-microphone probe and a per-band validity check; the
surface-velocity route abandons the microphone altogether and pays for it with
a radiation factor. Each method guide walks its routes in turn.

**The two parts of ISO 3743** reach engineering grade on *small movable*
sources and share a guide,
[Sound Power in Small Test Rooms](https://docs.jmrp.io/phonometry/devices/emission/sound-power-test-rooms/):
Part 1 compares the source with a reference sound source in an ordinary
hard-walled room, and Part 2 reads the power off a room built so that its
reverberation time follows a prescribed curve. Their in situ relative,
ISO 3747, has its own guide, [Sound Power in Situ by Comparison](https://docs.jmrp.io/phonometry/devices/emission/sound-power-in-situ/):
the comparison formula of the
[ISO 3741 page](https://docs.jmrp.io/phonometry/devices/emission/sound-power-reverberation/#2-the-comparison-method-eq-21)
applied to the room the machine works in instead of a qualified one, with the
background corrected at every microphone position and the grade read from the
excess of sound pressure level. The plate below draws seven of the ten: the
in situ comparison and the two small test rooms share the sound pressure
row's algebra with ISO 3741 and are drawn on their own pages.

![The seven sound power routes, one row for each measured quantity, every cell carrying the same five attributes. The sound pressure row: the ISO 3744/3746 hemispherical enveloping surface over a reflecting plane, the ISO 3745 fixed microphone array in a wedge-lined anechoic room, and the ISO 3741 diffuse field of a reverberation room. The sound intensity row: the ISO 9614-1 measurement surface cut into ten segments with the probe held still at a point in each, the ISO 9614-2 serpentine scan swept over that same surface, and the tighter ISO 9614-3 precision scan. The surface velocity row carries one route across the full width, ISO/TS 7849 with an accelerometer on a radiating casing and no microphone at all](https://docs.jmrp.io/phonometry/media/sound_power_methods.svg)

All ten deliver the same quantity, and the last one gets there without a
microphone. The claim that they *agree* is checkable, and worth checking once:
the block below drives one synthetic source through three of the routes and
compares their spectra. It is not a reproduction of the diagram above (a setup
diagram has no matplotlib equivalent), but a cross-check you can run.

**Show three routes computed on one source**

```python
import matplotlib.pyplot as plt
import numpy as np
from phonometry import emission

# One steady source (octave-band LW below) determined by three routes.
freqs = np.array([125.0, 250.0, 500.0, 1000.0, 2000.0, 4000.0])
lw_true = np.array([85.0, 88.0, 90.0, 89.0, 86.0, 82.0])

# ISO 3744: SPL at 10 positions on a hemisphere, r = 2 m (10 lg(S/S0) = 14 dB).
pres = emission.sound_power_pressure(np.tile(lw_true - 14.0, (10, 1)),
                                     "hemisphere", radius=2.0,
                                     frequencies=freqs)
# ISO 9614-2: uniform normal intensity scanned over six 0.5 m2 segments.
# In this free field Lp reads the intensity level, and the probe's
# delta_pI0 = 18 dB makes the clause 10.6 b screening evaluable.
i_n = np.tile(10.0 ** (lw_true / 10.0) * 1e-12 / 3.0, (6, 1))
lp = np.tile(lw_true - 10.0 * np.log10(3.0), (6, 1))
inten = emission.sound_power_intensity(i_n, np.full(6, 0.5),
                                       pressure_levels=lp,
                                       pressure_residual_index=18.0,
                                       frequencies=freqs, fraction=1)
# ISO 3741: comparison against a reference source of known LW = 84 dB per band.
comp = emission.sound_power_comparison(lw_true - 20.0, np.full(6, 64.0),
                                       np.full(6, 84.0), frequencies=freqs)

fig, ax = plt.subplots()
for res, style, ms, label in ((pres, "-o", 11, "pressure (ISO 3744)"),
                              (inten, "--s", 8, "intensity (ISO 9614-2)"),
                              (comp, ":^", 5, "reference source (ISO 3741)")):
    ax.semilogx(freqs, res.sound_power_level, style, markersize=ms,
                label=f"{label}: LWA = {res.sound_power_level_a:.1f} dB")
ax.set(xlabel="Frequency [Hz]", ylabel="Sound power level LW [dB]")
ax.legend()
plt.show()
```

### Before the method: the operating and mounting conditions

Choosing a route is the second decision. The first is what, exactly, is being
measured, and for most machine families it has already been made for you. A
**noise test code** is the type C standard for one family of machines: it fixes
the basic determination standard, the microphone array, the mounting and the
operating mode, and it overrides the general specifications of ISO 3744 or
ISO 3741 wherever the two disagree. The same installation, mounting and
operating conditions must then be used for the sound power level *and* for the
emission sound pressure level that stands beside it in a declaration, or the two
numbers describe two different machines.

Two rules from ISO 3744 clause 6 decide whether a determination is repeatable at
all. **Auxiliary equipment** shall be outside the test environment if it can be;
what cannot be removed or adequately quietened is counted as *part of the source*
and the reference box grows to enclose it. And the **mounting** shall be the
typical one, used or simulated: resilient mounts only if the machine is
resiliently mounted in the field, because a small machine that is a poor
low-frequency radiator on its own becomes a good one through a large panel it is
bolted to.

That is also where $\sigma_\text{omc}$ comes from. It is not a table lookup: it
is estimated from repeated determinations on the same source at the same
location, by the same operators, with the same instruments and positions,
**readjusting the mounting and re-establishing the operating conditions before
each one**. ISO 3744 Annex H notes 0.5 dB for a well-defined installation, and
much larger values where material flow or process variability dominates.

### A decision path

The table compresses into a short sequence of questions (ISO 3740:2019 draws
the same decision as two flowcharts in its clause 5, Figure 1 for the whole
family and Figure 2 for measurements in situ). Work through them in order;
the first match names the standard, and the chart after the list draws the
path.

1. **What is the number for?** A datasheet declaration or a limit check
   normally asks for engineering grade (grade 2, the preferred grade for
   noise declarations); a reference source, a product ranking or a dispute
   calls for precision (grade 1); a first walk-through of a noisy plant
   tolerates survey grade (grade 3). Grade 1 exists only in a qualified
   laboratory room (ISO 3741, ISO 3745) or via the precision intensity
   methods, both of which are here: ISO 9614-1 at discrete points in
   [Sound Intensity](https://docs.jmrp.io/phonometry/devices/emission/intensity/), ISO 9614-3 by
   scanning in [Sound Power by Intensity
   Scanning](https://docs.jmrp.io/phonometry/devices/emission/sound-power-intensity/).
2. **Can the source travel to a laboratory?** ISO 3741 wants the source
   small next to the room (volume no more than about 2 % of the room volume)
   and its noise steady; ISO 3745 wants it inside a qualified anechoic or
   hemi-anechoic room with a characteristic dimension below half the
   measurement radius, and it is the route that also yields directivity. A
   machine bolted to its foundation rules both out and leaves the in-situ
   methods. A small movable source with no qualified room at hand still has
   a laboratory route at grade 2: the small test rooms of ISO 3743.
3. **How quiet and how dry is the site?** ISO 3744 needs the background at
   least 6 dB below the source (preferably more than 15 dB) and
   $K_2 \le 4\ \text{dB}$. If only a
   3 dB margin or $K_2 \le 7\ \text{dB}$ can be met, the same microphones and formulae
   degrade gracefully to ISO 3746 at survey grade. A room too reverberant
   for either (ISO 3746 stops at $K_2 = 7\ \text{dB}$) is the case ISO 3747
   is written for: with a calibrated reference sound source at hand, the
   machine is compared with it at the same microphone positions, at
   engineering grade where the excess of sound pressure level $\Delta L_f$
   is at least 7 dB at every position and the source is not strongly
   directional, and at survey grade otherwise, as [Sound Power in Situ by
   Comparison](https://docs.jmrp.io/phonometry/devices/emission/sound-power-in-situ/) sets out.
4. **Is the background the problem?** When neighbouring machines cannot be
   switched off, or the margin is outright negative, the pressure methods
   are out. Intensity (ISO 9614-1 at discrete points, ISO 9614-2 by scanning,
   ISO 9614-3 for grade 1 by scanning) tolerates steady extraneous noise even
   some 10 dB *above* the source, because only the net energy flux through the
   surface counts; the per-band field indicators then decide the grade
   actually achieved.
5. **Can you put a microphone there at all?** A machine that cannot be stopped,
   an environment that destroys a capsule, an enclosure with no room for a
   surface: when no acoustic measurement is possible, the surface-velocity route
   estimates the same quantity from accelerometers on the casing
   (ISO/TS 7849-1 and -2). It answers a subtly different question: it
   characterises what the *structure* radiates, and stays blind to sound
   escaping through openings, intakes and gaps, and it costs a radiation factor
   you must either assume (Part 1, an upper limit) or measure once (Part 2).
   If a microphone can be put there and nothing above matched, the route is
   intensity again: a note to clause 4.3 of ISO 3746 sends a room beyond its
   $K_2$ limit to ISO 3747 or ISO 9614, and Figure 1 of ISO 3740:2019 lists
   ISO 9614 at every one of its in situ exits.
6. **How repeatable is the machine itself?** None of the above bounds
   $\sigma_\text{omc}$, and on a process machine it can outrank the whole grade
   choice: a grade-1 facility around a source whose own operating conditions
   scatter by 2 dB buys a $\sigma_\text{tot}$ no better than grade 2. Estimate it
   before paying for the room.

![A flow chart of this guide's six questions, read downwards. A dashed gate across the top, question 1, what is the number for, holds three chips: grade 2 engineering for a declaration or a check against a limit, grade 1 precision for a reference source, a ranking or a dispute, grade 3 survey for a first walk-through, with a note that grade 1 is reached only by ISO 3741, ISO 3745, ISO 9614-1 and ISO 9614-3. Four diamonds follow down the left, each with its exits in green boxes to the right. Question 2, can the source travel to a laboratory: yes leads to ISO 3741 in a qualified reverberation room and ISO 3745 in an anechoic or hemi-anechoic room, both grade 1, and to the small test rooms of ISO 3743-1 and -2 at grade 2. Question 3, how quiet and how dry is the site, has three exits: ISO 3744 at grade 2 with the background at least 6 dB below the source, preferably more than 15 dB, and K2 at most 4 dB; ISO 3746 at grade 3 with a 3 dB margin and K2 at most 7 dB; and ISO 3747, the comparison in situ at grade 2 or 3, for a room too reverberant for both, given a reference sound source and an excess of sound pressure level of at least 7 dB. Question 4, is the background the problem: yes leads to intensity by ISO 9614-1 and -2, at discrete points or by scanning. Question 5, can a microphone be put there at all: yes returns to the intensity box, no leads to ISO/TS 7849-1 and -2, the surface velocity of the casing with a radiation factor. A grey line collects every exit down the right-hand side into a second dashed box, question 6, how repeatable is the machine itself, with the total standard deviation as the root sum of squares of the method's and the operating and mounting one](https://docs.jmrp.io/phonometry/media/diagram_sound_power_decision_path.svg)

*The six questions as a path. Question 1 is a gate: it fixes the grade and names
no standard. Question 6 is a check every route passes through on its way out.
The four between them are the decisions, read downwards, and the first green
box reached names the standard. The order is this page's own: Figure 1 of
ISO 3740:2019 asks whether only the structure-borne part is wanted before
anything else, and carries neither the gate nor the check. Where the
background is the obstacle, both its flowcharts name ISO 9614-1 and -2 only,
and the plate follows them there.*

### What the accuracy grades mean

The grade is a claim about **reproducibility**: $\sigma_{R0}$ is the standard
deviation you would see if different laboratories measured the same source,
each following the standard correctly. Typical A-weighted values are
$\sigma_{R0} \approx 0.5\ \text{dB}$ for grade 1 (ISO 3741), 1.5 dB for grade 2 (ISO 3744, ISO 3743-1, ISO 3747,
ISO 9614-2; 2.0 dB for ISO 3743-2) and 3 dB or more for grade 3 (larger still when $K_2$ is
large or the spectrum is tonal). Per-band values are larger at the
spectrum edges. The `uncertainty` field of the pressure-method results
(enveloping surface and anechoic) is the expanded uncertainty
$U = 2\sigma_\text{tot}$ (95 % coverage), where
$\sigma_\text{tot} = \sqrt{\sigma_{R0}^2 + \sigma_\text{omc}^2}$ also folds
in the operating/mounting instability
$\sigma_\text{omc}$ that you estimate and pass in; the grade only bounds the method's
share of the budget.

In practice: a grade-2 $L_{W\mathrm{A}}$ of 92.4 dB carries $U \approx 3\ \text{dB}$, so two grade-2
results 2 dB apart are statistically indistinguishable, and checking that
same source against a 93 dB limit is a coin flip. Choose the grade from the
decision the number has to support, not from the facility that happens to
be free. And do not leave $\sigma_\text{omc}$ at its default of zero while doing
so: `emission.sound_power_pressure(..., sigma_omc_db=2.0)` on a machine whose
installation really does scatter by 2 dB turns a $U$ of 3.0 dB into 5.0 dB, and
the honest number is the one the declaration has to absorb.

**All these routes describe a steady source.** Each formula energy-averages over
an interval and reports a *rate* of energy flow, so it characterises the machine
only if the machine is time-invariant over that interval. A cyclic machine is
admissible when the averaging time spans a whole number of cycles, and the cycle
then becomes part of the declared operating condition. For a single event (one
press stroke, one door slam), the descriptor is not $L_W$ at all but the **sound
energy level $L_J$**, the same surface integral taken over the event rather than
per second, defined in the same clauses of ISO 3741, ISO 3744 and ISO 3746 and
determined by `sound_energy_pressure` over the [enveloping
surface](https://docs.jmrp.io/phonometry/devices/emission/sound-power-pressure/#4-sound-energy-level-of-a-burst-clause-83),
by `sound_energy_reverberation`, or `sound_energy_comparison` against a
reference source, in the [reverberation
room](https://docs.jmrp.io/phonometry/devices/emission/sound-power-reverberation/#4-sound-energy-level-of-a-single-event-clause-92),
by `sound_energy_hard_walled` in the hard-walled room of ISO 3743-1, in [Sound
Power in Small Test Rooms](https://docs.jmrp.io/phonometry/devices/emission/sound-power-test-rooms/#bursts-the-sound-energy-level-eq-15-to-20),
and by `sound_energy_in_situ` where the machine works, in [Sound Power in Situ
by Comparison](https://docs.jmrp.io/phonometry/devices/emission/sound-power-in-situ/#2-impulsive-sources-the-sound-energy-level-eq-13-to-20).
The practical test: if the A-weighted level
at a fixed microphone moves by more than a few decibels over the averaging time,
the source is not steady for these methods.

![Two panels. Left, one measured A-weighted sound power level of 92.4 dB drawn three times with the expanded uncertainty of each accuracy grade, 1.0 dB at grade 1, 3.0 dB at grade 2 and 6.0 dB at grade 3, against a dashed declared limit of 93 dB(A) that only the grade-1 interval clears. Right, the ISO 4871 Annex B example as two stacked bars: operating mode 1 at 88 dB plus a 2 dB uncertainty giving a declared 90 dB, verified by a level of 89 dB, and operating mode 2 at 95 plus 2 giving 97 dB, not verified by a level of 98 dB](https://docs.jmrp.io/phonometry/media/sound_power_grades_declaration.svg)

*What the grade buys, and what the declaration has to absorb. The left bar is
the width the method contributes; the right panel is the same width turned into
$K_{W\mathrm{A}}$ and added to the measured value, which is why a wide grade produces a
high declared limit and a machine that is easy to declare.*

**Show the code for this figure**

```python
import matplotlib.pyplot as plt
import numpy as np

# The interval each grade puts around one measured LWA, through the library's
# own uncertainty combination; then the ISO 4871 Annex B declaration.
fig, (axl, axr) = plt.subplots(1, 2, figsize=(12, 5.2))
for k, (name, sigma_r0) in enumerate([("Grade 1", 0.5), ("Grade 2", 1.5),
                                      ("Grade 3", 3.0)]):
    u = float(emission.precision_uncertainty(sigma_r0, 0.0, 2.0))
    axl.errorbar([k], [92.4], yerr=[[u], [u]], fmt="o", capsize=10)
axl.axhline(93.0, linestyle="--")
axl.set(ylabel="LWA [dB re 1 pW]", ylim=(84.0, 101.0))

for k, (level, k_wa, l1) in enumerate([(88.0, 2.0, 89.0), (95.0, 2.0, 98.0)]):
    axr.bar([k], [level], width=0.5)
    axr.bar([k], [k_wa], width=0.5, bottom=level)
    axr.plot([k], [l1], "D", markersize=10)
axr.set(ylabel="A-weighted sound power level [dB]", ylim=(80.0, 106.0))
plt.show()
```

## Declaring the noise emission (ISO 4871)

A measured sound power level is not yet a *declaration*. ISO 4871:1996 is the
standard for the noise-emission declaration a manufacturer prints in technical
documents: which quantities are stated, in which form, and how a declared value
is verified. The preferred quantity is the A-weighted sound power level
$L_{W\mathrm{A}}$, optionally accompanied by the A-weighted emission sound pressure level
$L_{p\mathrm{A}}$ at a work station.

![A calculation chain in three tiers. At the top, two boxes: the measured A-weighted sound power level L_WA, from a basic standard, preferably of grade 2 or better, on one machine and not rounded, eighty-eight decibels in operating mode 1 and ninety-five in operating mode 2; and its uncertainty K_WA, from the reproducibility in the noise test code or, without a code, two point five decibels at grade 2 and four at grade 3, two decibels in both modes. Joined lines carry both into the two forms the noise test code chooses between: the dual-number form, with L_WA and K_WA each rounded to the nearest decibel and always stated together, and the single-number form, L_WAd, the unrounded sum rounded once, ninety and ninety-seven decibels. Below a dashed line, a box for the verification value L_1, measured to the same noise test code or, where there is none, a basic standard of the same or better grade, under the same operating conditions, with a lower grade only by agreement. The declared value runs down the right-hand side into both verdicts, and two arrows from L_1 lead to them: declaration verified for mode 1, where L_1 is eighty-nine decibels against ninety, and declaration not verified for mode 2, where it is ninety-eight against ninety-seven. A box at the foot carries K_WA equal to one point six four five times sigma R for one machine, L_WAd equal to L_WA plus K_WA, and L_1 no greater than L_WAd, or than L_WA plus K_WA, whichever form was declared; a note underneath says that L_pA at the work station is declared the same way, with a K_pA from ISO 11201 to ISO 11204](https://docs.jmrp.io/phonometry/media/diagram_noise_declaration_chain.svg)

*What a declaration adds to a measured sound power: an uncertainty, one of two
forms chosen by the noise test code, and a declared value that a new measurement
on one machine either verifies or does not.*

A declaration takes one of two alternative forms (clause 4):

- the **dual-number** form (clause 3.16): the measured value $L_{W\mathrm{A}}$ and its
  uncertainty $K_{W\mathrm{A}}$ stated together but separately; and
- the **single-number** form (clause 3.15): the derived declared value
  $L_{W\mathrm{Ad}} = L_{W\mathrm{A}} + K_{W\mathrm{A}}$, an upper limit that repeated measurements are unlikely
  to exceed at the stated confidence level.

$K_{W\mathrm{A}}$ combines the measurement (reproducibility) and, for a batch, the
production spread; for a single machine $K = 1.645\,\sigma_\mathrm{R}$ (Annex A.2.2),
where 1.645 is the 95 % one-sided quantile of the normal distribution: the
declared value is an upper limit, so only one tail is being covered. A
[`NoiseEmissionDeclaration`](https://docs.jmrp.io/phonometry/reference/api/power/sound-power/)
holds one or more per-operating-mode declarations and renders the ISO 4871
fiche through `.report()`. The quickest route is to `declare()` straight from a
measured sound power:

```python
import numpy as np
from phonometry import emission
from phonometry import ReportMetadata

# ISO 3744: SPL at 10 positions on a hemisphere, r = 1 m (10 lg(S/S0) = 8 dB).
levels = np.tile(lw_true - 8.0, (10, 1))
result = emission.sound_power_pressure(levels, "hemisphere", radius=1.0,
                                 frequencies=freqs)

declaration = result.declare(
    uncertainty=2.0,                 # K_WA in dB (defaults to the expanded U)
    machine="Type 990, Model 11-TC",
    operating_conditions="50 Hz, 230 V, rated load",
    basic_standards="ISO 3744",
    verification_level=result.sound_power_level_a,  # L_1 for clause 6.2
)
declaration.report(
    "iso4871.pdf",
    metadata=ReportMetadata(measurement_standard="ISO 3744"),
)   # -> L_WAd = L_WA + K_WA, verified when L_1 <= L_WAd
```

Or build the declaration directly, reproducing the ISO 4871 Annex B example
(two operating modes, $L_{W\mathrm{A}} = 88$ and 95 dB with $K_{W\mathrm{A}} = 2$ dB, giving
declared $L_{W\mathrm{Ad}} = 90$ and 97 dB):

```python
mode1 = emission.OperatingModeDeclaration(
    "Operating mode 1", sound_power_level=88.0, sound_power_uncertainty=2.0,
    emission_pressure_level=78.0, emission_pressure_uncertainty=2.0,
    verification_level=89.0,          # passes: 89 <= 90
)
mode2 = emission.OperatingModeDeclaration(
    "Operating mode 2", sound_power_level=95.0, sound_power_uncertainty=2.0,
    emission_pressure_level=86.0, emission_pressure_uncertainty=2.0,
    verification_level=98.0,          # fails: 98 > 97
)
emission.NoiseEmissionDeclaration(
    (mode1, mode2), machine="Type 990, Model 11-TC",
    basic_standards=("ISO 3744", "ISO 11202"), form="dual-number",
).report("iso4871.pdf")
```

### Verifying a declared value

Verification is a separate measurement, made by somebody else. One machine of
the declared type is measured, giving a single level $L_1$, and the declaration
holds when $L_1$ does not exceed the declared value (clause 6.2). That is the
`verification_level` argument: a number the manufacturer does not supply.

What makes such a measurement admissible is clause 6.1, and it is the part most
often skipped: the verification shall follow the **same noise test code** as the
declaration or, where none exists, a basic determination standard of accuracy
grade at least equal to the one behind the declared value, under **identical
operating conditions**. A lower grade may be used only by agreement
between manufacturer and user, and the difference in accuracy accounted for.
Without that, the comparison is measuring the procedure and not the machine.

The rounding is a trap of its own. A **single-number** declaration is verified
against $L_{W\mathrm{Ad}}$, the unrounded sum $L_{W\mathrm{A}} + K_{W\mathrm{A}}$ rounded once; a
**dual-number** declaration is verified against
$\mathrm{round}(L_{W\mathrm{A}}) + \mathrm{round}(K_{W\mathrm{A}})$, which can differ by a whole
decibel. The library exposes both verdicts separately: `verified` for the
single-number rule and `verified_dual` for the dual-number one, so a
dual-number declaration must be read from the second, not from the first.

One caveat on the first snippet above: it passes
`verification_level=result.sound_power_level_a`, which verifies the declaration
against the very value it was built from. That always passes and is an
illustration of the plumbing, not of a verification.

The other two arguments deserve their provenance. `emission_pressure_level` is
the A-weighted **emission sound pressure level** $L_{p\mathrm{A}}$ at the work station,
determined by ISO 11201 (essentially free field), ISO 11202 (survey, with an
environmental correction) or ISO 11204 (with local corrections): none of which
is implemented here, so it arrives as a measured value from elsewhere, taken
under the same installation and operating conditions as the sound power.
`basic_standards` names the determination standards behind those two numbers,
and the **noise test code** in the fiche footnote is the type C standard for the
machine family that fixed the operating and mounting conditions and the
work-station positions in the first place.

The example fiche is regenerated with `make reports` and kept rendered in the
repository; click the preview to open the PDF.

**Report:** [ISO 4871 noise emission declaration example report (PDF)](https://docs.jmrp.io/phonometry/media/iso4871_declaration_example.pdf): One-page ISO 4871 noise-emission declaration fiche: a header with the machine identification and operating conditions, the declared dual-number table across two operating-mode columns listing the measured A-weighted sound power level L_WA, its uncertainty K_WA, the emission sound pressure level L_pA and the derived declared value L_WAd = L_WA + K_WA (90 and 97 dB), the noise-test-code and basic-standards footnote, and a clause 6.2 verification table where mode 1 passes and mode 2 fails.

Noise emission declaration fiche (NoiseEmissionDeclaration.report), the ISO 4871 Annex B dual-number table with the declared L_WAd = L_WA + K_WA and the clause 6.2 verification verdict.

## What this guide covers

- **Covered.** The ISO 3740 decision guidance condensed into the method table and the
  decision path, the operating and mounting conditions that precede it, what the
  accuracy grades claim about reproducibility, and the ISO 4871 noise-emission
  declaration (`NoiseEmissionDeclaration`, `OperatingModeDeclaration`,
  `declare()`): the dual- and single-number forms, $L_{W\mathrm{Ad}} = L_{W\mathrm{A}} + K_{W\mathrm{A}}$
  and the clause 6.2 verification with both rounding rules (`verified`,
  `verified_dual`), rendered as the declaration fiche through `.report()`.

- **Not covered.** The ten determinations themselves live in their method guides: the enveloping
  surface and the anechoic array in [Sound Power by Pressure
  Methods](https://docs.jmrp.io/phonometry/devices/emission/sound-power-pressure/), the
  diffuse-field method in [Sound Power in the Reverberation
  Room](https://docs.jmrp.io/phonometry/devices/emission/sound-power-reverberation/), both scanning
  methods in [Sound Power by Intensity
  Scanning](https://docs.jmrp.io/phonometry/devices/emission/sound-power-intensity/), the
  discrete-point summation of ISO 9614-1 in [Sound
  Intensity](https://docs.jmrp.io/phonometry/devices/emission/intensity/), beside the Annex A field
  indicators it lends the scanning parts, and the surface-velocity route in
  [Sound power from surface
  vibration](https://docs.jmrp.io/phonometry/devices/emission/vibration-sound-power/), the small
  test rooms of ISO 3743-1 and ISO 3743-2 in [Sound Power in Small Test
  Rooms](https://docs.jmrp.io/phonometry/devices/emission/sound-power-test-rooms/), and the in
  situ comparison of ISO 3747 in [Sound Power in Situ by
  Comparison](https://docs.jmrp.io/phonometry/devices/emission/sound-power-in-situ/).
  The sound *energy* level $L_J$ of a single event is covered for ISO 3744,
  ISO 3746, ISO 3741, ISO 3743-1 and ISO 3747, each on its own page; ISO 3745
  and ISO 3743-2 define none.
  The emission sound pressure level $L_{p\mathrm{A}}$ that
  stands beside $L_{W\mathrm{A}}$ in a declaration is consumed, not determined: ISO
  11201, ISO 11202 and ISO 11204 are outside the library. ISO 4871's batch
  criteria of clause 6.3 are not evaluated: only the clause 6.2 single-machine
  verification, and its batch statistics beyond the single-machine
  $K = 1.645\,\sigma_\mathrm{R}$ case (Annex A.2.2) are stated, not derived.

## See also

- [Sound Power by Pressure Methods (ISO 3744 / ISO 3746 / ISO 3745)](https://docs.jmrp.io/phonometry/devices/emission/sound-power-pressure/):
  the enveloping surface with its $K_1$/$K_2$ corrections, and the precision
  anechoic array.
- [Sound Power in the Reverberation Room (ISO 3741)](https://docs.jmrp.io/phonometry/devices/emission/sound-power-reverberation/):
  the precision diffuse-field method with the Waterhouse and meteorological
  corrections.
- [Sound Power in Small Test Rooms (ISO 3743)](https://docs.jmrp.io/phonometry/devices/emission/sound-power-test-rooms/):
  a small movable source in a hard-walled room against a reference sound
  source, or in a special reverberation room read directly or against one.
- [Sound Power in Situ by Comparison (ISO 3747)](https://docs.jmrp.io/phonometry/devices/emission/sound-power-in-situ/):
  the reference-source comparison where the machine works, with the sound
  energy level of an impulsive source.
- [Sound Power by Intensity Scanning (ISO 9614-2 / ISO 9614-3)](https://docs.jmrp.io/phonometry/devices/emission/sound-power-intensity/):
  the routes that tolerate steady background noise, qualified by their field
  indicators.
- [Sound power from surface vibration (ISO/TS 7849)](https://docs.jmrp.io/phonometry/devices/emission/vibration-sound-power/):
  the surface-velocity route, for a machine no microphone can approach.
- [Sound Intensity (p-p)](https://docs.jmrp.io/phonometry/devices/emission/intensity/): the two-microphone
  probe behind every intensity route, and the ISO 9614-1 determination at
  discrete points.
- [Room Acoustics](https://docs.jmrp.io/phonometry/buildings/rooms/room-acoustics/): the reverberation
  time and equivalent absorption area that feed the room corrections.
- [Predicting Sound Insulation (EN 12354)](https://docs.jmrp.io/phonometry/buildings/design/insulation-prediction/): the
  airborne and impact insulation ratings and the EN 12354 emission predictions
  that consume a source $L_W$.
- [Levels](https://docs.jmrp.io/phonometry/signals/levels/levels/): energy averaging and the A-weighting behind $L_{W\mathrm{A}}$.
- [Theory](https://docs.jmrp.io/phonometry/reference/theory/environment-transport/): the Waterhouse, $K_1$/$K_2$ and $C_1$/$C_2$ derivations.
- API reference: [`emission.sound_power`](https://docs.jmrp.io/phonometry/reference/api/power/sound-power/), [`emission`: sound power in a reverberation room](https://docs.jmrp.io/phonometry/reference/api/power/sound-power-reverberation/) and [`emission`: sound power by intensity scanning](https://docs.jmrp.io/phonometry/reference/api/power/sound-power-intensity/).
- Theory: [Sound power determination](https://docs.jmrp.io/phonometry/reference/theory/environment-transport/#sound-power-determination-iso-374437453746-iso-3741-iso-9614-123): the enveloping-surface, reverberation-room and intensity derivations behind the six acoustic routes.

## Quick answers

### What is the difference between sound power and sound pressure?

Sound pressure depends on where you stand and on the room; sound power does
not. The sound power level $L_W$ is the total acoustic energy per second a
source radiates, referenced to $P_0 = 1\ \text{pW}$, and it is the device-independent
emission descriptor that goes on a datasheet or is checked against a
noise-emission limit; ISO 3744, ISO 3741, ISO 3743-1, ISO 3743-2, ISO 3747,
ISO 9614-1, ISO 9614-2, ISO 3745, ISO 9614-3 and ISO/TS 7849 all determine
it.

### What do the accuracy grades in sound power measurement mean?

The grade is a claim about reproducibility: $\sigma_{R0}$ is the standard deviation
you would see if different laboratories measured the same source, each
following the standard correctly. Typical A-weighted values are
$\sigma_{R0} \approx 0.5\ \text{dB}$ for grade 1 (ISO 3741), 1.5 dB for grade 2 (ISO 3744, ISO 3743-1, ISO 3747,
ISO 9614-2; 2.0 dB for ISO 3743-2) and 3 dB or more for grade 3. A grade-2 $L_{W\mathrm{A}}$ carries
$U \approx 3\ \text{dB}$, so two grade-2 results 2 dB apart are statistically
indistinguishable.

### How do I measure sound power when background noise cannot be switched off?

Use intensity: ISO 9614-1 at discrete points, ISO 9614-2 by scanning (grade 2
or 3) or ISO 9614-3 by precision scanning (grade 1). Because sound intensity
is the net energy flux through the measurement surface, steady extraneous
noise even some 10 dB above the source is tolerated, whereas the ISO 3744
pressure method needs the background at least 6 dB below the source. The
per-band field indicators then decide the grade actually achieved.

## References

- **Beranek, L. L., & Mellow, T. J. (2012).** *Acoustics: Sound fields and transducers*. Academic Press. https://doi.org/10.1016/C2011-0-05897-0
  ISBN 978-0-12-391421-7. Radiation and sound fields: the free-field and diffuse-field relations between pressure and power that the whole method family rests on.
- **International Organization for Standardization. (1996).** [*Acoustics — Declaration and verification of noise emission values of machinery and equipment*](https://www.iso.org/standard/10855.html) (ISO 4871:1996).
  The declaration section: the dual/single-number forms, L_WAd = L_WA + K_WA (clause 3.15) and the clause 6.2 verification.
- **International Organization for Standardization. (2019).** [*Acoustics — Determination of sound power levels of noise sources — Guidelines for the use of basic standards*](https://www.iso.org/standard/45107.html) (ISO 3740:2019).
  The selection guide behind 'Choosing a method': grades, environments, source-size and background criteria for the whole family.